Library › Devices and mechanical support › Temporary mechanical circulatory support
Temporary mechanical circulatory support: selection, escalation, weaning and complications
Temporary mechanical circulatory support in adults from device design and pressure-volume physiology through selection, access, anticoagulation, complications, escalation, weaning, and bridging to recovery, durable support, or transplantation.
Temporary mechanical circulatory support replaces part or all of the work of a failing ventricle for days to weeks while the cause is treated or a definitive decision is made. The devices differ in what they do: a balloon pump adds about 0.5 L/min and lowers afterload, a microaxial pump unloads the left ventricle and adds about 3.5–6 L/min, and venoarterial extracorporeal membrane oxygenation (VA-ECMO) provides full flow and gas exchange while raising left ventricular afterload. Only one strategy has lowered mortality in a randomized trial: a microaxial pump in selected shock related to ST-elevation myocardial infarction (STEMI) (DanGer Shock). Device complications were more frequent with the pump (24.0% against 6.2%; relative risk 4.74). US guidelines give routine balloon pump use and routine VA-ECMO use in infarct-related shock a Class 3: No Benefit recommendation.
Definition, purposes, and classes of temporary support
Temporary mechanical circulatory support (temporary MCS) is any device that generates or augments blood flow for a limited period, usually hours to a few weeks, in a patient whose native heart cannot sustain organ perfusion or is expected to fail during a procedure. The device does not treat the cause. It buys time for a coronary artery to be opened, an inflamed myocardium to recover, a mechanical lesion to be repaired, a donor heart to be found, or a team and family to learn whether any of these is possible.
Devices are placed with a stated destination. Bridge to recovery applies when native function is expected to return (reperfused infarction, myocarditis, stunning after cardiotomy, takotsubo syndrome). Bridge to decision applies when neurologic status, organ function, or candidacy for advanced therapy is unknown at the time of placement. Bridge to bridge describes escalation from a temporary device to a durable left ventricular assist device (LVAD). Bridge to transplantation applies to a listed candidate. Support used during an elective high-risk procedure, such as complex percutaneous coronary intervention (PCI) or ventricular tachycardia ablation, is called protected or prophylactic support. When none of these destinations is plausible, the device prolongs dying, and the decision to place it is a decision about goals of care.
Devices are classified by mechanism. The intra-aortic balloon pump (IABP) is a counterpulsation device that moves no blood of its own. Transvalvular microaxial pumps (the Impella family) aspirate blood from the left ventricle and eject it into the ascending aorta, or from the inferior vena cava into the pulmonary artery. Extracorporeal centrifugal pumps move blood outside the body: from the left atrium to a femoral artery (TandemHeart), from the right atrium to the pulmonary artery (ProtekDuo), from the venous system to the aorta through an oxygenator (VA-ECMO), or through surgically placed cannulas (CentriMag and similar pumps).
| Class | Devices | Inflow → outflow | Pump type | Typical duration |
|---|---|---|---|---|
| Counterpulsation | IABP | None (displaces aortic blood volume) | Helium-driven balloon | Days; weeks with axillary placement |
| Left-sided transvalvular microaxial | Impella CP, Impella 5.5; Impella ECP (low-profile, investigational in the United States for high-risk PCI) | Left ventricle → ascending aorta | Axial impeller on a catheter | CP up to 4 days in shock (US labeling); 5.5 up to 14 days |
| Right-sided transvalvular microaxial | Impella RP, RP Flex | Inferior vena cava or right atrium → pulmonary artery | Axial impeller on a catheter | Up to 14 days |
| Percutaneous left atrial bypass | TandemHeart | Left atrium (transseptal) → femoral artery | Extracorporeal centrifugal | Days to weeks |
| Percutaneous right atrial bypass | ProtekDuo dual-lumen cannula with centrifugal pump | Right atrium → pulmonary artery (oxygenator optional) | Extracorporeal centrifugal | Days to weeks |
| Venoarterial ECMO | Peripheral (femoral, axillary) or central circuits | Venous system → aorta through a membrane oxygenator | Extracorporeal centrifugal | Days; the Extracorporeal Life Support Organization (ELSO) advises reassessing at 5 to 7 days if weaning fails |
| Surgical temporary VADs | CentriMag and other centrifugal pumps with surgical cannulas | Left atrium or ventricle → aorta; right atrium → pulmonary artery; or both | Extracorporeal magnetically levitated centrifugal | Weeks |
Classes of temporary mechanical circulatory support in adults. Durations follow US labeling and device reviews; actual duration is governed by the clinical course and complications. The Impella 2.5 is no longer marketed. VAD, ventricular assist device.
Use of temporary support and the evidence that governs it
Temporary support is used most often in cardiogenic shock, which is defined, staged, and phenotyped on the cardiogenic shock page. Acute myocardial infarction–related cardiogenic shock (AMI-CS) and heart failure–related cardiogenic shock (HF-CS) differ in the devices used and in the evidence. AMI-CS is abrupt, usually left ventricle–dominant, and recoverable when the artery is opened; HF-CS arises in a ventricle that has been failing for months, tolerates lower output, recovers less often, and more often ends in a durable LVAD or transplantation.
The randomized evidence in AMI-CS is consistent in one direction: support strategies that were applied routinely to unselected patients did not lower mortality. IABP-SHOCK II (balloon pump, 600 patients), ECLS-SHOCK (VA-ECMO, 420 patients), and ECMO-CS (VA-ECMO, 117 patients) were neutral, and the extracorporeal strategies increased bleeding and vascular complications. DanGer Shock is the exception. It enrolled 360 patients with STEMI-related shock selected by lactate, blood pressure, and ejection fraction, and excluded comatose survivors of out-of-hospital arrest and patients with predominant right ventricular failure. An Impella CP lowered 180-day mortality from 58.5% to 45.8% at the cost of a composite device complication rate of 24.0% against 6.2%. An individual patient data meta-analysis of nine trials (1,114 patients) found no significant 6-month mortality reduction from early active support as a whole (hazard ratio 0.87; 95% confidence interval [CI] 0.74 to 1.03). Benefit was confined to STEMI-related shock without risk of hypoxic brain injury. HF-CS has one randomized device trial, Altshock-2, which was neutral for the balloon pump.
Large US observational analyses comparing microaxial pumps with balloon pumps in AMI-CS before DanGer Shock reported higher in-hospital mortality and bleeding with the microaxial pump. Those analyses compared patients who received different devices for unmeasured reasons and cannot separate device effect from severity. Device use outside the DanGer Shock phenotype therefore rests on physiology, center expertise, and individual judgment; benefit outside that phenotype has not been established.
Pressure-volume physiology of the devices
The left ventricular pressure-volume loop explains what the devices can and cannot do. Myocardial oxygen consumption tracks the pressure-volume area, the area of the loop plus the potential energy under the end-systolic pressure-volume relationship. A device that lowers end-diastolic volume and peak pressure reduces oxygen demand; a device that raises afterload increases it.
Balloon pump. Diastolic inflation raises aortic diastolic pressure, which increases the coronary perfusion gradient when coronary autoregulation is exhausted. Presystolic deflation lowers aortic end-diastolic pressure, so the isovolumic contraction phase ends at a lower pressure and the ventricle ejects against a lower afterload. The loop narrows modestly, end-systolic volume falls slightly, and stroke volume rises slightly. The end-systolic pressure-volume relationship does not move: the balloon adds no contractility, and the forward flow it adds, about 0.5 L/min, depends on native ejection and a stable rhythm.
Microaxial pump. The pump empties the left ventricle continuously, in systole and diastole. As pump flow rises, end-diastolic volume and pressure fall, the isovolumic phases disappear because blood leaves the ventricle throughout the cycle, and the loop changes from a rectangle to a triangle shifted to the left. Aortic pressure is maintained by pump flow and is now uncoupled from left ventricular pressure. Pressure-volume area and myocardial oxygen demand fall, wall stress falls, and the gradient between aortic diastolic pressure and left ventricular diastolic pressure, the coronary driving pressure, widens. The pump can deliver only what reaches the left ventricle, so right ventricular output and pulmonary vascular resistance set the ceiling on its flow.
Left atrial bypass (TandemHeart). Draining the left atrium lowers left ventricular preload. End-diastolic volume and stroke volume fall, systemic flow is maintained by the pump, and the loop shifts to the left and becomes narrower. Because the outflow is retrograde into the femoral artery, aortic pressure and afterload rise as pump flow increases, partly offsetting the preload reduction.
