Phase-contrast and 4D flow CMR: regurgitation, shunts, and stenosis

Clinical phase-contrast and 4D flow CMR in adults: acquisition choices, regurgitant volume and fraction for all four valves, severity thresholds, shunt ratios, stenosis velocities, advanced 4D flow measures, and the checks that separate a true result from an artifact.

Phase-contrast cardiovascular magnetic resonance (CMR) measures the volume of blood crossing a plane per heartbeat, forward and backward, without assumptions about jet shape or orifice geometry. Aortic and pulmonary regurgitation are measured directly from reverse flow above the valve; mitral and tricuspid regurgitation are measured indirectly as ventricular stroke volume from cine minus net forward flow in the corresponding great artery. The American College of Cardiology/American Heart Association (ACC/AHA) thresholds for severe aortic or mitral regurgitation are a regurgitant volume of 60 mL or more and a fraction of 50% or more. In the absence of CMR-specific consensus values, they are applied to CMR. Outcome studies place the CMR thresholds lower: an aortic regurgitant fraction above 33% and a mitral regurgitant volume above 55 mL or fraction above 40%. All results are checked against a second, independent measurement of the same flow before it is reported.

What phase-contrast flow CMR measures and where it fits

Phase-contrast CMR encodes the velocity of moving blood into the phase of the MR signal. Integrating velocity over the area of a vessel frame by frame gives instantaneous flow, and integrating flow over the cardiac cycle gives the forward volume, the reverse volume, and the net volume per beat. From these three numbers, CMR derives regurgitant volume and fraction for the individual valves, the pulmonary-to-systemic flow ratio (Qp:Qs), cardiac output, the distribution of flow between the lungs, and collateral flow. For some of these, the numbers are combined with ventricular stroke volumes from a cine short-axis stack. Peak velocity through a stenosis or across a defect gives an estimate of the pressure gradient.

The method has three properties that echocardiography lacks. It measures volume flow at any plane in the chest, independent of acoustic windows. It needs no geometric assumption about a regurgitant orifice or the shape of an outflow tract, so a crescentic orifice, an eccentric jet, or several jets are measured as their total effect on forward and ventricular volumes. It also pairs flow with ventricular volumes measured by CMR, which are more reproducible than two-dimensional echocardiographic volumes. Its weaknesses are that spatial and temporal averaging blunt peak velocities, small uncorrected phase errors accumulate into large volume errors, arrhythmia degrades segmented acquisitions, and the result depends on careful plane placement and contouring.

The 2020 ACC/AHA valvular heart disease guideline places CMR as a test for resolving uncertainty after transthoracic echocardiography. In moderate or severe aortic regurgitation with suboptimal transthoracic images or a discrepancy between clinical and echocardiographic findings, transesophageal echocardiography, CMR, or catheterization is indicated (Class 1). The indication covers assessment of left ventricular systolic function, systolic and diastolic volumes, aortic size, and regurgitation severity. In chronic primary mitral regurgitation, CMR is indicated to assess left and right ventricular volumes, function, or regurgitant severity when transthoracic echocardiography does not answer these questions satisfactorily (Class 1). The 2017 American Society of Echocardiography (ASE) recommendations on native valvular regurgitation, written with the Society for Cardiovascular Magnetic Resonance (SCMR), cite the free choice of imaging planes without acoustic-window limits and the high interstudy reproducibility of CMR volumes, which suits serial assessment. The 2022 European Association of Cardiovascular Imaging position paper takes the same integrative approach, with echocardiography first and CMR as the complementary quantitative method.

Flow quantification in adult congenital heart disease follows the 2025 ACC/AHA adult congenital heart disease guideline, in which CMR supplies Qp:Qs, right ventricular volumes, pulmonary regurgitant fraction, branch pulmonary artery flow, and conduit and baffle function. Echocardiographic Doppler quantitation is covered on the Doppler hemodynamics page, and invasive shunt and output measurement on the right heart catheterization page.

Velocity encoding: phase shift, velocity-encoding value, and encoding direction

Protons moving along a magnetic field gradient accumulate a phase shift proportional to their velocity. A phase-contrast sequence applies a bipolar gradient, two lobes of equal area and opposite sign, which leaves stationary spins with no net phase and moving spins with a phase proportional to velocity. Two acquisitions with different velocity sensitivity are subtracted, which removes phase contributions unrelated to motion and leaves a phase-difference (velocity) image in which pixel intensity is linear with velocity. Mid-gray is zero, brighter is flow in one direction, and darker is flow in the other. An acquisition also produces a magnitude image that shows anatomy and is used to draw the vessel contour (Zoghbi 2017; Nayak 2015).

The velocity-encoding value (VENC) is the velocity that produces a phase shift of 180°. Velocities above the VENC wrap around and appear as flow in the opposite direction (aliasing), seen as a black spot inside a white lumen or the reverse. Velocity noise is inversely proportional to the VENC, so a VENC far above the true peak velocity makes measurement of slow flow imprecise. Clinical practice therefore uses the lowest VENC that avoids aliasing. The physics of encoding, the velocity-to-noise ratio, and phase unwrapping are covered in the CMR physics and sequences reference.

Through-plane encoding measures velocity perpendicular to the imaging slice and is the standard for flow volume: the slice transects the vessel, and all pixels within the lumen contribute their velocity times their area. In-plane encoding measures velocity along a direction within the slice; it shows the course, direction, and timing of a jet along its length and is used to find where a stenotic jet reaches its peak. A through-plane slice is then placed perpendicular to the jet at that point. Three-directional encoding measures all three velocity components; when applied to a three-dimensional volume over the cardiac cycle it becomes 4D flow.

Acquisition protocol for clinical flow measurement

The 2020 SCMR standardized protocols specify a through-plane, retrospectively gated cine gradient-echo phase-contrast sequence for blood flow quantitation. The plane is centered in the vessel of interest, aligned orthogonally to the expected main flow direction in two spatial directions, and positioned at the isocenter of the scanner, where background phase errors are smallest. Orthogonal alignment is confirmed on two perpendicular views of the vessel, such as a left ventricular outflow tract view and a coronal view for the ascending aorta. A plane tilted from the vessel axis cuts an elliptical cross section. The volume flow is largely preserved, because the larger area offsets the smaller through-plane velocity component, but peak velocity is underestimated by the cosine of the angle (Schulz-Menger 2020).

ParameterSCMR 2020 specificationReason
SequenceOne-direction (through-plane) velocity-encoded cine gradient echoVolume flow through a plane
Slice thickness5–8 mmThicker slices average high and low velocities and stationary tissue
In-plane resolutionAt least one-tenth of the vessel diameterLimits partial-volume error at the vessel edge
Temporal resolution≤50 ms between acquired framesCaptures peak flow and brief reversal; interpolated frames add no information
ECG gatingRetrospective, covering the whole cycleIncludes late diastole and atrial contraction, when reversal and regurgitation continue
VENCLowest velocity without aliasing; phase images checked after a scan and VENC raised if aliasing is presentMaximizes velocity-to-noise ratio
Echo timeShortest available, particularly across stenosesReduces signal loss from intravoxel dephasing in jets
Respiratory strategyBreath-hold with k-space segmentation, or navigator-gated or multiple-average free breathingFree breathing suits patients who cannot hold their breath

Adapted from the 2020 SCMR standardized protocols (Kramer 2020) and post-processing recommendations (Schulz-Menger 2020). The post-processing document also asks for 8 to 16 pixels across the vessel and at least 11 to 16 frames per cardiac cycle for through-plane flow.

Choosing the VENC. The VENC is set from the expected peak velocity at the plane: a recent echocardiogram, a previous CMR study, or a rapid velocity scout gives the starting point. In the absence of stenosis, about 150 cm/s suits the aorta and main pulmonary artery, 100 to 150 cm/s intracardiac flow, and 50 to 80 cm/s venous flow and the pulmonary arteries of a Fontan circulation. A dissection false lumen needs 50 to 150 cm/s. When stenosis is suspected and no prior velocity is known, 250 cm/s is a reasonable starting point (Bissell 2023). In the SCMR protocol example, a VENC of 250 cm/s aliased at the sinotubular junction in aortic stenosis and was repeated at 350 cm/s without aliasing (Kramer 2020).

Breath-holding and free breathing. A segmented breath-hold acquisition takes roughly 15 to 20 heartbeats. Flow changes during breath-holding: large inspiratory and end-expiratory breath-holds alter venous return and great-artery flow, so breath-hold measurements are not identical to measurements during quiet breathing (Nayak 2015). Free-breathing acquisitions with multiple signal averages or respiratory navigators average flow over the respiratory cycle and allow higher spatial or temporal resolution at the cost of time. For Qp:Qs and regurgitant fraction, the aortic and pulmonary planes are acquired with the same respiratory strategy so the two measurements reflect the same physiology.

Gating and arrhythmia. Retrospective gating sorts data into phases normalized to the average RR interval. It covers the whole cycle but assumes beats of similar length, and it can conceal errors when beat-to-beat variation is large (Kramer 2020). Atrial fibrillation and frequent ectopy blur the velocity curve and make stroke volume from segmented cine and flow less reliable. Arrhythmia rejection, prospective triggering, or real-time phase-contrast sequences that acquire beats separately are used in those patients, and ratios such as regurgitant fraction are more robust than absolute volumes (Bissell 2023).

Where flow planes are placed for common questions

The site of a plane follows from the lesion. For a semilunar valve, the plane sits just above the valve so that reverse flow into the ventricle is captured before it is diluted by coronary filling or vessel compliance. For the atrioventricular valves, flow is taken at the semilunar valve of the same ventricle and compared with ventricular stroke volume.

QuestionPlane position (SCMR 2020)Comment
Aortic regurgitationPerpendicular to the aortic root about 5 mm above the valve, prescribed at end-diastole; a second plane across the descending aorta at the level of the main pulmonary artery for diastolic flow reversalProximal planes record more reverse flow than distal ascending planes
Mitral regurgitationAortic plane at the sinotubular junction at end-diastole, with a full left ventricular short-axis cine stackIndirect method
Tricuspid regurgitationPulmonary plane about 5 mm above the valve at end-diastole, with right ventricular volumesIndirect method
Pulmonary regurgitationPerpendicular to the main pulmonary artery about 5 mm above the valve at end-diastoleDirect method
Aortic, pulmonary, mitral, or tricuspid stenosisThrough-plane, parallel to the valve at the point of peak flow disturbance found on a long-axis cine; alternatively in-plane along the jetShortest echo time; VENC above the expected peak
Left ventricular outflow obstructionIn-plane in the three-chamber view or through-plane perpendicular to the outflow jetGradient estimate in hypertrophic cardiomyopathy
Shunt (Qp:Qs)Ascending aorta at the sinotubular junction and main pulmonary arteryNet flows; branch pulmonary arteries and pulmonary veins as checks
CoarctationDescending aorta just distal to the coarctation and at the diaphragmCollateral flow from the increase in flow toward the diaphragm
Lung perfusion distributionRight and left pulmonary arteriesSum checked against main pulmonary artery flow

Plane positions from the 2020 SCMR standardized protocols (Kramer 2020) and post-processing recommendations (Schulz-Menger 2020).

The proximal ascending aorta and main pulmonary artery move toward the apex in systole. A fixed plane prescribed at end-diastole therefore lies in a slightly different part of the vessel in systole and, near the valve, may cut through the valve or the sinuses in part of the cycle. Through-plane motion of the root adds an apparent velocity to all pixels in the lumen and to the surrounding tissue; it is reduced by placing the plane a little above the sinuses, and 4D flow avoids it with valve tracking.

Post-processing: contours, offset correction, and derived values

The SCMR post-processing recommendations describe a fixed sequence. Magnitude and phase images are reviewed side by side, windowed so that the vessel border is sharp. The images are checked for aliasing and for a VENC so high that the lumen shows little contrast. The vessel border is traced on all frames so that only the lumen is included and surrounding noise is excluded; contours are propagated automatically and corrected by hand because the root moves and changes shape. The flow area on the phase image can appear slightly larger than on the magnitude image, and the contour is adapted to include all flowing blood (Schulz-Menger 2020).

Measured values are forward volume, reverse volume, flow rate, and peak and mean velocity. Derived values follow from them:

Background phase offset. Eddy currents induced by the rapidly switched gradients, concomitant (Maxwell) gradient terms, and gradient nonlinearity leave a small residual phase in stationary tissue after subtraction. The offset appears as an apparent velocity in all pixels, varies gradually across the image, and grows with distance from the isocenter. Peak velocity is hardly affected. Volume is the integral of velocity over the lumen area and over the whole cycle, so a small offset in a large vessel becomes a large volume error, and a dilated vessel magnifies it. In a multicenter, multivendor phantom study of twelve 1.5-T systems at a VENC of 150 cm/s, uncorrected offsets in a 30-mm "great vessel" region up to 70 mm from isocenter ranged from 0.4 to 4.9 cm/s and averaged 2.7 cm/s. An offset of 0.6 cm/s, only 0.4% of the VENC, was calculated to produce about a 5% error in cardiac output and up to a 10% error in shunt measurement (Gatehouse 2010). Correction uses a fit to stationary tissue around the vessel or a repeat acquisition of a stationary phantom with identical parameters; the ASE recommendations advise phantom or background correction whenever possible. The 2020 SCMR post-processing document regards baseline correction as optional because methods are not standardized. The 2023 4D flow consensus states that all 4D flow software should provide static-tissue correction (Zoghbi 2017; Schulz-Menger 2020; Bissell 2023).

Aliasing in post-processing. When peak velocity exceeds the VENC in a few frames, some software shifts the velocity range. A VENC of 150 cm/s spans −150 to +150 cm/s, and the window can be moved to −100 to +200 cm/s when forward flow reaches 175 cm/s and reverse flow is slow. Phase unwrapping algorithms do the same pixel by pixel. Aliased pixels normally lie in the center of the jet; an apparent alias at the edge of the vessel usually lies outside it. Unwrapping fails where the alias is not an isolated island within the flow, and the acquisition is then repeated at a higher VENC (Schulz-Menger 2020; Nayak 2015).

Peak velocity within the contour. Software reports either the single highest pixel or the average of a few adjacent pixels. The single pixel is noisy; the averaged value is smoother and lower than the true peak. Through-plane peak velocity also underestimates the true peak if the plane does not lie at the vena contracta (Schulz-Menger 2020).

Stroke volume, cardiac output, and normal flow relationships

Without regurgitation or a shunt, the net forward volume in the ascending aorta equals left ventricular stroke volume, net main pulmonary artery flow equals right ventricular stroke volume, and the two are equal. Three small physiological terms modify these identities. Coronary flow leaves the aorta from the sinuses, so a plane at or above the sinotubular junction records about 5% less than left ventricular output. The 4D flow consensus adds 5% to aortic flow above the sinuses when comparing it with pulmonary flow. Bronchial arterial flow, about 5% of aortic output, returns to the left atrium through the pulmonary veins and bypasses the pulmonary artery. Caval flow, measured as superior vena cava plus descending aortic flow, equals ascending aortic flow only when collateral flow is negligible (Schulz-Menger 2020; Bissell 2023).

Cardiac output by phase contrast is the net aortic volume multiplied by heart rate. It agrees with left ventricular stroke volume from a cine stack when the mitral and aortic valves are competent, and the comparison of the two is the simplest internal check in any study. Ventricular volumes, contouring rules, and the reference values for left and right ventricular stroke volume belong to the CMR quantitation reference.

Direct and indirect methods for regurgitant volume and fraction

The ASE divides CMR regurgitation methods into direct and indirect approaches. The direct method measures reverse flow with a through-plane phase-contrast slice. It is accurate for the semilunar valves, where the plane can sit in a relatively stationary vessel above the valve, and it is the preferred technique for aortic and pulmonary regurgitation. Direct measurement at the mitral or tricuspid valve is difficult because the annulus moves substantially toward the apex in systole and the plane captures a changing part of the valve and atrium. The indirect method compares ventricular stroke volume, by planimetry of a short-axis cine stack, with net forward flow in the great artery of the same ventricle; the difference is the regurgitant volume of the atrioventricular valve (Zoghbi 2017).

LesionPreferred methodSecondary methodsCorroborating signs
Mitral regurgitationLeft ventricular stroke volume − total aortic forward volumeLeft ventricular stroke volume − pulmonary forward volume; left minus right ventricular stroke volume; mitral inflow volume − aortic forward volumeLeft ventricular and left atrial dilation
Aortic regurgitationDirect diastolic reverse volume in the aortic rootTotal aortic forward volume − pulmonary forward volume; left minus right ventricular stroke volumeLeft ventricular dilation
Tricuspid regurgitationRight ventricular stroke volume − total pulmonary forward volumeRight ventricular stroke volume − aortic forward volume; right minus left ventricular stroke volumeRight ventricular and right atrial dilation
Pulmonary regurgitationDirect diastolic reverse volume in the main pulmonary arteryRight minus left ventricular stroke volume; right ventricular stroke volume − aortic forward volumeRight ventricular dilation

Adapted from the 2017 ASE recommendations developed with SCMR (Zoghbi 2017, Table 3). The preferred method is generally unaffected by other regurgitant lesions; the secondary methods are reliable only when other regurgitant lesions are absent or can be accounted for.

In mitral regurgitation, left ventricular stroke volume equals mitral regurgitant volume plus total aortic forward volume, and total aortic forward volume includes any aortic regurgitant volume that returned to the ventricle in diastole. The preferred formula, stroke volume minus total aortic forward volume, therefore remains valid when aortic regurgitation coexists; stroke volume minus net aortic volume would add the aortic regurgitant volume to the mitral result. The secondary method using mitral inflow minus aortic forward volume measures mitral regurgitant volume minus aortic regurgitant volume and underestimates mitral regurgitation when both are present. Left minus right ventricular stroke volume gives the difference between left-sided and right-sided regurgitation and is invalid when a shunt or any right-sided regurgitation is present.