Venoarterial ECMO. Drainage of the right atrium reduces right and left ventricular preload, while retrograde arterial return raises aortic pressure. In a ventricle with preserved contractile reserve the net effect is a smaller loop. In a severely impaired ventricle the rise in afterload dominates: stroke volume falls, end-systolic and end-diastolic volumes and pressures rise, the loop shifts to the right, and oxygen demand increases. When the ventricle cannot generate enough pressure to open the aortic valve, it fills from bronchial, Thebesian, and transpulmonary return, and left atrial pressure climbs. This is the mechanism of left ventricular distension on ECMO and the reason for combined VA-ECMO and Impella support (ECPELLA).
Right-sided support. A right atrium–to–pulmonary artery pump decompresses the right ventricle and delivers flow to the pulmonary circulation, which raises left ventricular preload. Left ventricular filling pressure can rise, and pulmonary edema can follow, when the left ventricle cannot handle the added volume.
| Variable | IABP | Microaxial pump | TandemHeart | VA-ECMO | Right-sided pump |
|---|---|---|---|---|---|
| Left ventricular preload (end-diastolic pressure) | Slight fall | Falls | Falls | Variable; rises if the ventricle does not eject | Rises |
| Left ventricular afterload | Falls | Uncoupled from aortic pressure | Rises with flow | Rises | No direct effect |
| Mean arterial pressure | Slight rise; diastolic augmentation | Rises | Rises | Rises | Rises if the left ventricle can use the added preload |
| Myocardial oxygen demand | Falls | Falls | Falls | Can rise | Right ventricular demand falls |
| Coronary perfusion pressure | Rises (diastolic) | Rises | Little change | Variable | Right coronary perfusion improves with systemic pressure |
| Pulmonary capillary wedge pressure | Slight fall | Falls | Falls | Variable; can rise | Can rise |
| Right atrial pressure | No direct effect | No direct effect; can rise if right ventricle fails | No direct effect | Falls | Falls |
Hemodynamic effects of temporary devices on the left ventricle and circulation, summarized from pressure-volume analyses of mechanical unloading (Uriel 2018) and the 2015 SCAI/ACC/HFSA/STS consensus. Effects depend on pump speed or flow, native contractility, volume status, and vascular resistance.
Native and device flows add, and the cardiac output that a pulmonary artery catheter reports depends on the device. With a microaxial pump, thermodilution across the right heart measures total systemic flow, because all blood passes through the lungs and the pump. On VA-ECMO, blood drained to the circuit bypasses the pulmonary artery, so thermodilution measures only native right ventricular output, and total systemic flow is the sum of circuit flow and native output. Cardiac power output (mean arterial pressure × cardiac output / 451) calculated on support therefore combines native and device contributions; native cardiac power is assessed with support turned down. Cardiac power output and the pulmonary artery pulsatility index (PAPi) are calculated with the cardiogenic shock calculator. Their derivation and limitations are described on the right heart catheterization page.
Indications, candidacy, and contraindications
Temporary support is indicated when tissue perfusion cannot be maintained by treatment of the cause, correction of preload, and a reasonable dose of vasoactive drugs, or when escalating drugs are producing unacceptable arrhythmia or myocardial oxygen demand. The decision integrates the Society for Cardiovascular Angiography and Interventions (SCAI) SHOCK stage and its trajectory, the failing chamber (left, right, or both), filling pressures and cardiac output, lactate and end-organ function, the respiratory status, vascular anatomy, bleeding risk, neurologic prognosis, and the exit strategy. No single threshold of cardiac power, cardiac index, lactate, PAPi, or vasoactive dose mandates a device. Staging, the initial vasoactive regimen, and revascularization are covered on the cardiogenic shock, vasopressors and inotropes, and STEMI pages.
Common indications outside AMI-CS are acute decompensation of chronic heart failure with hypoperfusion, fulminant myocarditis, postcardiotomy shock, mechanical complications of infarction awaiting repair, refractory ventricular arrhythmia or electrical storm, right ventricular failure after LVAD implantation, transplantation, or pulmonary embolism, cardiotoxic drug poisoning, septic cardiomyopathy in selected patients, and protected high-risk PCI. Refractory cardiac arrest treated with extracorporeal cardiopulmonary resuscitation (ECPR) is a separate pathway.
| Device | Absolute or strong contraindications | Relative contraindications |
|---|---|---|
| IABP | Moderate to severe aortic regurgitation; aortic dissection; abdominal or thoracic aortic aneurysm at risk of rupture | Severe iliofemoral disease (axillary route); uncontrolled bleeding or contraindication to anticoagulation; tachyarrhythmia or highly irregular rhythm that defeats timing |
| Left-sided microaxial pump | Left ventricular thrombus; mechanical aortic valve; moderate to severe aortic regurgitation; aortic dissection; post-infarct ventricular septal rupture | Severe aortic stenosis (weighed case by case); obstructive hypertrophic cardiomyopathy; ascending aortic aneurysm; significant peripheral artery disease (alternative access); severe right ventricular failure without right-sided support; disorders predisposing to hemolysis |
| TandemHeart | Ventricular septal rupture (right-to-left shunting with left atrial drainage); left atrial thrombus | Severe peripheral artery disease; aortic regurgitation; contraindication to anticoagulation |
| VA-ECMO | No plausible exit (no recovery and no candidacy for durable LVAD or transplantation); severe aortic regurgitation; aortic dissection with extensive branch involvement; prolonged anoxic brain injury | Severe aortoiliac disease (axillary or central route); Child-Pugh class B or C cirrhosis; malignancy with poor life expectancy; severe coagulopathy |
| Right-sided microaxial pump or ProtekDuo | Mechanical tricuspid or pulmonary valve; right atrial, right ventricular, or caval thrombus | Pulmonary valve stenosis or severe regurgitation; vena cava filter or venous anatomy that prevents passage |
Contraindications from the 2015 SCAI/ACC/HFSA/STS consensus, the 2021 ELSO venoarterial guideline, device reviews, and labeling. Most contraindications are relative and are weighed against death without support. Device-independent exclusions, such as terminal illness or refusal, apply to all devices.
Shock teams, regional hub-and-spoke systems, and the Level 1 to 3 center designations of the 2025 ACC guidance determine where support is offered. Placement at a center that cannot manage complications, escalate, or offer durable LVAD and transplantation commits the patient to transfer on a device; the advantages and limits of team-based care are summarized on the cardiogenic shock page.
Guideline recommendations for temporary support
| Source | Recommendation | Class (level) |
|---|---|---|
| 2025 ACC/AHA/ACEP/NAEMSP/SCAI ACS guideline | Microaxial flow pump in selected patients with STEMI and severe or refractory cardiogenic shock to reduce death | 2a |
| 2025 ACC/AHA/ACEP/NAEMSP/SCAI ACS guideline | Routine IABP in AMI-CS | 3: No Benefit |
| 2025 ACC/AHA/ACEP/NAEMSP/SCAI ACS guideline | Routine VA-ECMO in AMI-CS | 3: No Benefit |
| 2022 AHA/ACC/HFSA heart failure guideline | Temporary MCS when end-organ function cannot be maintained by pharmacologic means | 2a |
| 2022 AHA/ACC/HFSA heart failure guideline | Management of cardiogenic shock by a multidisciplinary team experienced in shock | 2a |
| 2022 AHA/ACC/HFSA heart failure guideline | Triage of patients who do not respond rapidly to initial measures to centers that can provide temporary MCS | 2b |
| 2021 ACC/AHA/SCAI revascularization guideline | Elective insertion of an appropriate hemodynamic support device as an adjunct to PCI in selected high-risk patients to prevent hemodynamic compromise | 2b (B-R) |
| 2025 ACC concise clinical guidance on cardiogenic shock | Routine temporary MCS in all patients with shock is strongly discouraged; escalation to a microaxial pump is suggested in selected STEMI-related shock with left ventricle–dominant failure and hypoperfusion or deterioration | Consensus (not graded) |
| 2022 AHA scientific statement on escalation and de-escalation | Structured, phenotype-based escalation and stepwise de-escalation with predefined hemodynamic and perfusion targets | Statement (not graded) |
| 2023 ISHLT/HFSA acute MCS guideline | Definitions of shock, device selection, management, and complications of temporary support, including bridging to durable therapy | Graded recommendations (ISHLT system) |
| 2021 ELSO venoarterial ECMO guideline | Indications, cannulation, distal perfusion, left ventricular venting, monitoring, and weaning of VA-ECMO | Consensus (not graded) |
US guideline and society positions on temporary mechanical circulatory support. Where a guideline grades a recommendation, the ACC/AHA class is shown. The ISHLT/HFSA acute MCS guideline carries a published corrigendum (2023;42:1770).
European guidance differs in emphasis. The 2023 European Society of Cardiology (ESC) acute coronary syndrome guideline, written before DanGer Shock, states that short-term MCS may be considered in acute coronary syndrome with severe or refractory cardiogenic shock (Class IIb) and recommends against routine IABP use (Class III). The 2021 ESC heart failure guideline gives short-term MCS in cardiogenic shock a Class IIa recommendation.