Regurgitant fraction is regurgitant volume divided by the forward volume of the same valve: for the direct semilunar method, reverse volume divided by forward volume in the same vessel; for the indirect atrioventricular method, regurgitant volume divided by ventricular stroke volume. Fraction is less sensitive than volume to body size and to systematic offset error, and in atrial fibrillation it is the more reliable of the two (Bissell 2023).

Grading regurgitation severity by CMR

There is no consensus set of CMR-specific thresholds. The ASE recommendations note the paucity of outcome data and adopt the general echocardiographic and ACC/AHA cutoffs for regurgitant volume and fraction, while listing the CMR studies that point to lower values (Zoghbi 2017). Three groups of numbers are in use: the guideline staging values, values chosen to match echocardiographic grades, and values derived from clinical outcomes.

LesionGuideline grading (applied to CMR)CMR-specific evidence
Aortic regurgitationMild: volume <30 mL, fraction <30%. Moderate: 30–59 mL, 30–49%. Severe: ≥60 mL, ≥50% (ACC/AHA 2020)Fraction >33% predicted progression to symptoms or surgery (Myerson 2012). Echocardiographic concordance: mild ≤15%, moderate 16–25%, moderate to severe 26–48%, severe >48% (Gelfand 2006). After transcatheter aortic valve replacement, a fraction of 30% or more was associated with higher mortality and heart failure hospitalization (Ferreira-Neto 2019)
Mitral regurgitationProgressive: volume <60 mL, fraction <50%. Severe: ≥60 mL, ≥50%, effective orifice ≥0.40 cm² (ACC/AHA 2020)Volume >55 mL or fraction >40% predicted progression to symptoms or surgery (Myerson 2016). Volume ≥60 mL defined severe in an outcome cohort; CMR volume was the strongest predictor of mortality (Penicka 2018). Echocardiographic concordance thresholds as for aortic regurgitation (Gelfand 2006)
Tricuspid regurgitationSevere by echocardiography: regurgitant volume ≥45 mL (ASE 2017); CMR fractions borrowed from mitral grading: mild ≤15%, moderate 16–25%, moderate to severe 25–48%, severe >48% (ASE 2017)Functional tricuspid regurgitation: volume ≥45 mL or fraction ≥50% identified the highest-risk stratum for mortality (Zhan 2020)
Pulmonary regurgitationFraction <20% mild, 20–40% moderate, >40% severe (ASE 2017)In repaired tetralogy of Fallot, pulmonary valve replacement is timed by right ventricular volumes and function, with referral at a right ventricular end-systolic volume index above 80 mL/m² (ACC/AHA 2025; tetralogy of Fallot)

Guideline values from the 2020 ACC/AHA valvular heart disease guideline and the 2017 ASE recommendations. Outcome-derived CMR thresholds come from single cohorts and have not been tested as triggers for intervention in randomized trials.

The lower outcome thresholds reflect differences in what the two methods measure. The indirect CMR method measures the regurgitant volume that enters the atrium during systole, whereas Doppler methods based on the proximal isovelocity surface area assume a hemispheric, holosystolic orifice and overestimate late-systolic and crescentic regurgitation. In a direct comparison, echocardiographic mitral regurgitant volume averaged 17 mL more than the CMR value (Penicka 2018). With the aortic direct method, a plane above the coronary ostia can count diastolic coronary flow as reverse flow, and aortic compliance below the plane hides part of the regurgitant volume. CMR aortic regurgitant fractions therefore also run lower than echocardiographic estimates, and measurement in the distal ascending aorta gives lower values than at the sinotubular junction (Gabriel 2011).

The grade combines the fraction with ventricular size and function, the morphology of the valve, the concordance of the direct and indirect methods, and the echocardiographic findings.

CMR flow calculator: regurgitant volumes, fractions, Qp:Qs, and output

Aortic regurgitant fraction (volume)—
Mitral regurgitant volume—
Mitral regurgitant fraction—
Pulmonary regurgitant fraction (volume)—
Tricuspid regurgitant volume—
Tricuspid regurgitant fraction—
Qp:Qs (net pulmonary ÷ net aortic)—
Cardiac output (net aortic)—
—

Aortic and pulmonary regurgitant fractions are reverse ÷ forward flow in the same vessel (direct method). Mitral regurgitant volume = left ventricular stroke volume − total aortic forward flow, and tricuspid regurgitant volume = right ventricular stroke volume − total pulmonary forward flow; both fractions divide by the ventricular stroke volume (indirect method). Qp:Qs uses net flows, so semilunar regurgitation does not inflate it. The indirect results assume no ventricular-level shunt, which would make stroke volume exceed great-artery flow for reasons other than regurgitation. Thresholds are those in the severity table above.

Aortic regurgitation by phase-contrast CMR

Aortic regurgitation is the lesion phase contrast measures best. The plane is placed perpendicular to the aortic root about 5 mm above the valve, at or just below the sinotubular junction, and prescribed in end-diastole. The diastolic portion of the flow curve below zero is the regurgitant volume, and reverse divided by forward volume is the regurgitant fraction. In 107 patients, fractions measured at the sinotubular junction were higher than in the distal ascending aorta (mean 13% versus 9.4%), and the proximal ascending aorta gave the best agreement with echocardiographic grading (Gabriel 2011). In systole the compliant aorta stores part of the stroke volume, and in diastole its recoil drives some of it back through the incompetent valve. A plane farther from the valve misses the recoil volume of the segment below it, so the measured regurgitant volume falls with distance (Chatzimavroudis 1998).

Two technical issues are specific to the aortic root. Through-plane motion of the annulus moves the valve toward the plane in systole; a plane too close to the valve then includes leaflets and sinus in some frames. In a dilated aorta, slow and vortical flow lowers the velocity-to-noise ratio, and a small background offset applied to a large lumen becomes a large volume error. The SCMR post-processing recommendations allow an alternative when the aorta is large: aortic regurgitant volume is calculated as total aortic forward flow minus net pulmonary artery flow, or minus caval flow, when aortopulmonary collateral flow is not significant. This estimate slightly overestimates regurgitation because bronchial flow is about 5% of aortic output (Schulz-Menger 2020).

Holodiastolic flow reversal. A second plane in the descending thoracic aorta shows whether reverse flow continues throughout diastole. In 80 patients with a validation group of 42, holodiastolic reversal in the mid-descending aorta predicted severe aortic regurgitation by integrated echocardiographic grading with a sensitivity of 100% and a specificity of 93%, and it was 100% specific but 61% sensitive for moderate to severe regurgitation. Combining reversal with the regurgitant fraction stratified severity in the validation group (Bolen 2011). The sign is subject to the same offset error as any slow flow, so it is read only after background correction. A patent ductus, an aortopulmonary window, or a surgical systemic-to-pulmonary shunt also produces diastolic reversal in the descending aorta without valve regurgitation.

Outcome evidence. In 113 asymptomatic patients with moderate or severe aortic regurgitation by echocardiography, followed for a mean of 2.6 years after CMR, 85% of the 39 patients with a regurgitant fraction above 33% progressed to symptoms or another indication for surgery, compared with 8% of the 74 with a fraction of 33% or less. The area under the curve for the fraction was 0.93. A left ventricular end-diastolic volume above 246 mL also discriminated outcome (area under the curve 0.88), and the combination of fraction and volume performed best (Myerson 2012). The ACC/AHA guideline recognizes that left ventricular volume thresholds differ among two-dimensional echocardiography, three-dimensional echocardiography, and CMR and that outcome data for volumes are limited.

CMR also measures the ascending aorta and left ventricular ejection fraction in the same study, which together with regurgitant severity set the timing of surgery. Surgical indications in aortic regurgitation are covered in the aortic regurgitation reference.

Mitral regurgitation by the indirect volumetric method

Mitral regurgitant volume is left ventricular stroke volume from the short-axis cine stack minus total forward flow at the sinotubular junction, and regurgitant fraction is that volume divided by left ventricular stroke volume (Zoghbi 2017). Right ventricular stroke volume or net pulmonary flow serves as the internal control: with no other regurgitant lesion or shunt, either equals net aortic flow. Because the method subtracts two independently measured volumes, an error in either passes into the regurgitant volume. A 10-mL overestimate of left ventricular stroke volume from a basal slice that includes atrium, or a 10-mL underestimate of aortic flow from an uncorrected offset, adds 10 mL of apparent regurgitation. Agreement with pulmonary flow confirms the aortic flow, and agreement with mitral inflow, measured by 4D flow or by an inflow plane, confirms the stroke volume.

In 103 patients assessed by both techniques, agreement between CMR and echocardiographic grading of mitral regurgitation was modest (r = 0.6). Among the 26 who had a postoperative scan after isolated mitral surgery, the decrease in left ventricular end-diastolic volume correlated strongly with CMR regurgitant volume (r = 0.85) and not with echocardiographic severity (r = 0.32) (Uretsky 2015). In 258 asymptomatic patients with primary regurgitation, grading agreed in 76%; disagreement was concentrated in late-systolic regurgitation and multiple jets, and CMR regurgitant volume had the largest area under the curve for mortality (Penicka 2018). In 109 asymptomatic patients followed for a mean of 2.5 years, 91% of those with a CMR regurgitant volume of 55 mL or less were free of surgery at 5 years, compared with 21% of those above 55 mL. A fraction of 40% separated the groups similarly (Myerson 2016).

The indirect method suits the cases in which Doppler quantitation is weakest: eccentric jets that hug the atrial wall, multiple jets, late-systolic prolapse, and the crescentic orifice of secondary regurgitation. It is weakest when another lesion confounds the subtraction or when atrial fibrillation degrades the cine and flow data. Mechanism, leaflet anatomy, and the indications for repair are covered on the mitral regurgitation page.

Tricuspid regurgitation by CMR

Tricuspid regurgitant volume is right ventricular stroke volume minus total forward flow in the main pulmonary artery. Without aortic regurgitation, aortic forward flow can replace pulmonary flow, and without other regurgitant lesions, right minus left ventricular stroke volume gives the same result (Zoghbi 2017). Direct measurement across the tricuspid valve is of limited value because the annulus moves more than the mitral annulus and the plane captures varying amounts of atrium and ventricle. The accuracy of the indirect method depends on right ventricular contouring, which is harder than left: the basal slices include the right atrium and the outflow tract, and trabeculation is coarse. Many centers contour the right ventricle on an axial stack, which shows the tricuspid plane more clearly.

The ASE recommendations noted no CMR-specific tricuspid thresholds and borrowed the mitral grading by fraction. Outcome data have since appeared. In 547 patients with functional tricuspid regurgitation, both regurgitant fraction (adjusted hazard ratio 1.26 per 10% increment) and regurgitant volume (1.15 per 10 mL) were associated with all-cause mortality after adjustment for clinical variables and right ventricular ejection fraction. A regurgitant volume of 45 mL or more or a fraction of 50% or more identified the highest-risk stratum (Zhan 2020). The rapid growth of transcatheter tricuspid repair and replacement has increased the use of CMR to define right ventricular volumes and function before intervention, because those values carry prognostic weight that the regurgitant grade alone does not. Spin dephasing in the right atrium on steady-state free precession cine is not used to grade tricuspid regurgitation, because modern sequences minimize it (Zoghbi 2017).

Pulmonary regurgitation and repaired tetralogy of Fallot

Pulmonary regurgitation is measured directly in the main pulmonary artery, with the plane perpendicular to the vessel about 5 mm above the valve. Severe pulmonary regurgitation after repair of tetralogy of Fallot often has little color Doppler signature: with normal pulmonary pressures, the regurgitant flow is laminar, low in velocity, and brief, because pulmonary and right ventricular diastolic pressures equalize early. CMR measures the volume directly, and the ASE grades the regurgitant fraction as mild below 20%, moderate from 20% to 40%, and severe above 40% (Zoghbi 2017). An echocardiographic ratio of jet width to pulmonary annulus diameter above 0.5 correlates with severe regurgitation by CMR.

Measurement in the main pulmonary artery after repair has particular limitations. The artery is short and often dilated or aneurysmal after a transannular patch, so the plane may lie close to the bifurcation, where flow is complex and vortical. Branch pulmonary artery stenosis redirects regurgitant volume, and branch flows measured separately show how much the individual lungs contribute; the sum of the branch net flows checks the main pulmonary measurement. Through-plane motion and a large lumen magnify offset errors. When the main pulmonary artery measurement is unreliable, the regurgitant volume is cross-checked as right ventricular stroke volume minus net aortic flow, valid only without tricuspid regurgitation or a residual shunt.

The 2025 ACC/AHA adult congenital heart disease guideline bases referral for pulmonary valve replacement after tetralogy repair on a right ventricular end-systolic volume index above 80 mL/m² and other metrics, in place of the end-diastolic volume criteria of earlier practice (Gurvitz 2025). In asymptomatic adults, the decision weighs right ventricular end-systolic volume with right and left ventricular ejection fraction, the ratio of right to left ventricular volume, and the trend in objective exercise capacity.

The 2020 ESC guideline recommends replacement in symptomatic patients with severe pulmonary regurgitation or at least moderate right ventricular outflow obstruction (Class I). In asymptomatic patients with severe regurgitation, replacement should be considered (Class IIa) when one of the following is present: a decrease in objective exercise capacity; progressive right ventricular dilation to an end-systolic index of at least 80 mL/m² or an end-diastolic index of at least 160 mL/m²; progression of tricuspid regurgitation to at least moderate; or progressive right ventricular systolic dysfunction (Baumgartner 2020). After replacement, right ventricular volumes normalized when the preoperative end-systolic index was below 82 mL/m² (Oosterhof 2007), and an index above 95 mL/m² was associated with death, sustained ventricular tachycardia, or heart failure (hazard ratio 2.89) (Bokma 2016). Right ventricular volumes come from the cine stack and are covered in the CMR quantitation reference; the complete replacement criteria are covered in the tetralogy of Fallot reference.

Pulmonary regurgitation also has consequences for shunt calculations by echocardiography. Right ventricular outflow tract stroke volume by Doppler includes the regurgitant volume, so an echocardiographic Qp:Qs is inflated in severe pulmonary regurgitation. CMR uses net pulmonary flow, which removes the regurgitant component.

Mixed and multivalve regurgitation and prosthetic valves

When regurgitation affects more than one valve, the lesions are quantified by the methods least affected by the others, and the set of results is reconciled by conservation of mass. The direct semilunar measurements are unaffected by atrioventricular regurgitation. Mitral regurgitation is calculated from left ventricular stroke volume minus total aortic forward flow, which remains valid with aortic regurgitation; tricuspid regurgitation is calculated from right ventricular stroke volume minus total pulmonary forward flow, which remains valid with pulmonary regurgitation. The difference between left and right ventricular stroke volumes then equals the left-sided regurgitant volumes minus the right-sided regurgitant volumes, a relationship that checks the whole set. Combined stenosis and regurgitation of one valve is graded by forward and reverse flow together; velocity across the stenosis rises because the forward volume includes the regurgitant volume.

Prosthetic valves. Mechanical and bioprosthetic valves can be scanned at 1.5 and 3 T; the force from the beating heart greatly exceeds the force from the magnetic field (Kramer 2020). Metal in the sewing ring, stent frame, or occluder produces a susceptibility void that distorts the velocity field next to the prosthesis, so the aortic plane is moved downstream of the artifact. Paravalvular regurgitation after transcatheter aortic valve replacement is often eccentric and multiple, which limits echocardiographic grading. In 448 patients studied by CMR at a median of 12 days after transcatheter implantation, the aortic regurgitant fraction predicted mortality (hazard ratio 1.06 per 10% increase) and heart failure hospitalization (1.15 per 10%). A fraction of 30% or more, found in 3%, was associated with a hazard ratio for mortality of 2.63 (Ferreira-Neto 2019). Multi-VENC 2D and 4D flow sequences agreed with standard 2D flow for paravalvular regurgitation after transcatheter replacement, with no substantial bias (Rooijakkers 2024). Prosthetic valve assessment by echocardiography and CT is covered on the prosthetic heart valves page.

Shunt quantification: Qp:Qs by phase-contrast CMR

Qp:Qs is the ratio of net pulmonary flow to net systemic flow. For atrial and ventricular shunts, and for anomalous pulmonary venous return, Qp is net flow in the main pulmonary artery and Qs is net flow in the ascending aorta. For a patent ductus arteriosus or an aortopulmonary window, the shunt arises beyond the aortic valve: the left ventricle ejects the pulmonary venous return, so ascending aortic flow is Qp, and main pulmonary artery flow, proximal to the ductal insertion, is Qs. When both great arteries are unreliable, pulmonary venous flow (Qp) and caval flow (Qs) give an independent ratio, and the right and left pulmonary arteries and the four pulmonary veins localize flow to the individual lungs (Schulz-Menger 2020; Bissell 2023).

Phase contrast agrees closely with invasive methods. In 21 adults and adolescents (age 15 to 72 years) who had proximal aortic and pulmonary flow measured immediately before catheterization, CMR shunt magnitude correlated with oximetry and indicator dilution (r = 0.94). CMR correctly separated all patients with a Qp:Qs below 1.5 from all patients with a ratio of 1.5 or more (Hundley 1995). Unlike oximetry, CMR does not depend on the mixed venous sampling site, and unlike echocardiography it does not square a measured outflow diameter. Direct through-plane measurement of flow across an atrial septal defect is inaccurate because the defect moves and changes shape and the jet is not perpendicular to any fixed plane; Qp:Qs comes from the great arteries.