Intra-aortic balloon pump
The IABP is a polyethylene balloon of 25 to 50 mL on a 7 to 8 Fr catheter, positioned in the descending thoracic aorta with its tip about 2 cm below the origin of the left subclavian artery and its lower end above the renal arteries. Balloon volume is chosen by patient height, with 40 mL the most common adult size and smaller or larger balloons for short and tall patients; manufacturers' height bands differ slightly, and torso length matters more than height. Helium is the driving gas because its low density allows rapid inflation and deflation. The device is placed at the bedside or under fluoroscopy, usually through the femoral artery; the axillary or subclavian route allows the patient to sit and walk during weeks of support in HF-CS awaiting transplantation.
Inflation is timed to the dicrotic notch and deflation to the onset of systole, triggered by the electrocardiogram R wave, the arterial pressure waveform, or a pacemaker spike. Correct timing is confirmed on the arterial tracing in 1:2 mode: diastolic augmentation exceeds unassisted systolic pressure, and assisted aortic end-diastolic pressure and the following systolic peak are lower than the unassisted values. Late deflation, with the balloon still inflated when ejection begins, increases afterload and oxygen demand and is the most harmful timing error. Marked tachycardia, atrial fibrillation with a highly irregular rhythm, and low arterial pulse pressure reduce benefit. The waveform analysis and timing errors are illustrated on the cardiogenic shock page.
Evidence for the balloon pump. IABP-SHOCK II randomized 600 patients with AMI-CS and planned early revascularization to a balloon pump or no balloon pump. Thirty-day mortality was 39.7% and 41.3% (relative risk 0.96; P = 0.69), and lactate, catecholamine dose, renal function, and time to hemodynamic stabilization did not differ. Altshock-2 randomized 101 patients with HF-CS in SCAI stages B to D who were candidates for heart replacement therapy to early IABP plus standard care or standard care. It was stopped for futility at a prespecified interim analysis: survival or successful bridge to transplantation or durable LVAD at 60 days occurred in 81% and 75% (hazard ratio 0.72; 95% CI 0.31 to 1.68; P = 0.45), with similar complications. Six patients (13%) assigned to standard care crossed over to a balloon pump.
Selective indications for the balloon pump. The balloon pump retains a selective place where its specific effects match the problem. Examples are acute mitral regurgitation from papillary muscle rupture and ventricular septal rupture awaiting surgery, where afterload reduction lowers the regurgitant fraction or shunt; refractory ischemia awaiting coronary bypass; HF-CS as a bridge to transplantation, especially through axillary access; and left ventricular venting during VA-ECMO when a microaxial pump is not available. It requires only a small arteriotomy, has the lowest vascular complication rate of the devices, and is the most widely available. Complications include limb ischemia, bleeding, thrombocytopenia, balloon rupture with helium embolism, malposition with renal or mesenteric ischemia when the balloon sits too low, and left subclavian occlusion when it sits too high.
Microaxial flow pumps: the Impella family
The Impella is a catheter-mounted axial pump. On left-sided models the cannula crosses the aortic valve, the inlet area sits in the left ventricular cavity, and the outlet ejects into the ascending aorta. A motor drives an impeller at up to tens of thousands of revolutions per minute; performance is set as a level (P-level), and the console estimates flow from motor current and the pressure difference between aorta and left ventricle. Current models use an optical pressure sensor and software (SmartAssist) that displays a placement signal, estimated left ventricular end-diastolic pressure, and suction alerts. A continuous purge of fluid through the motor housing prevents blood from entering the motor.
| Device | Maximum flow | Access and sheath | Principal use | Labeled duration (US) |
|---|---|---|---|---|
| Impella CP | About 3.7 L/min | Femoral (or axillary) artery, 14 Fr | AMI-CS; protected high-risk PCI | Up to 4 days in shock; up to 6 hours in high-risk PCI |
| Impella 5.5 | 5.5 to about 6.2 L/min | Surgical axillary graft (direct aortic access possible) | HF-CS; bridge to durable LVAD or transplantation; postcardiotomy shock | Up to 14 days |
| Impella ECP | Peak above 3.5 L/min | Femoral, 9 Fr insertion, expands after insertion | High-risk PCI; investigational in the United States | Not approved by the US Food and Drug Administration (FDA) |
| Impella RP and RP Flex | About 4 L/min | Femoral vein (RP, 23 Fr); internal jugular vein (RP Flex) | Acute right ventricular failure | Up to 14 days |
Microaxial pumps in current use, from device reviews and manufacturer specifications; peak flows are reached only with adequate preload and right ventricular output. The Impella 2.5 (2.5 L/min, 13 Fr) is no longer marketed.
Placement and position. The femoral CP is advanced over a wire across the aortic valve under fluoroscopy, with the inlet about 3.5 cm below the aortic annulus. The Impella 5.5 is placed through a graft sewn to the axillary artery, with the inlet about 5.0 cm below the annulus. Position is measured on parasternal long-axis or transesophageal long-axis views from the annulus to the middle of the inlet area, and the inlet points toward the apex, free of the mitral apparatus and the walls. Position is rechecked daily and after transport, repositioning, cardiopulmonary resuscitation, or any change in alarms, because the pump migrates. A shallow pump recirculates blood around the aortic valve and can slip into the aorta, a deep pump abuts the wall or entangles the mitral chordae, and both produce suction, hemolysis, and ventricular ectopy. Mosaic flow in the proximal ascending aorta is the expected outlet jet and is distinguished from aortic regurgitation by its location above the valve.
Suction and low flow. A suction alarm indicates that the inlet is collapsing the cavity or contacting tissue. Causes are hypovolemia, right ventricular failure or high pulmonary vascular resistance, malposition, tamponade, and a P-level set above what the ventricle can supply. Management follows the cause: volume, inotropes or right-sided support, repositioning, pericardial drainage, or a lower P-level. Repeated suction is a common cause of hemolysis.
Evidence for the microaxial pump. IMPRESS in Severe Shock randomized 48 mechanically ventilated patients with AMI-CS, most after cardiac arrest, to Impella CP or IABP; 30-day mortality was 46% and 50%. DanGer Shock randomized 360 patients with STEMI-related shock (lactate 2.5 mmol/L or higher, systolic pressure below 100 mm Hg or vasopressor support, and left ventricular ejection fraction below 45%) within 24 hours of shock onset. Of these, 355 were analyzed. Death at 180 days occurred in 45.8% with Impella CP plus standard care and 58.5% with standard care (hazard ratio 0.74; 95% CI 0.55 to 0.99; P = 0.04). The composite of severe bleeding, limb ischemia, hemolysis, device failure, or worsening aortic regurgitation occurred in 24.0% and 6.2% (relative risk 4.74; 95% CI 2.36 to 9.55), and renal replacement therapy in 41.9% and 26.7%. Over follow-up extending to 10 years, the survival difference persisted (hazard ratio 0.70; 95% CI 0.54 to 0.92). Use of a microaxial pump outside the DanGer Shock phenotype, including comatose arrest survivors, right ventricle–dominant shock, mechanical complications, and HF-CS, rests on physiology and observational data.
Impella purge solution: heparin and sodium bicarbonate
The purge is a pressurized infusion that flows through the motor housing and exits at the junction of the motor and the rotor, opposing the entry of blood into the gap where thrombus would stop the pump. The console adjusts purge flow to hold a target purge pressure. The fluid is dextrose in water, which is viscous enough to maintain the barrier, conventionally with unfractionated heparin added. Dextrose concentrations from 5% to 20% have been used; in a US survey of high-volume centers, 52.4% used a heparin concentration of 50 units/mL and 41.3% used lower concentrations.
The purge delivers heparin systemically. At a purge flow of 10 mL/h, a solution of 25 units/mL delivers 250 units of heparin per hour and a solution of 50 units/mL delivers 500 units per hour, and the dose changes whenever the console changes purge flow. Systemic heparin is therefore given as a separate infusion titrated to the combined effect of both sources, and the purge is part of all anticoagulation calculations, bleeding evaluations, and evaluations for heparin-induced thrombocytopenia (HIT).