Six cardiac MR panels. Top row: end-diastolic and end-systolic horizontal long-axis cine frames showing a large defect in the basal ventricular septum. Middle row: magnitude image and through-plane phase-contrast velocity map at the aortic valve, both with a dotted red circle around the vessel lumen. Bottom row: magnitude image and velocity map at the pulmonary valve with the same dotted region of interest.
Qp:Qs from through-plane phase contrast in a large, long-standing ventricular septal defect. Cine frames in end-diastole (a) and end-systole (b) show the defect in the basal septum. Regions of interest (dotted circles) drawn on the magnitude images and applied to the velocity maps at the aortic valve (c, d) and pulmonary valve (e, f) gave a Qp:Qs close to 1:1, consistent with advanced pulmonary vascular disease limiting flow across even a large defect. Credit: Nayak KS, Nielsen JF, Bernstein MA, et al. Cardiovascular magnetic resonance phase contrast imaging. J Cardiovasc Magn Reson. 2015;17:71, Figure 11, doi:10.1186/s12968-015-0172-7, CC BY 4.0.

Interpretation of Qp:Qs. The 2025 ACC/AHA adult congenital heart disease guideline treats a Qp:Qs of 1.5 or more with enlargement of the chambers that receive the shunted volume as a hemodynamically significant atrial or ventricular shunt. The receiving chambers are right-sided for an atrial shunt and left-sided for a ventricular shunt. A patent ductus arteriosus is judged significant by left atrial or left ventricular enlargement attributable to the shunt. Closure of an atrial septal defect additionally requires an acceptable pulmonary vascular resistance and no significant left ventricular disease (Gurvitz 2025); these conditions, and the trials behind them, are covered on the atrial septal defect and patent foramen ovale page. A Qp:Qs near 1 does not exclude a large defect. Rising pulmonary vascular resistance reduces left-to-right flow, and in Eisenmenger physiology the ratio falls to 1 or below. A large defect with a balanced ratio calls for invasive measurement of pulmonary vascular resistance. Small or intermittent shunts, such as a patent foramen ovale, do not change Qp:Qs measurably, and a borderline ratio is interpreted with chamber size and anatomy.

Anatomy determines repair. Once the magnitude of a shunt is measured, its location decides treatment. A superior or inferior sinus venosus defect lies outside the true septum at the junction of a vena cava with the right atrium, commonly with partial anomalous pulmonary venous return; it has no septal rims for a device and is repaired surgically. Anomalous pulmonary veins are traced on contrast angiography or 3D whole-heart imaging, and pulmonary vein flows show how much of Qp the individual anomalous veins carry. A left ventricle-to-right atrial (Gerbode) communication appears as a high-velocity systolic jet into the right atrium that can mimic tricuspid regurgitation on echocardiography. Coronary artery fistulas and aortopulmonary collaterals are additional sources of a Qp:Qs above 1 that are found by tracing flow. The ratio alone does not locate them.

Other sources of error in Qp:Qs. Uncorrected background offset is the dominant error in shunt measurement (Gatehouse 2010). The aortic plane above the sinuses excludes coronary flow, so a perfectly balanced circulation reads a Qp:Qs of about 1.05.

Aortic coarctation and collateral flow

In coarctation of the aorta, phase-contrast flow in the descending aorta measures collateral circulation, which reflects hemodynamic severity independently of the resting gradient. In normal subjects, flow falls by a small amount from the proximal to the distal descending thoracic aorta as intercostal arteries leave it; in 10 volunteers the decrease was 7% (standard deviation 6%). In 23 patients, those with moderate to severe coarctation had an 83% (standard deviation 50%) increase in flow from just below the coarctation to the diaphragm, as intercostal and other collaterals fed blood back into the aorta; patients with mild coarctation had the normal pattern. The increase in flow correlated with the reduction in luminal diameter (r = 0.94) and the clinical gradient (r = 0.84) (Steffens 1994).

The jet velocity across a coarctation gives a Bernoulli estimate of the gradient, but collateral flow lowers the velocity across a severe narrowing, and a long-segment or tortuous narrowing defeats both Doppler and a single CMR plane. CMR angiography or CT defines the anatomy, and CMR flow and collateral volume define the physiology. The same descending aortic planes measure flow after repair or stenting, and an increase in distal flow after stenting indicates persistent collateral dependence. Coarctation management is covered in the adult congenital heart disease references.

Stenosis: peak velocity, gradient, and the limits of CMR

Peak velocity through a stenosis gives the instantaneous peak pressure gradient by the simplified Bernoulli equation, ΔP = 4V², with V in m/s and ΔP in mm Hg; a peak velocity of 4.0 m/s corresponds to a gradient of 64 mm Hg. Integrating the gradient over systole gives a mean gradient, and the velocity-time integrals in the outflow tract and at the valve give a continuity-equation valve area. In 24 patients with aortic stenosis, CMR peak and mean gradients correlated with Doppler (r = 0.83 and 0.87) and the continuity valve area correlated with the echocardiographic area (r = 0.83), with good repeatability (Caruthers 2003).

CMR tends to record lower peak velocities than continuous-wave Doppler for structural reasons. A voxel averages the velocities within it, and in a narrow jet the voxel includes slower flow at the jet edge; a 5- to 8-mm slice averages along the jet; temporal resolution of 30 to 50 ms can miss a brief peak; and a through-plane slice that is not at the vena contracta records a lower velocity. Intravoxel dephasing in turbulent jets causes signal loss that makes the velocity in the core unreliable, which is reduced by the shortest available echo time. Continuous-wave Doppler, which records the highest velocity along its beam with millisecond resolution, remains the reference for stenosis grading when its alignment is adequate. CMR has the advantage when the jet is directed away from all acoustic windows: misalignment underestimates Doppler velocity, and CMR can place its plane perpendicular to the jet whatever the direction (Nayak 2015).

The protocol therefore uses in-plane encoding along the jet to locate the peak, followed by a through-plane slice perpendicular to the jet at that level with a VENC above the expected peak (Kramer 2020). A stack of thin cine slices through the orifice allows direct planimetry of the anatomic valve area, which is valid only where the orifice is clearly delineated. In the ACC/AHA classification, a peak velocity of 4.0 m/s or more or a mean gradient of 40 mm Hg or more defines severe aortic stenosis; these thresholds were developed with Doppler and are applied to CMR, although CMR tends to underestimate peak velocity. Grading, low-flow states, and valve replacement are covered on the aortic stenosis page.

The same principles apply to pulmonary stenosis and right ventricular outflow obstruction in adults with repaired congenital disease, where distorted outflow geometry is easier to interrogate by CMR, to branch pulmonary artery stenosis, and to conduit obstruction. In mitral stenosis, CMR peak early diastolic velocity and pressure half-time agree with Doppler (Nayak 2015). In hypertrophic cardiomyopathy, phase contrast in the three-chamber view shows the outflow jet and systolic anterior motion, but the dynamic, late-peaking gradient and its provocation are measured by Doppler, as described on the hypertrophic cardiomyopathy page.

Pulmonary hypertension and pulmonary artery flow

Phase contrast in the main pulmonary artery shows how a raised pulmonary vascular resistance changes flow. Mean velocity falls, acceleration time shortens, the artery dilates, and its pulsatility falls. In 59 patients undergoing right heart catheterization, average velocity through the cardiac cycle correlated with mean pulmonary artery pressure (r = −0.73) and pulmonary vascular resistance index (r = −0.86); an average velocity of 11.7 cm/s or less identified pulmonary arterial hypertension, then defined as a mean pressure above 25 mm Hg, with a sensitivity of 92.9% and a specificity of 82.4% (Sanz 2007). With 4D flow, a vortex of blood forms in the main pulmonary artery when pressure is elevated; its relative duration in the cardiac cycle correlated with mean pulmonary artery pressure (r = 0.94), and pressure estimated from vortex duration differed from catheter values with a standard deviation of 3.6 mm Hg (Reiter 2008).

These flow markers are supportive. The hemodynamic definition of pulmonary hypertension now uses a mean pressure above 20 mm Hg. The published thresholds were derived against the older definition, and the diagnosis and classification still require catheterization. In adults with pulmonary hypertension, CMR is used for right ventricular volumes and function, for Qp:Qs in suspected shunts, and for prognosis. Pulmonary hypertension is covered on the pulmonary hypertension page.

4D flow CMR: acquisition, resolution, and scan time

4D flow CMR acquires velocity in three directions over a three-dimensional volume, resolved through the cardiac cycle. A single volume placed over the heart and great vessels allows flow through any plane to be measured after the patient has left the scanner, along with visualization of flow paths, jet direction, and flow-derived parameters. The 2015 consensus described typical spatial resolution of 1.5 to 3 mm isotropic, temporal resolution of 30 to 40 ms, and acquisition times of 5 to 25 minutes (Dyverfeldt 2015). The 2023 update gives clinical acquisition recommendations from centers using the technique routinely.

ParameterRecommendedAcceptableTrade-off
Spatial resolution, whole heart (adults)2.5 mm isotropic3 mm isotropicAccuracy against scan time
Spatial resolution, vessels (adults)2 mm isotropic2.5 mm isotropicAt least 6 voxels across the vessel diameter
Acquired temporal resolution30 ms50 msLonger allowed for visualization only
VENCMaximum expected velocityAbout 10% above the maximum, and at most 25% above itVelocity-to-noise ratio against aliasing
ECG gatingRetrospectiveProspective with arrhythmia rejection in irregular rhythmComplete cycle coverage
Respiratory compensationOptionalSelf-gating or bellows where navigators are not feasibleImage quality against scan time
Contrast agentNot requiredConsidered in dissection and other challenging casesHigher blood signal after gadolinium
Clinical scan time5–10 min—Fits in the delay before late gadolinium enhancement imaging

Adult values adapted from the 2023 4D flow consensus (Bissell 2023, Table 1). Volume coverage includes at least the four valves and the aortic and pulmonary sinuses so that internal consistency can be checked.