The FDA has approved a labeling change that allows sodium bicarbonate, 25 or 50 mEq/L in 5% dextrose, as an alternative to heparin in the purge for patients who are intolerant of heparin or in whom heparin is contraindicated. A further approved labeling change allows a bicarbonate purge whenever, in the physician's judgment, heparin should not be used. Bicarbonate neutralizes the acidic pH of dextrose and is thought to reduce protein denaturation and deposition in the narrow gap between motor and rotor. The solution provides no systemic anticoagulation, so systemic anticoagulation is maintained separately, with heparin, or with bivalirudin or argatroban in HIT. Direct thrombin inhibitors were placed in the purge before bicarbonate was available; bicarbonate is the labeled alternative. In a single-center comparison of 83 patients on Impella CP or 5.5 for at least 48 hours, the composite of major and clinically relevant bleeding did not differ between bicarbonate-based and heparin-based purge (45.7% against 56.3%; P = 0.34). Major bleeding on the Impella 5.5 was lower with bicarbonate (4.8% against 27.3%; P = 0.04). Device thrombosis did not differ in that cohort, and larger prospective data are not available. A 2021 multidisciplinary consensus of 42 recommendations on Impella anticoagulation describes a heparinized dextrose purge as the standard and bicarbonate purge as an alternative in selected situations.
Purge alarms. A rising purge pressure suggests kinking, clot in the purge lumen, or a change of purge fluid composition; a falling purge pressure suggests a leak or disconnection. Purge alarms are investigated promptly, because loss of the purge barrier allows blood into the motor and causes pump stoppage within hours.
TandemHeart and percutaneous left atrial drainage
The TandemHeart drains oxygenated blood from the left atrium through a large-bore cannula placed from the femoral vein across the interatrial septum, and returns it with an extracorporeal centrifugal pump to the femoral artery. It can deliver flows of about 3.5 to 5 L/min and unload the left atrium and ventricle. It does not depend on aortic valve anatomy, so it can be used with aortic stenosis or a mechanical aortic valve where a microaxial pump is contraindicated. An oxygenator can be added to the circuit. The transseptal puncture requires interventional expertise, and the cannula can migrate back into the right atrium with patient movement. Migration causes right-to-left shunting of desaturated blood and loss of support, and ventricular septal rupture is a contraindication for the same reason.
Two small randomized trials compared TandemHeart with IABP in AMI-CS. In 41 patients randomized after PCI, cardiac power index rose from 0.22 to 0.37 W/m² with TandemHeart and from 0.22 to 0.28 W/m² with IABP, but severe bleeding (19 against 8 patients) and limb ischemia (7 against 0) were more frequent with TandemHeart, and 30-day mortality was similar (43% against 45%). A multicenter trial of 42 patients (33 randomized) also showed larger improvements in cardiac index, mean arterial pressure, and wedge pressure without a difference in 30-day survival or severe adverse events. The same left atrial drainage principle is used to vent the left ventricle during VA-ECMO (left atrial venoarterial ECMO, or LAVA-ECMO).
Venoarterial ECMO: circuit, cannulation, and evidence
A VA-ECMO circuit consists of a venous drainage cannula, a centrifugal pump, a membrane oxygenator with a gas blender and heat exchanger, and an arterial return cannula. Blood flow is set by pump speed and limited by venous drainage, cannula size, and the resistance of the arterial return. Oxygenation is set by the fraction of delivered oxygen and by blood flow; carbon dioxide removal is set by sweep gas flow. Post-oxygenator blood gases are checked after any setting change, and a post-oxygenator Po₂ near 150 mm Hg avoids gross hyperoxemia. Flow sufficient to replace the cardiac output, commonly 4 to 6 L/min in an adult, supports the circulation independent of both ventricles and of the lungs.
Peripheral cannulation through the common femoral vein and artery is the usual approach and can be performed at the bedside, in the catheterization laboratory, or during cardiopulmonary resuscitation. The 2021 ELSO guideline states that a 15 to 17 Fr arterial cannula usually supplies sufficient flow. Cannulas of 19 or 21 Fr are reserved for situations such as sepsis that require higher flow, because larger cannulas cause more limb ischemia. Venous and arterial cannulas are preferably placed in different limbs. The right femoral vein, which runs straight to the right atrium, is preferred for the venous cannula, with the drainage tip at the junction of the inferior vena cava and right atrium or, with multistage cannulas, in the superior vena cava. Ultrasound guidance is recommended, and position is confirmed by fluoroscopy, echocardiography, or chest radiograph. Axillary or subclavian arterial return avoids leg ischemia, reduces differential hypoxemia because return is closer to the aortic arch, and allows mobilization during long runs. Upper limb hyperperfusion with arm swelling is its characteristic complication. Central cannulation of the right atrium and ascending aorta is used mainly after cardiac surgery and in severe peripheral vascular disease. ELSO considers peripheral cannulation preferable when feasible, including after cardiotomy, because it has been linked to better outcomes.
Evidence for VA-ECMO. ECLS-SHOCK randomized 420 patients with AMI-CS and planned early revascularization to early VA-ECMO or no ECMO. Thirty-day mortality was 47.8% and 49.0% (relative risk 0.98), moderate or severe bleeding 23.4% and 9.6%, and peripheral vascular complications requiring intervention 11.0% and 3.8%. ECMO-CS randomized 117 patients with rapidly deteriorating or severe shock of mixed causes to immediate VA-ECMO or an early conservative strategy that allowed downstream ECMO, which 39% of the conservative group received. Death, resuscitated arrest, or another device at 30 days occurred in 63.8% and 71.2% (hazard ratio 0.72; P = 0.21). An individual patient data meta-analysis of four trials (567 patients) found no reduction in 30-day mortality (odds ratio 0.93) and more major bleeding (odds ratio 2.44) and peripheral ischemic complications (odds ratio 3.53). VA-ECMO therefore remains a rescue therapy for refractory biventricular failure, combined cardiac and respiratory failure, refractory arrest, and other selected situations at experienced centers.
Prognostic scores developed in ECMO registries estimate survival at the time of cannulation. The SAVE score (Survival After Venoarterial ECMO) assigns points for diagnosis, age, weight, pre-ECMO organ failures, duration of intubation, peak inspiratory pressure, pre-ECMO arrest, diastolic and pulse pressure, and bicarbonate; predicted survival ranges from 75% in risk class I to 18% in class V. The ENCOURAGE score was derived in AMI-CS and the REMEMBER score in postcardiotomy shock. These scores describe populations. They do not describe individual patients and do not establish futility.
In the ELSO registry from 2014 to 2018 (14,580 adult venoarterial runs), overall survival to discharge was 44%. The registry is the source of most complication rates quoted for VA-ECMO, summarized in the complication table below.
Left ventricular distension and unloading during VA-ECMO
Retrograde aortic flow raises the pressure that the left ventricle must generate to open the aortic valve. When it cannot, the aortic valve stays closed, the ventricle fills from bronchial and Thebesian venous return and from any blood that crosses the lungs, and left ventricular and left atrial pressures rise. The consequences are pulmonary edema, subendocardial ischemia from high wall stress, mitral regurgitation, and stasis with thrombus in the ventricle and aortic root. Arterial pressure can look adequate while this develops.
ELSO lists pulmonary edema on chest radiograph, an arterial pulse pressure below 5 to 10 mm Hg, spontaneous echocardiographic contrast ("smoke") in the left ventricle, and a closed aortic valve with a dilating ventricle as triggers for prompt intervention. A pulmonary artery catheter detects a rising pulmonary artery diastolic or wedge pressure before radiographic edema. Echocardiography documents left ventricular size, aortic valve opening on M-mode, the outflow tract velocity-time integral, mitral regurgitation, stasis, and thrombus.
Noninvasive measures come first: reduction of ECMO flow to the lowest level that maintains organ perfusion, a lower mean arterial pressure target with vasodilators where pressure permits, a moderate dose of inotrope to restore ejection, diuresis or ultrafiltration, and higher positive end-expiratory pressure to reduce pulmonary venous return. When distension persists, the ventricle is vented mechanically:
| Method | How it unloads | Advantages | Limitations |
|---|---|---|---|
| Microaxial pump (ECPELLA, ECMELLA) | Pumps blood from the left ventricle to the aorta | Direct left ventricular unloading; added forward flow; later transition to isolated microaxial support | Second large-bore arterial access; more bleeding, hemolysis, and limb ischemia; worse differential hypoxemia if the lungs fail |
| IABP | Lowers aortic end-diastolic pressure and afterload | Small access; widely available; restores pulsatility | Limited unloading; depends on some native ejection |
| Atrial septostomy (balloon or blade) | Creates a left-to-right atrial shunt drained by the venous cannula | Percutaneous; no second device | Shunt may need closure after recovery; does not empty a ventricle that does not eject |
| Transseptal left atrial cannula (LAVA-ECMO) | Drains the left atrium into the venous limb of the circuit | Effective decompression; one circuit | Transseptal expertise; residual septal defect |
| Surgical vent (left ventricular apex, left atrium, pulmonary vein) | Drains the vented chamber into the circuit | Complete decompression in central configurations | Requires surgery; bleeding |
| Pulmonary artery vent | Drains the pulmonary artery, reducing left-sided return | Percutaneous | Indirect unloading |
Methods of left ventricular decompression during VA-ECMO, from the 2021 ELSO venoarterial guideline and device reviews. Choice depends on anatomy, the need for oxygenation, anticipated duration, and center expertise.
The evidence for routine unloading is mixed. In an international propensity-matched cohort of 510 patients, unloading with a microaxial pump during VA-ECMO was associated with lower 30-day mortality (hazard ratio 0.79) and with more severe bleeding (38.4% against 17.9%), access-site ischemia, hemolysis, and renal replacement therapy. EARLY-UNLOAD, a single-center trial of 116 patients, randomized patients on VA-ECMO to transseptal left atrial cannulation within 12 hours or to a conventional approach that permitted rescue cannulation for a clear indication. Thirty-day mortality was 46.6% and 44.8% (hazard ratio 1.02; 95% CI 0.59 to 1.74), 50% of the conventional group crossed over to rescue venting at a median of 21.8 hours, and pulmonary congestion resolved sooner in the early group. These results support venting when distension is present and do not support routine prophylactic venting of all patients.
Differential hypoxemia (Harlequin, north-south syndrome)
With femoral arterial return, oxygenated ECMO blood travels retrograde up the aorta and meets blood ejected antegrade by the native left ventricle. The meeting point is the mixing zone. When native ejection is minimal, the mixing zone lies in the ascending aorta and ECMO blood perfuses the coronary arteries and the arch vessels. As the ventricle recovers, the mixing zone moves distally through the arch into the descending aorta. If the native lungs are failing at the same time, as with pulmonary edema, pneumonia, or acute respiratory distress syndrome, the recovering ventricle ejects poorly oxygenated pulmonary venous blood into the coronary arteries and the vessels of the head and arms. The lower body receives fully oxygenated ECMO blood. The coronary arteries, the first branches of the aorta, are the first to receive desaturated blood, and the brain follows.
Detection. Arterial blood gases and pulse oximetry are taken from the right radial artery and right hand, because the innominate artery is the first arch branch and reflects the blood reaching the right carotid and, closely, the coronary arteries. ELSO recommends right radial arterial gases and right upper-body oximetry routinely, with cerebral near-infrared spectroscopy (NIRS); a left radial or femoral line can show normal oxygenation while the brain and heart are hypoxemic. A gradient between right radial and femoral saturation, or between right and left cerebral NIRS, establishes the diagnosis. Cyanosis of the head and upper torso with pink lower limbs is a late sign.
Management. The first steps improve native lung function: higher fraction of inspired oxygen and positive end-expiratory pressure, recruitment, diuresis, and treatment of pneumonia. Increasing ECMO flow pushes the mixing zone proximally, at the cost of more left ventricular afterload. When the problem persists, the circuit is reconfigured: a second return cannula into the internal jugular vein converts the circuit to venoarterial-venous (VAV) ECMO, which delivers oxygenated blood to the right heart and lungs so that native ejection carries oxygenated blood; or arterial return is moved to the axillary artery or the ascending aorta. When cardiac function has recovered and the lungs remain the problem, conversion to venovenous ECMO is the definitive change. Adding a microaxial pump for left ventricular unloading increases antegrade flow of poorly oxygenated blood and can worsen differential hypoxemia when the lungs are failing.
Extracorporeal cardiopulmonary resuscitation
ECPR is VA-ECMO initiated during cardiac arrest refractory to conventional advanced life support. Its outcome depends on the interval from collapse to extracorporeal flow, case selection, and a system able to cannulate during mechanical compressions. ARREST (30 patients with refractory ventricular fibrillation) reported survival to discharge of 43% with ECMO-facilitated resuscitation against 7% with standard care. Prague OHCA (256 patients) found survival with good neurologic outcome at 180 days of 31.5% and 22.0% (P = 0.09), and INCEPTION (134 patients) found 30-day favorable neurologic survival of 20% and 16%. The AHA classes, selection criteria, and trial details are covered on the adult cardiac arrest and advanced life support page, and care after return of circulation on the post–cardiac arrest care page. Once flow is restored, ECPR patients face the same device problems as other VA-ECMO patients, with a higher rate of left ventricular distension, differential hypoxemia, and severe brain injury.
Right ventricular and biventricular support
Acute right ventricular failure occurs in right ventricular infarction, acute pulmonary embolism, after LVAD implantation, after heart transplantation, after cardiotomy, in decompensated pulmonary arterial hypertension, and as part of biventricular failure in myocarditis or advanced cardiomyopathy. A failing right ventricle limits all left-sided devices, because no left-sided pump can deliver more blood than crosses the lungs. Suction alarms on a microaxial pump, a rising right atrial pressure with a low PAPi, and a small, underfilled left ventricle on echocardiography indicate that the right ventricle has become the limiting chamber.
Right-sided support is considered when right ventricular output remains inadequate after restoration of coronary flow, correction of preload, rhythm, and atrioventricular synchrony, reduction of pulmonary vascular resistance, and inotropic therapy. The patient must also have a plausible path to recovery or definitive therapy. In acute inferior infarction, a PAPi below 0.9 identified severe right ventricular dysfunction; before LVAD implantation, a lower PAPi was independently associated with the need for a right ventricular assist device. PAPi supports the decision and does not by itself establish the need for a device.
- Impella RP and RP Flex are microaxial pumps that draw blood from the inferior vena cava or right atrium and deliver it to the pulmonary artery at about 4 L/min. In RECOVER RIGHT, 30 patients with refractory right ventricular failure after LVAD implantation, cardiotomy, or infarction received the Impella RP; cardiac index rose from 1.8 to 3.3 L/min/m², central venous pressure fell from 19.2 to 12.6 mm Hg, and 30-day survival was 73.3%.
- ProtekDuo is a dual-lumen cannula placed through the right internal jugular vein with its tip in the main pulmonary artery; it drains the right atrium and returns blood to the pulmonary artery through an extracorporeal centrifugal pump at about 4 to 5 L/min. An oxygenator can be spliced in (an oxygenated right ventricular assist device, OxyRVAD), which supports a right ventricle and lungs that fail together. Jugular access allows the patient to sit and walk. In a series of 60 patients supported for a mean of 14.9 days, central venous pressure fell from 17.8 to 11.8 mm Hg and lactate and creatinine improved.
- Surgical right ventricular assist devices use a centrifugal pump with cannulas from the right atrium to the pulmonary artery, usually at the time of LVAD implantation or cardiotomy.
- VA-ECMO supports the right ventricle and lungs together and is the usual choice in massive pulmonary embolism with refractory shock or arrest, as described on the pulmonary embolism page.
Biventricular failure is supported with VA-ECMO, with VA-ECMO plus a left ventricular vent, with left and right microaxial pumps together (often called BiPella), with a left microaxial pump plus ProtekDuo, or with surgical biventricular centrifugal pumps. The choice depends on whether oxygenation is needed, the anticipated duration, the vascular access available, and whether mobilization will be needed while awaiting recovery, durable support, or transplantation. Right-sided flow that exceeds what the left ventricle or left-sided pump can clear produces pulmonary edema. Left-sided flow that exceeds what the right ventricle or right-sided pump can deliver causes left ventricular suction and low flow. Both pumps are titrated together against the wedge and right atrial pressures.
Surgical temporary ventricular assist devices
Extracorporeal centrifugal pumps with surgically placed cannulas, of which the magnetically levitated CentriMag is the most widely used, provide full left, right, or biventricular support for weeks. The left-sided configuration drains the left atrium or ventricular apex and returns to the ascending aorta, and the right-sided configuration drains the right atrium and returns to the pulmonary artery. The pump has no bearings in contact with blood, which limits hemolysis and thrombosis during prolonged runs, and an oxygenator can be added. These devices are used in postcardiotomy shock, in fulminant myocarditis where prolonged support may be needed, in primary graft dysfunction after transplantation, and as a bridge to decision or to durable LVAD when percutaneous devices are inadequate or have produced complications. Their cost is sternotomy or thoracotomy, with surgical bleeding and infection. ELSO describes using the centrifugal ECMO system as a temporary left, right, or biventricular assist device, with the oxygenator removed once lung function recovers. Candidates include patients with myocarditis in whom recovery is possible but prolonged support may be required.