Acceleration. Without acceleration, three-directional encoding over a whole-heart volume at adequate resolution would take far longer than a clinical examination allows. Parallel imaging, which uses the sensitivity profiles of the receiver coil elements, and k-t methods brought acquisitions into the 5- to 15-minute range. Compressed sensing exploits the sparsity of the data to undersample further: an aortic 4D flow acquisition with a compressed-sensing acceleration factor of 7.7 took 1 minute 52 seconds on average, compared with 7 minutes 25 seconds conventionally, with net flow within 3.5% in volunteers. Peak velocity and peak flow were underestimated, generally within 13% of conventional 4D flow (Ma 2019). The trade-off applies to all heavily accelerated 4D flow: net volumes survive acceleration better than peak velocities.

Field strength and contrast. In adults, 4D flow gives adequate signal at 1.5 T. The higher signal at 3 T matters most when finer spatial resolution is needed, and 3 T gives good image quality without contrast (Bissell 2023). 4D flow uses a spoiled gradient-echo readout, so its signal depends on blood T1; after gadolinium, blood signal and velocity-to-noise ratio rise, and a higher flip angle becomes optimal. Placing the 4D flow acquisition after contrast angiography or during the wait for late enhancement uses this effect without extra contrast (Dyverfeldt 2015; Bissell 2023).

VENC and dual-VENC. A single 4D flow VENC must accommodate both the fastest jet and the slowest flow of interest. Multi-VENC and variable-VENC encoding schemes acquire high- and low-VENC data and combine them, preserving velocity-to-noise ratio in slow flow while avoiding aliasing in fast jets, at the cost of time.

4D flow post-processing: offset correction, unwrapping, valve tracking, and visualization

The 2023 consensus describes a five-step clinical workflow. Background phase offset is corrected first, by fitting linear or polynomial surfaces to static tissue, which performs as well as a stationary phantom scan; offsets increase with distance from isocenter, which matters for the large fields of view used in 4D flow. Aliasing is then corrected by phase unwrapping, which most software performs automatically, followed by visual inspection of the peak systolic and diastolic phases in all three encoding directions; regions with uncorrectable aliasing are excluded. Vessels are segmented in three dimensions or by placing two-dimensional planes orthogonal to the vessel, and contours are propagated through the cycle. Visualization and quantification follow (Bissell 2023).

Multi-panel workflow diagram for 4D flow valve tracking. Preprocessing panels show a velocity image before and after anti-aliasing, a four-chamber image with static tissue shaded and a flow curve for phase offset correction, and color velocity overlays before and after registration. Valve tracking panels show a mitral annulus plane with forward flow pathlines, a regurgitant jet plane, and a through-plane velocity curve of mitral annular motion. Velocity mapping panels show a reformatted valve contour and a flow plot of net forward volume through all four valves over time.
Post-processing workflow for 4D flow valve tracking. Velocity data are corrected for aliasing (1), background phase offset (2), and misregistration with the cine images (3). The valve annulus is tracked on two orthogonal cine views to measure forward valvular flow (4), and the regurgitant jet is tracked on a plane perpendicular to it (5). Through-plane motion of the valve is subtracted from the measured velocities (6), velocities are mapped on the reformatted plane (7), and net forward volume through the four valves serves as an internal consistency check (8). Credit: Bissell MM, Raimondi F, Ait Ali L, et al. 4D flow cardiovascular magnetic resonance consensus statement: 2023 update. J Cardiovasc Magn Reson. 2023;25:40, Figure 2, doi:10.1186/s12968-023-00942-z, CC BY 4.0.

Retrospective valve tracking. The atrioventricular annuli move toward the apex in systole and back in diastole, which is why a fixed 2D plane cannot measure mitral or tricuspid flow directly. In 4D flow, the annulus is marked on two orthogonal cine views in all frames, for example the two- and four-chamber views for the mitral valve, and a reformatted plane follows the valve through the cycle. The velocity of the annulus itself is subtracted from the measured velocities so that motion of the plane does not count as flow. Misregistration between cine and 4D flow data is corrected first. The result is a measurement of forward flow through all four valves from the same averaged cardiac cycle (Bissell 2023).

Visualization. Velocity can be displayed as color-coded planes, vector maps, maximum velocity projections, instantaneous streamlines, or time-resolved pathlines. Streamlines are drawn tangent to the velocity field at one instant and show its direction at that moment; they do not represent the paths of blood in pulsatile flow, and short streamlines limit misinterpretation. Pathlines trace where virtual particles released at a point travel through the time-varying velocity field and show the route of blood over time; they accumulate errors in proportion to the noise in the velocity data. Visualization shows the location and direction of regurgitant and stenotic jets, shunt flow, and helical and vortical patterns, and it serves as a quick quality check, because inverted velocity directions are obvious on a pathline display (Dyverfeldt 2015; Bissell 2023).

Regurgitation by 4D flow: direct jet and indirect valve-tracking methods

4D flow offers two approaches. The direct jet method places a plane perpendicular to the regurgitant jet, close to the orifice, and integrates reverse flow. The 2023 consensus recommends it for a single central jet, as in aortic and pulmonary regurgitation, functional mitral regurgitation, and atrioventricular valve regurgitation after repair of an atrioventricular septal defect. It makes no assumption about jet shape or about conservation of mass across other valves or the septa. The indirect method with valve tracking subtracts net aortic flow from mitral forward flow (or net pulmonary flow from tricuspid forward flow), with all four flows taken from the same dataset. It is recommended when there are multiple jets in different directions or when the jet has uncorrectable aliasing, which is common because regurgitant jets are fast, turbulent, and incoherent (Bissell 2023).

A systematic review of 4D flow quantification of mitral regurgitation found seven eligible studies; retrospective valve tracking, examined in five, was reliable and accurate, and 4D flow was judged potentially as accurate as echocardiography and complementary to it, with outcome data still lacking (Fidock 2019). In 21 patients studied with both conventional multiplanar CMR and 4D flow at 3 T, direct and indirect 4D flow measurements of mitral and tricuspid regurgitation agreed closely with conventional CMR (r = 0.813 to 0.985), and direct interrogation of the jet had high interobserver and intraobserver reproducibility (Feneis 2018).

The consensus advises cross-checking all 4D flow regurgitation measurements against a standard method. A discrepancy above 15 mL or 10% between methods prompts re-analysis using conservation of mass: left ventricular stroke volume from cine equals mitral forward flow plus aortic backward flow, and equals aortic forward flow plus mitral backward flow. Valve tracking can be applied in atrial fibrillation with caution, because flows may be underestimated, and regurgitant fraction is then more reliable than regurgitant volume (Bissell 2023). All published prognostic thresholds come from 2D phase contrast and cine volumetry.

Internal consistency checks for 2D and 4D flow

All flow measurements in a CMR study have at least one independent counterpart, and agreement between them is the main protection against an error that looks plausible. The checks below come from conservation of mass and apply to both 2D and 4D flow.

CheckExpected relationshipA mismatch suggests
Aortic versus pulmonary net flowEqual without a shunt (aortic flow above the sinuses is about 5% lower because of coronary flow)Background offset, plane misplacement, or an unrecognized shunt
Net aortic flow versus left ventricular stroke volumeEqual without mitral or aortic regurgitationRegurgitation, basal-slice contouring error, or flow error
Net pulmonary flow versus right ventricular stroke volumeEqual without tricuspid or pulmonary regurgitationRegurgitation or right ventricular contouring error
Left minus right ventricular stroke volumeEquals left-sided minus right-sided regurgitant volumes without a shuntInconsistent regurgitant estimates
Right plus left pulmonary artery flowEquals main pulmonary artery net flowMain pulmonary plane error, often in a dilated artery
Branch pulmonary artery flow versus pulmonary vein flow on the same sideEqualAnomalous pulmonary venous connection or collateral flow
Superior vena cava plus descending aortic flowEquals ascending aortic flowCollateral flow or plane error
Several planes along the ascending aortaWithin 5% of one another between sinuses and first branchPlane-specific error, vortical flow, or offset
Four-valve net forward flows (4D flow)Equal without a shuntValve-tracking error

Relationships from the 2020 SCMR post-processing recommendations (Schulz-Menger 2020) and the 2023 4D flow consensus (Bissell 2023). In local validation of 4D flow, differences of 5% or less between methods and up to 10% between scan and rescan are acceptable.

A regurgitant fraction or shunt ratio that conflicts with chamber size, cine appearance, or Doppler findings is reviewed at the level of the acquisition and the contours before it is accepted. A mitral regurgitant fraction of 35% in a normal-sized left ventricle and left atrium, or a Qp:Qs of 1.6 with a normal right ventricle, is more often a measurement error than a new diagnosis.