Choosing a device: support level, phenotype, and duration
Device choice follows from four questions: which ventricle has failed, how much flow is needed, whether gas exchange has failed, and how long support will be needed.
| Clinical situation | Usual first device | Escalation or alternative |
|---|---|---|
| STEMI-related shock, left ventricle–dominant, without post-arrest coma, resembling DanGer Shock | Impella CP, placed early around revascularization | Impella 5.5; addition of right-sided support; VA-ECMO if biventricular failure or hypoxemia |
| HF-CS, left ventricle–dominant, expected need for days to weeks | Axillary Impella 5.5 or axillary IABP | Surgical VAD; durable LVAD; transplantation |
| Right ventricle–dominant shock (right ventricular infarction, post-LVAD, post-transplant) | Impella RP or ProtekDuo | OxyRVAD if hypoxemic; VA-ECMO; surgical RVAD |
| Biventricular failure, refractory arrest, or cardiac plus respiratory failure | VA-ECMO | Venting device (microaxial pump, septostomy, left atrial cannula); VAV configuration for differential hypoxemia |
| Massive pulmonary embolism with refractory shock or arrest | VA-ECMO | Surgical or catheter embolectomy on support |
| Papillary muscle or ventricular septal rupture awaiting repair | IABP | VA-ECMO; a microaxial pump for papillary muscle rupture only (contraindicated in septal rupture because of right-to-left shunting) |
| Fulminant myocarditis with refractory shock or arrhythmia | VA-ECMO or microaxial pump by phenotype | Impella 5.5; surgical biventricular support; transplantation |
| Dynamic left ventricular outflow obstruction (takotsubo, hypertrophic cardiomyopathy) | VA-ECMO when drugs fail | Left-sided microaxial pump and IABP avoided because both can worsen obstruction |
| Elective high-risk PCI with severe left ventricular dysfunction | Impella CP (or IABP) | Standby VA-ECMO |
Common pathways of device selection, assembled from the 2025 ACC guidance, the 2022 AHA statement on escalation and de-escalation, the 2023 ISHLT/HFSA guideline, and ELSO guidance. Center expertise, access, and contraindications modify these pathways. HF-CS, heart failure–related cardiogenic shock; RVAD, right ventricular assist device.
| Feature | IABP | Impella CP | Impella 5.5 | TandemHeart | VA-ECMO |
|---|---|---|---|---|---|
| Flow added | About 0.5 L/min | About 3.7 L/min | 5.5–6.2 L/min | 3.5–5 L/min | 4–6 L/min or more |
| Arterial access | 7–8 Fr | 14 Fr | Surgical axillary graft | Femoral arterial cannula plus large-bore venous transseptal cannula | 15–17 Fr arterial plus large venous cannula |
| Insertion time and setting | Minutes, bedside or fluoroscopy | Minutes, fluoroscopy | Surgical, operating room or hybrid suite | Longer, fluoroscopy and transseptal puncture | Minutes, bedside, fluoroscopy, or during cardiopulmonary resuscitation |
| Mobilization | Possible with axillary access | Limited (femoral) | Yes | Limited | Limited; possible with axillary or jugular configurations |
| Requires stable rhythm | Yes | No, but needs right ventricular output | No, but needs right ventricular output | No, but needs right ventricular output | No |
| Gas exchange | No | No | No | Only with oxygenator | Yes |
Practical comparison of left-sided and full support devices. Flows are from device reviews and the 2021 ELSO guideline. Trial evidence and guideline classes by device are compared on the cardiogenic shock page.
Temporary support selection guide
The guide matches the failing ventricle, the need for oxygenation, the clinical setting, and selected contraindications to the device pathways, support levels, exclusions, and cautions on this page.
Pathways, support levels, and exclusions come from the tables on this page. Cautions come from the sections on unloading, differential hypoxemia, guidelines, and complications.
Large-bore vascular access and limb protection
Large-bore arterial access causes most of the bleeding and limb ischemia on support. Access planning begins before the emergency when possible, with computed tomography or angiography of the iliofemoral arteries in elective or semielective cases and ultrasound in all cases. The femoral artery is entered under ultrasound and fluoroscopic guidance with a single anterior wall puncture over the femoral head, below the inferior epigastric artery and above the bifurcation. A high puncture risks retroperitoneal bleeding, and a low puncture into the superficial femoral artery risks occlusion and pseudoaneurysm. Suture-based closure devices deployed before upsizing ("preclose") allow percutaneous removal of 14 Fr and larger sheaths. Large-bore sheaths are removed in the catheterization laboratory or operating room, where balloon tamponade from the contralateral side and covered stents are available.
When the iliofemoral arteries are small, calcified, or tortuous, alternatives are the axillary artery (percutaneous or through a surgical graft), the transcaval route from the femoral vein into the abdominal aorta, and, for surgical devices, the ascending aorta. Axillary access is preferred for prolonged support because it allows the patient to sit and walk and avoids groin infection.
Limb ischemia. In the ELSO registry, limb ischemia occurred in 5.3% of venoarterial runs and amputation in 0.7%. An ipsilateral distal perfusion cannula, a short 6 to 8 Fr sheath placed antegrade into the superficial femoral artery under ultrasound and connected to the side port of the arterial return cannula, lowered the incidence of leg ischemia from 25.4% to 9.7% in the data cited by ELSO. ELSO recommends distal perfusion at the time of femoral arterial cannulation. With 15 and 17 Fr cannulas, NIRS can guide the decision, with tissue saturation above 50% (preferably 60%), less than 20% difference between legs, and measured distal perfusion flow of at least 100 mL/min where possible. Retrograde perfusion through the posterior tibial or dorsalis pedis artery is an alternative. Microaxial pumps occupy the femoral artery with a 14 Fr sheath and can also produce limb ischemia; options are a distal perfusion catheter or an external bypass from the contralateral femoral artery. Pulse checks and Doppler or NIRS monitoring occur at 1- to 2-hour intervals, and compartment syndrome is treated with fasciotomy.
Monitoring during temporary support
Monitoring on a device combines the hemodynamic and perfusion surveillance used in shock with device-specific checks. The structured reassessment schedule of the 2025 ACC guidance is reproduced on the cardiogenic shock page.
- Invasive hemodynamics. A pulmonary artery catheter measures right atrial, pulmonary artery, and wedge pressures, mixed venous saturation, and cardiac output. ELSO recommends considering one during VA-ECMO to detect rising left-sided pressures, recognizing that thermodilution on ECMO measures only native right ventricular output. Arterial pulse pressure reports native ejection during continuous-flow support.
- Device parameters. Microaxial pumps: P-level, estimated flow, motor current, placement signal, purge flow and pressure, suction alerts. VA-ECMO: pump speed and flow, inlet (drainage) pressure, pre- and post-oxygenator pressures with the transmembrane gradient, sweep and fraction of delivered oxygen, and post-oxygenator gases. IABP: timing, augmentation, and helium alarms. A drop in ECMO flow with "chattering" or vibration of the venous line indicates inadequate drainage from hypovolemia, tamponade, or cannula malposition.
- Echocardiography. Daily transthoracic or transesophageal studies document device position, ventricular size and function, aortic valve opening, valvular regurgitation, pericardial effusion, and thrombus; position is rechecked after transport or alarms.
- Laboratory surveillance. Lactate, mixed or central venous saturation, creatinine, liver chemistries, hemoglobin, platelets, fibrinogen, anticoagulation tests, and hemolysis markers (plasma free hemoglobin, lactate dehydrogenase, haptoglobin, bilirubin) are measured at least daily and more often in the first 24 hours.
- Neurologic monitoring. Sedation is minimized to allow neurologic examination; ELSO recommends standardized multimodal neuromonitoring and cerebral NIRS, with neuroimaging for new deficits.
Response to support is judged by the trajectory of perfusion: falling lactate, improving urine output and mentation, falling vasoactive dose, and improving cardiac power and filling pressures over the first 6 to 24 hours. Failure to improve prompts a search for inadequate support, an unrecognized right or left ventricular problem, a mechanical complication, tamponade, bleeding, sepsis, or malposition.
Anticoagulation on mechanical support
All devices place foreign surfaces in contact with blood, and most require systemic anticoagulation. The targets balance device thrombosis and thromboembolism against the bleeding that dominates complications on support. Evidence for any specific assay or target range is limited, and targets are set by device, bleeding risk, and center protocol.
| Device | Usual agent | Monitoring and targets in common use | Device-specific points |
|---|---|---|---|
| IABP | Unfractionated heparin | aPTT or anti-factor Xa (anti-Xa) activity at a low therapeutic range; some centers omit systemic heparin at 1:1 support in patients at high bleeding risk | Balloon not left stationary when support is paused, to avoid thrombus formation on the balloon |
| Microaxial pump | Unfractionated heparin through the purge plus a titrated systemic infusion | ACT 160–180 s per manufacturer guidance during support; aPTT or anti-Xa used in many centers for maintenance | Purge heparin counted in the total dose; bicarbonate purge with separate systemic anticoagulation when heparin is not to be used |
| TandemHeart | Unfractionated heparin | ACT or aPTT at therapeutic range | Low-flow states increase circuit thrombosis |
| VA-ECMO | Unfractionated heparin; bivalirudin or argatroban | aPTT, anti-Xa, or ACT by center protocol; anti-Xa about 0.3–0.7 IU/mL commonly used for heparin | Reduced or interrupted anticoagulation for major bleeding at adequate circuit flow; low flows during weaning require full anticoagulation |
| Right-sided pump or ProtekDuo | Unfractionated heparin | As for the corresponding left-sided device | Impella RP uses the same purge system |
Anticoagulation on temporary support, from the 2021 ELSO anticoagulation guideline, the 2021 multidisciplinary Impella anticoagulation consensus, and manufacturer guidance. ACT, activated clotting time; aPTT, activated partial thromboplastin time.
Assays. The activated clotting time is a whole-blood point-of-care test used at the bedside and during procedures; it is affected by platelet count, hypothermia, hemodilution, and fibrinogen and is imprecise at the low ranges used for maintenance. The aPTT is affected by factor deficiencies, lupus anticoagulant, and high factor VIII and fibrinogen during inflammation. Anti-Xa activity measures heparin effect directly but can be falsely low with hyperbilirubinemia and hemolysis, and does not reflect antithrombin deficiency, which produces apparent heparin resistance. Discordance between aPTT and anti-Xa is frequent in shock. Viscoelastic tests (thromboelastography, rotational thromboelastometry) add information on platelet function and fibrinogen during bleeding. The properties of these assays are described on the parenteral anticoagulants and HIT page.
Heparin-induced thrombocytopenia. Thrombocytopenia on support has many causes: consumption by the circuit, shear, dilution, sepsis, drugs, and disseminated intravascular coagulation, so most falls in platelet count on a device are not HIT. When HIT is suspected by clinical probability and confirmed by testing, all heparin is stopped, including the Impella purge and heparin-coated catheters where feasible, and a direct thrombin inhibitor is substituted: bivalirudin (cleared mainly by plasma proteolysis with a partial renal component) or argatroban (hepatic clearance). The Impella purge is changed to the bicarbonate solution. ELSO notes that bivalirudin binds circulating and clot-bound thrombin but can undergo local proteolysis and clotting in low-flow areas, such as a nonejecting left ventricle or stagnant parts of a circuit. Diagnosis and treatment of HIT are described on the parenteral anticoagulants and HIT page.
Combined support. With ECPELLA, systemic anticoagulation serves both devices, and the heparin delivered by the purge adds to the circuit infusion, so the two are titrated as one regimen.
Hemolysis on mechanical support
Hemolysis on a device results from shear stress on red cells in high-velocity flow, contact with the impeller, and negative pressure at the inlet. It is more frequent with microaxial pumps than with the balloon pump, and on VA-ECMO it reflects circuit or oxygenator thrombosis, high pump speed with poor drainage (negative inlet pressure), or kinking. With microaxial pumps the common causes are malposition (inlet too deep against the wall or chordae, or too shallow at the valve), suction from hypovolemia or right ventricular failure, high P-levels, and thrombus in the pump; a chorda caught in the inlet has been reported. In a single-center series of Impella patients supported for more than 6 hours, hemoglobin fell significantly within 24 hours and lactate dehydrogenase and bilirubin rose during support.
Markers. Plasma free hemoglobin is the most specific measure. ELSO registry data used plasma free hemoglobin above 50 mg/dL to define hemolysis, which was reported in 3.4% of adult venoarterial runs. Lactate dehydrogenase, bilirubin, haptoglobin, and dark or pink urine are supportive, though all are affected by liver injury, ischemia, and transfusion.
Consequences. Free hemoglobin scavenges nitric oxide, causes pigment nephropathy, and contributes to acute kidney injury, which in turn worsens outcomes on support. Severe hemolysis falsely lowers anti-Xa activity and interferes with other laboratory assays.
Management. Treatment follows the cause: echocardiographic repositioning of a malpositioned microaxial pump; correction of volume status and right ventricular failure to stop suction; a lower P-level or ECMO pump speed when perfusion permits; investigation and exchange of a thrombosed oxygenator, circuit, or pump; and avoidance of nephrotoxins. Hemolysis that persists despite correct position and adequate filling is an indication to change or remove the device.
Bleeding, thrombocytopenia, and acquired von Willebrand syndrome
Bleeding is the most frequent major complication of temporary support. In DanGer Shock the composite device complication rate, driven largely by severe bleeding, was 24.0% with the microaxial pump and 6.2% without it; in ECLS-SHOCK moderate or severe bleeding occurred in 23.4% with VA-ECMO and 9.6% without it. In the ELSO registry, cannulation-site bleeding occurred in 12.5% of venoarterial runs, surgical-site bleeding in 14.5%, cardiac tamponade in 4.4%, and pulmonary hemorrhage in 2.3%.
The causes are additive: large-bore arterial access, systemic anticoagulation, thrombocytopenia from consumption and shear, platelet dysfunction, acquired von Willebrand syndrome from shear-induced cleavage of high-molecular-weight von Willebrand multimers, hepatic dysfunction with reduced factor synthesis, hyperfibrinolysis, and hemodilution. Access-site and retroperitoneal bleeding predominate early; gastrointestinal, pulmonary, and intracranial bleeding emerge with longer runs.
Management combines source control (manual or balloon compression, covered stents, surgical repair, endoscopy) with reduction or interruption of anticoagulation when circuit flow is adequate, correction of fibrinogen and platelets guided by laboratory and viscoelastic testing, and antifibrinolytic therapy in selected patients. ELSO guidance allows heparin to be reduced or stopped for excessive or life-threatening hemorrhage, with close monitoring of the circuit for thrombus. Unnecessary invasive procedures are avoided during support.
Neurologic, limb, kidney, infectious, and device-related complications
| Complication | Frequency | Main mechanisms | Prevention and management |
|---|---|---|---|
| Central nervous system infarction | 3.5% of VA-ECMO runs (ELSO registry) | Circuit or ventricular thrombus, air, aortic root stasis, hypoperfusion | Anticoagulation monitoring; aortic valve opening and venting; daily neurologic examination; neuroimaging for new deficits |
| Intracranial hemorrhage | 1.4% of VA-ECMO runs; survival 11% in affected patients | Anticoagulation, thrombocytopenia, rapid hypocapnia at initiation, hypertension | Lowest effective anticoagulation; avoidance of rapid carbon dioxide reduction at ECMO start; platelet and fibrinogen correction |
| Limb ischemia | 5.3% of VA-ECMO runs; amputation 0.7% | Occlusive arterial cannula or sheath, embolism, dissection | Small arterial cannula; distal perfusion cannula; NIRS and Doppler checks; fasciotomy for compartment syndrome |
| Acute kidney injury requiring renal replacement | 29.6% of VA-ECMO runs; 41.9% with microaxial pump against 26.7% without in DanGer Shock | Shock, hemolysis, contrast, atheroembolism, venous congestion | Hemolysis control; avoidance of nephrotoxins; continuous renal replacement therapy, which can be connected to the ECMO circuit |
| Culture-proven infection | 7.6% of VA-ECMO runs | Prolonged cannulation, open chest, central lines, ventilator-associated pneumonia | Shortest effective support; axillary or jugular access for long runs; standard line and ventilator bundles |
| Circuit and oxygenator thrombosis | 9.2% of VA-ECMO runs (circuit clots) | Low flow, inadequate anticoagulation, prolonged runs | Monitoring of transmembrane pressure, gas transfer, D-dimer, and hemolysis; elective component exchange |
| Aortic valve injury and aortic regurgitation (microaxial pump) | Included in DanGer Shock safety composite | Cannula across the valve; leaflet trauma | Echocardiographic surveillance; shortest effective duration |
| Left ventricular perforation and tamponade | Uncommon | Wire or pump perforation during placement | Fluoroscopic guidance and soft-tipped wires; urgent echocardiography for hypotension after placement |
Complications of temporary support with frequencies from the ELSO registry (2014–2018, 14,580 adult venoarterial runs, as reported in the 2021 ELSO venoarterial guideline) and from DanGer Shock. Rates in individual trials depend on definitions and follow-up.
Neurologic injury on VA-ECMO includes hypoxic-ischemic injury from the arrest or low-flow period that preceded cannulation, ischemic stroke, intracranial hemorrhage, and brain death. ELSO recommends awakening patients as soon as stable support is established to assess neurologic status. It recommends repeating the examination after 1 to 2 days when there is evidence of major brain damage and considering termination of support for futility when there is no neurologic improvement. The 2024 ELSO neuromonitoring guideline recommends standardized multimodal monitoring with clinical examination, NIRS, and neuroimaging, and avoidance of large, rapid falls in carbon dioxide tension at initiation.