Advanced 4D flow measures: wall shear stress, kinetic energy, vorticity, and helicity

Beyond volumes and velocities, 4D flow data support derived parameters that the 2015 and 2023 consensus statements classify as research tools not yet validated for clinical decisions: wall shear stress, pressure difference mapping, kinetic energy, turbulent kinetic energy and energy loss, ventricular flow components, flow eccentricity and displacement, vortices, helicity, and pulse wave velocity (Dyverfeldt 2015; Bissell 2023).

Wall shear stress is the tangential viscous force of flowing blood on the vessel wall, calculated from the velocity gradient near the wall. It depends strongly on spatial resolution, because a 2- to 3-mm voxel cannot resolve the thin boundary layer, so absolute values underestimate the true shear and are comparable only within one method. Its clinical interest centers on the bicuspid aortic valve. In a 4D flow study of 60 subjects that included 15 patients with bicuspid valves, right-left cusp fusion produced an eccentric systolic jet striking the right anterior wall of the ascending aorta. Wall shear stress there was elevated compared with controls with tricuspid valves matched for age and aortic size (peak 0.9 versus 0.4 N/m²). The elevation was independent of stenosis severity. Right-noncoronary fusion produced a different jet pattern (Barker 2012). In 20 patients undergoing ascending aortic resection, wall regions with elevated shear stress on preoperative 4D flow showed more elastin degradation and more matrix metalloproteinase and transforming growth factor β-1 expression than adjacent regions with normal shear stress from the same patient (Guzzardi 2015). The findings implicate valve-related hemodynamics in bicuspid aortopathy. Whether shear stress maps should change resection extent or surveillance has not been established, and aortic size thresholds continue to govern surgery.

Kinetic energy of blood within a ventricle, calculated as ½mv² summed over voxels, describes how the ventricle moves blood through the cycle, and turbulent kinetic energy, estimated from the intravoxel velocity spread, measures energy lost to turbulence downstream of stenoses and in regurgitant jets. Ventricular flow components separate left ventricular blood into direct flow, retained inflow, delayed ejection, and residual volume by tracing pathlines from end-diastole; a smaller direct-flow fraction marks less efficient ventricular function. Vorticity quantifies local rotation of the fluid, and the vortex ring that forms behind the opening mitral leaflets has been extracted automatically and characterized by its vorticity and kinetic energy, which differed between healthy subjects and patients with ischemic heart disease (Kräuter 2020). Helicity describes the corkscrew component of flow along a vessel; in normal volunteers, helicity was higher in older than in younger adults, rose with aortic tortuosity, and was inversely related to turbulent kinetic energy (Dyverfeldt 2023). These parameters are measured in research cohorts with heart failure, cardiomyopathy, bicuspid valves, and aortic aneurysm; none has a validated clinical threshold.

4D flow in adult congenital heart disease

Complex congenital anatomy is where 4D flow has been adopted most widely, because a single acquisition covers vessels whose orientation is hard to predict and whose planes would require separate 2D scans. The 2015 consensus lists clinical uses in shunts and collateral vessels, complex congenital disease including single-ventricle and Fontan circulations and repaired tetralogy of Fallot, and aortic disease including aneurysm, coarctation, and dissection (Dyverfeldt 2015).

In repaired tetralogy of Fallot, 4D flow measures pulmonary regurgitation, branch pulmonary artery flow, and tricuspid regurgitation by valve tracking from one dataset, and it shows flow in a dilated right ventricular outflow tract that is difficult to measure with a fixed plane. In the Fontan circulation, flow in the superior and inferior venae cavae, the Fontan conduit, and the branch pulmonary arteries shows how caval return is distributed to the two lungs, and aortopulmonary collateral flow is estimated as the difference between total pulmonary venous return and branch pulmonary artery flow, or between aortic flow and caval return. In transposition after an atrial switch, baffle obstruction and leaks appear as accelerated or abnormally directed flow. In dissection, false lumen flow and entry and re-entry tears are characterized with a low VENC; dissection is covered on the acute aortic syndromes page. Anomalous pulmonary venous connections are identified by tracing pulmonary vein pathlines and by comparing the branch pulmonary arteries with their pulmonary veins.

Clinical adoption of 4D flow: validation, software, storage, and reporting

The 2023 consensus describes the practical barriers to clinical use and how centers have overcome them. Accuracy depends on the vendor sequence, acquisition parameters, and post-processing software, so a center validates its own implementation before clinical reporting and again after a sequence change, gradient service, major parameter change, or new software. The recommended validation acquires 2D and 4D flow in 10 subjects without shunts, at a minimum in the ascending and descending aorta, main and branch pulmonary arteries, superior vena cava, and pulmonary veins. It compares forward flow and peak velocity between methods, checks conservation of mass within the individual datasets, and tests intraobserver and interobserver agreement by repeated analysis at least 1 week apart. Differences of 5% or less are the target, and scan-rescan differences up to 10% are acceptable (Bissell 2023).

4D flow datasets are large and need storage capacity in the image archive. The data are analyzed interactively on dedicated software, most of which has regulatory clearance for basic flow quantification and offers advanced parameters for research use; standard clinical viewers cannot display them, so centers save reformatted planes and visualizations to the archive. Readers and technologists need training time before results are used clinically. 2D and 4D flow are reported side by side until the team has confidence in 4D flow alone, because all prognostic thresholds were derived with 2D phase contrast. Respiratory compensation, multiple VENCs, and a large field of view lengthen scans and can prevent 4D flow from fitting into a clinical protocol (Bissell 2023).

Arrhythmia, implanted devices, and other patient factors

Atrial fibrillation and ectopy. Segmented acquisitions average data across beats of different length and filling, producing blurred velocity curves and imprecise volumes. Arrhythmia rejection discards beats outside a chosen RR window at the cost of time; real-time phase-contrast sequences acquire beats separately and measure beat-to-beat variation directly; and ratios such as regurgitant fraction and Qp:Qs tolerate irregular rhythm better than absolute volumes. Left ventricular stroke volume from a cine stack and aortic flow from a phase-contrast sequence acquired in different breath-holds and different beats are least comparable in atrial fibrillation, which weakens the indirect mitral method most.

Implanted devices. MR-conditional pacemakers and defibrillators can be scanned under labeled conditions, and legacy systems under structured protocols; the generator and leads produce susceptibility artifacts that can distort velocity data in nearby vessels. Sternal wires, valve prostheses, stents, and septal occluders produce local signal voids and phase errors; planes are placed away from the artifact. The safety framework is covered in the CMR physics and safety reference.

Body habitus and breath-holding. Large patients may need a larger field of view, which moves vessels away from isocenter and increases offset error. Patients who cannot hold their breath are imaged with free-breathing averaged or navigator-gated acquisitions.

Sources of error in CMR flow quantification

Reporting a CMR flow study

A complete flow report gives the VENC and plane position for all measurements, states whether background correction was applied, and lists forward, reverse, and net volumes with peak velocity for the individual vessels. Derived values follow: regurgitant volume and fraction for the individual lesions with the method used (direct or indirect, and which subtraction), Qp:Qs with the vessels used, cardiac output and index, and branch pulmonary artery flow distribution where relevant. The internal consistency checks that were performed are reported with their agreement. The left and right ventricular volumes, ejection fractions, and indexed volumes are reported alongside the flow data because they establish the consequences of the lesion. Rhythm during acquisition and any limitation, such as atrial fibrillation, prosthetic artifact, or a dilated vessel, are stated. Severity is given with the threshold system used: the ACC/AHA staging values, with the CMR outcome thresholds stated where they apply.

References

  1. Kramer CM, Barkhausen J, Bucciarelli-Ducci C, Flamm SD, Kim RJ, Nagel E. Standardized cardiovascular magnetic resonance imaging (CMR) protocols: 2020 update. J Cardiovasc Magn Reson. 2020;22:17. doi:10.1186/s12968-020-00607-1
  2. Schulz-Menger J, Bluemke DA, Bremerich J, et al. Standardized image interpretation and post-processing in cardiovascular magnetic resonance: 2020 update. Society for Cardiovascular Magnetic Resonance (SCMR) Board of Trustees Task Force on Standardized Post-Processing. J Cardiovasc Magn Reson. 2020;22:19. doi:10.1186/s12968-020-00610-6
  3. Zoghbi WA, Adams D, Bonow RO, et al. Recommendations for noninvasive evaluation of native valvular regurgitation: a report from the American Society of Echocardiography developed in collaboration with the Society for Cardiovascular Magnetic Resonance. J Am Soc Echocardiogr. 2017;30:303–371. doi:10.1016/j.echo.2017.01.007
  4. Otto CM, Nishimura RA, Bonow RO, et al. 2020 ACC/AHA guideline for the management of patients with valvular heart disease: a report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice Guidelines. Circulation. 2021;143:e72–e227. doi:10.1161/CIR.0000000000000923
  5. Gurvitz M, Krieger EV, Fuller S, et al. 2025 ACC/AHA/HRS/ISACHD/SCAI guideline for the management of adults with congenital heart disease: a report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice Guidelines. Circulation. 2026;153:e115–e251. doi:10.1161/CIR.0000000000001402
  6. Lancellotti P, Pibarot P, Chambers J, et al. Multi-modality imaging assessment of native valvular regurgitation: an EACVI and ESC council of valvular heart disease position paper. Eur Heart J Cardiovasc Imaging. 2022;23:e171–e232. doi:10.1093/ehjci/jeab253
  7. Dyverfeldt P, Bissell M, Barker AJ, et al. 4D flow cardiovascular magnetic resonance consensus statement. J Cardiovasc Magn Reson. 2015;17:72. doi:10.1186/s12968-015-0174-5
  8. Bissell MM, Raimondi F, Ait Ali L, et al. 4D flow cardiovascular magnetic resonance consensus statement: 2023 update. J Cardiovasc Magn Reson. 2023;25:40. doi:10.1186/s12968-023-00942-z. Corrigendum: J Cardiovasc Magn Reson. 2026;28:101912. doi:10.1016/j.jocmr.2025.101912
  9. Nayak KS, Nielsen JF, Bernstein MA, et al. Cardiovascular magnetic resonance phase contrast imaging. J Cardiovasc Magn Reson. 2015;17:71. doi:10.1186/s12968-015-0172-7
  10. Gatehouse PD, Rolf MP, Graves MJ, et al. Flow measurement by cardiovascular magnetic resonance: a multi-centre multi-vendor study of background phase offset errors that can compromise the accuracy of derived regurgitant or shunt flow measurements. J Cardiovasc Magn Reson. 2010;12:5. doi:10.1186/1532-429X-12-5
  11. Gelfand EV, Hughes S, Hauser TH, et al. Severity of mitral and aortic regurgitation as assessed by cardiovascular magnetic resonance: optimizing correlation with Doppler echocardiography. J Cardiovasc Magn Reson. 2006;8:503–507. doi:10.1080/10976640600604856
  12. Gabriel RS, Renapurkar R, Bolen MA, et al. Comparison of severity of aortic regurgitation by cardiovascular magnetic resonance versus transthoracic echocardiography. Am J Cardiol. 2011;108:1014–1020. doi:10.1016/j.amjcard.2011.05.034
  13. Chatzimavroudis GP, Oshinski JN, Franch RH, Pettigrew RI, Walker PG, Yoganathan AP. Quantification of the aortic regurgitant volume with magnetic resonance phase velocity mapping: a clinical investigation of the importance of imaging slice location. J Heart Valve Dis. 1998;7:94–101. PMID: 9502146.
  14. Myerson SG, d'Arcy J, Mohiaddin R, et al. Aortic regurgitation quantification using cardiovascular magnetic resonance: association with clinical outcome. Circulation. 2012;126:1452–1460. doi:10.1161/CIRCULATIONAHA.111.083600
  15. Bolen MA, Popovic ZB, Rajiah P, et al. Cardiac MR assessment of aortic regurgitation: holodiastolic flow reversal in the descending aorta helps stratify severity. Radiology. 2011;260:98–104. doi:10.1148/radiol.11102064
  16. Myerson SG, d'Arcy J, Christiansen JP, et al. Determination of clinical outcome in mitral regurgitation with cardiovascular magnetic resonance quantification. Circulation. 2016;133:2287–2296. doi:10.1161/CIRCULATIONAHA.115.017888
  17. Uretsky S, Gillam L, Lang R, et al. Discordance between echocardiography and MRI in the assessment of mitral regurgitation severity: a prospective multicenter trial. J Am Coll Cardiol. 2015;65:1078–1088. doi:10.1016/j.jacc.2014.12.047
  18. Penicka M, Vecera J, Mirica DC, et al. Prognostic implications of magnetic resonance-derived quantification in asymptomatic patients with organic mitral regurgitation: comparison with Doppler echocardiography-derived integrative approach. Circulation. 2018;137:1349–1360. doi:10.1161/CIRCULATIONAHA.117.029332
  19. Zhan Y, Debs D, Khan MA, et al. Natural history of functional tricuspid regurgitation quantified by cardiovascular magnetic resonance. J Am Coll Cardiol. 2020;76:1291–1301. doi:10.1016/j.jacc.2020.07.036
  20. Ferreira-Neto AN, Merten C, Beurich HW, et al. Effect of aortic regurgitation by cardiovascular magnetic resonance after transcatheter aortic valve implantation. Am J Cardiol. 2019;124:78–84. doi:10.1016/j.amjcard.2019.04.007
  21. Rooijakkers MJP, El Messaoudi S, Stens NA, et al. Assessment of paravalvular regurgitation after transcatheter aortic valve replacement using 2D multi-velocity encoding and 4D flow cardiac magnetic resonance. Eur Heart J Cardiovasc Imaging. 2024;25:929–936. doi:10.1093/ehjci/jeae035
  22. Hundley WG, Li HF, Lange RA, et al. Assessment of left-to-right intracardiac shunting by velocity-encoded, phase-difference magnetic resonance imaging: a comparison with oximetric and indicator dilution techniques. Circulation. 1995;91:2955–2960. doi:10.1161/01.CIR.91.12.2955
  23. Steffens JC, Bourne MW, Sakuma H, O'Sullivan M, Higgins CB. Quantification of collateral blood flow in coarctation of the aorta by velocity encoded cine magnetic resonance imaging. Circulation. 1994;90:937–943. doi:10.1161/01.CIR.90.2.937
  24. Caruthers SD, Lin SJ, Brown P, et al. Practical value of cardiac magnetic resonance imaging for clinical quantification of aortic valve stenosis: comparison with echocardiography. Circulation. 2003;108:2236–2243. doi:10.1161/01.CIR.0000095268.47282.A1
  25. Sanz J, Kuschnir P, Rius T, et al. Pulmonary arterial hypertension: noninvasive detection with phase-contrast MR imaging. Radiology. 2007;243:70–79. doi:10.1148/radiol.2431060477
  26. Reiter G, Reiter U, Kovacs G, et al. Magnetic resonance-derived 3-dimensional blood flow patterns in the main pulmonary artery as a marker of pulmonary hypertension and a measure of elevated mean pulmonary arterial pressure. Circ Cardiovasc Imaging. 2008;1:23–30. doi:10.1161/CIRCIMAGING.108.780247
  27. Fidock B, Barker N, Balasubramanian N, et al. A systematic review of 4D-flow MRI derived mitral regurgitation quantification methods. Front Cardiovasc Med. 2019;6:103. doi:10.3389/fcvm.2019.00103
  28. Feneis JF, Kyubwa E, Atianzar K, et al. 4D flow MRI quantification of mitral and tricuspid regurgitation: reproducibility and consistency relative to conventional MRI. J Magn Reson Imaging. 2018;48:1147–1158. doi:10.1002/jmri.26040
  29. Ma LE, Markl M, Chow K, et al. Aortic 4D flow MRI in 2 minutes using compressed sensing, respiratory controlled adaptive k-space reordering, and inline reconstruction. Magn Reson Med. 2019;81:3675–3690. doi:10.1002/mrm.27684
  30. Barker AJ, Markl M, Bürk J, et al. Bicuspid aortic valve is associated with altered wall shear stress in the ascending aorta. Circ Cardiovasc Imaging. 2012;5:457–466. doi:10.1161/CIRCIMAGING.112.973370
  31. Guzzardi DG, Barker AJ, van Ooij P, et al. Valve-related hemodynamics mediate human bicuspid aortopathy: insights from wall shear stress mapping. J Am Coll Cardiol. 2015;66:892–900. doi:10.1016/j.jacc.2015.06.1310
  32. Kräuter C, Reiter U, Reiter C, et al. Automated mitral valve vortex ring extraction from 4D-flow MRI. Magn Reson Med. 2020;84:3396–3408. doi:10.1002/mrm.28361
  33. Dyverfeldt P, Trenti C, Ziegler M, Bjarnegård N, Lindenberger M. Helical flow in tortuous aortas and its relationship to turbulence: a whole-aorta 4D flow MRI study. Front Cardiovasc Med. 2023;10:1124604. doi:10.3389/fcvm.2023.1124604
  34. Baumgartner H, De Backer J, Babu-Narayan SV, et al. 2020 ESC guidelines for the management of adult congenital heart disease. Eur Heart J. 2021;42:563–645. doi:10.1093/eurheartj/ehaa554
  35. Oosterhof T, van Straten A, Vliegen HW, et al. Preoperative thresholds for pulmonary valve replacement in patients with corrected tetralogy of Fallot using cardiovascular magnetic resonance. Circulation. 2007;116:545–551. doi:10.1161/CIRCULATIONAHA.106.659664
  36. Bokma JP, Winter MM, Oosterhof T, et al. Preoperative thresholds for mid-to-late haemodynamic and clinical outcomes after pulmonary valve replacement in tetralogy of Fallot. Eur Heart J. 2016;37:829–835. doi:10.1093/eurheartj/ehv550