Escalation of support by failure pattern
Escalation is considered when perfusion, congestion, or end-organ function fails to improve on the present device after correction of reversible problems. Three patterns account for most failures: undersupport (cardiac index or power remains low, lactate does not clear), inadequate unloading (wedge pressure stays high, pulmonary edema persists, or the ventricle distends on ECMO), and right ventricular failure (rising right atrial pressure, low PAPi, suction on a left-sided pump). The patterns point to different escalations.
| Problem on present support | Findings | Escalation options |
|---|---|---|
| Undersupport on IABP | Persistent hypoperfusion, rising lactate, rising vasoactive dose | Microaxial pump; VA-ECMO if biventricular or hypoxemic |
| Undersupport on Impella CP | Low total cardiac power, rising lactate at maximal P-level with good position and filling | Impella 5.5; VA-ECMO with the CP as vent |
| Right ventricular failure on a left-sided pump | Suction alarms, low pump flow, high right atrial pressure, low PAPi, small left ventricle | Inhaled pulmonary vasodilator and inotrope; Impella RP or ProtekDuo; VA-ECMO |
| Left ventricular distension on VA-ECMO | Closed aortic valve, pulse pressure below 5–10 mm Hg, rising wedge pressure, pulmonary edema, stasis | Flow reduction and inotrope; microaxial pump, IABP, septostomy, or left atrial cannula |
| Differential hypoxemia on femoral VA-ECMO | Low right radial saturation with high femoral saturation | Ventilator optimization; VAV configuration; axillary or central return |
| No recovery after days of support | Continued device dependence after hemodynamic and metabolic abnormalities have been corrected | Surgical VAD, durable LVAD, or transplantation evaluation; goals-of-care discussion |
Escalation of temporary support by failure pattern, adapted from the 2022 AHA statement on escalation and de-escalation and the 2025 ACC guidance.
Escalation adds access sites, anticoagulation exposure, hemolysis, and cost, and added devices bring their own complications. Before escalation the team confirms that the hemodynamic and metabolic state is not already irreversible, that neurologic status is compatible with recovery, and that a destination exists for the more invasive device. When none of these holds, escalation is replaced by a goals-of-care discussion.
Weaning, de-escalation, and decannulation
De-escalation begins when the cause has been treated or has recovered, vasoactive support is minimal, volume status is near euvolemia, and end-organ function is recovering, with falling lactate, creatinine, and transaminases. Before weaning, an advanced heart failure specialist assesses whether recovery, durable LVAD, or transplantation is the realistic goal, because weaning a patient who will need durable therapy wastes the window for it. Weaning steps are followed by reassessment of clinical examination, perfusion markers, echocardiography, and hemodynamics. Weaning in HF-CS is usually slower than in AMI-CS, and some patients are bridged from the device to inotropes.
IABP. The assist ratio is reduced from 1:1 to 1:2 and 1:3, or augmentation is reduced, over hours; stable pressure, cardiac output, filling pressures, and lactate at 1:3 justify removal. Anticoagulation is held before removal of the sheath.
Microaxial pump. The P-level is reduced by one or two steps at 2- to 4-hour intervals, corresponding to about 0.5 to 1 L/min, to a minimal level. Stable mean arterial pressure, cardiac power, wedge and right atrial pressures, PAPi, and lactate at minimal support, with an adequate left ventricular outflow tract velocity-time integral on echocardiography, justify removal. The pump is withdrawn into the aorta at a low P-level before removal so that it does not regurgitate across the valve. Removal of the 14 Fr sheath uses the preclose sutures or surgical repair. The Impella 5.5 is removed in the operating room with ligation or repair of the graft.
VA-ECMO. ELSO describes weaning when recovery is sufficient to maintain a pulse pressure above 10 mm Hg, a mean arterial pressure above 65 mm Hg, and an adequate mixed or central venous saturation at 2 to 2.5 L/min of ECMO flow on minimal vasoactive support. During a weaning trial, flow is reduced in steps of 0.5 L/min at 5- to 10-minute intervals, and the patient is assessed after 3 to 5 minutes with the circuit clamped or at a minimum flow of about 1 L/min, with echocardiography and, where available, a pulmonary artery catheter. At reduced flow, ELSO expects successful weaning with a mean arterial pressure above 60 mm Hg, a left ventricular outflow tract velocity-time integral above 12 cm, a lateral mitral annular peak systolic velocity of 6 cm/s or more, a central venous pressure of 10 mm Hg or less, and a left ventricular ejection fraction of 25% to 30% or more. These thresholds apply on low doses of no more than 1 or 2 inotropes or vasopressors. In the study that derived most of these echocardiographic thresholds (Aissaoui 2011), all patients who weaned successfully had an aortic velocity-time integral of 10 cm or more, an ejection fraction above 20% to 25%, and a lateral mitral annular systolic velocity of 6 cm/s or more at minimal flow. Right ventricular function needs separate testing, because even low ECMO flow reduces right ventricular preload and clamping may not test the right ventricle fully. The pump-controlled retrograde trial off, in which about 1 L/min of flow is reversed through the venous cannula with the sweep off, tests the right heart and lungs together. After a successful trial, decannulation follows promptly, with surgical or percutaneous arterial repair and ultrasound evaluation for venous thrombosis.
| Low-dose vasoactive support during VA-ECMO weaning (ELSO) | Threshold |
|---|---|
| Dopamine | Below 3 µg/kg/min |
| Dobutamine | Below 3 µg/kg/min |
| Milrinone | Below 0.3 µg/kg/min |
| Norepinephrine | Below 0.06 µg/kg/min |
| Epinephrine | Below 0.1 µg/kg/min |
| Phenylephrine | Below 1 µg/kg/min |
| Vasopressin | Below 0.03 units/min |
Definitions of low-dose vasoactive support used with the VA-ECMO weaning criteria of the 2021 ELSO venoarterial guideline. The guideline prints the vasopressin threshold as 0.03 µg/kg/min; vasopressin is dosed in units per minute, and 0.03 units/min is the conventional low dose.
Right-sided support. Flow is reduced stepwise while right atrial pressure, cardiac index, PAPi, and right ventricular size and function are followed; inhaled pulmonary vasodilators and inotropes often accompany the transition.
Failure to wean. ELSO states that patients who cannot be weaned from VA-ECMO within 5 to 7 days may be considered for a temporary LVAD to allow more time for recovery when right ventricular function is preserved. Patients with known severe chronic left ventricular dysfunction may proceed earlier to durable LVAD or transplantation.
Bridge to recovery, durable LVAD, or transplantation
A temporary device starts a pathway, and the next step depends on recovery. Patients who recover native function are weaned and started on guideline-directed therapy for heart failure with reduced ejection fraction as soon as hemodynamics allow. Patients who do not recover are evaluated for durable support and transplantation, with the evaluation started early, often within the first days of support.
The acuity of patients considered for durable LVAD is described by the Interagency Registry for Mechanically Assisted Circulatory Support (INTERMACS) profiles: profile 1 is critical cardiogenic shock ("crash and burn"), and profile 2 is progressive decline despite inotropes. Outcomes of durable LVAD implanted in profile 1 are worse than in more stable profiles, so temporary support is used to stabilize the patient, allow recovery of kidney, liver, and neurologic function, and assess right ventricular function before implantation. Axillary microaxial pumps and balloon pumps allow ambulation and rehabilitation during this period. Recognition of advanced heart failure and referral are described on the heart failure with reduced ejection fraction page.
The current US heart allocation policy replaced a three-tier system with six statuses and assigns priority partly by device. Status 1 applies to patients on VA-ECMO, with a nondischargeable surgical biventricular assist device, or with mechanical support complicated by life-threatening ventricular arrhythmia. Status 2 applies to patients supported by a balloon pump or another percutaneous endovascular device, a nondischargeable surgical LVAD, or a malfunctioning device. Listing on VA-ECMO, a balloon pump, or a percutaneous endovascular device requires documented hemodynamic criteria for cardiogenic shock. The policy increased the number of patients transplanted directly from temporary support and the use of balloon pumps and microaxial pumps as bridges to transplantation.
Protected high-risk percutaneous coronary intervention
Hemodynamic support is used electively during PCI in patients whose coronary anatomy and ventricular function predict collapse during balloon inflation, atherectomy, or a complication: unprotected left main or last-remaining-vessel disease, complex multivessel disease, and severe left ventricular dysfunction, often in patients declined for surgery. The aim is to prevent hemodynamic instability and allow more complete revascularization.
PROTECT II randomized 452 symptomatic patients with complex three-vessel or unprotected left main disease and severely depressed ejection fraction to Impella 2.5 or IABP during nonemergent PCI. The Impella provided better hemodynamic support, but the primary end point, 30-day major adverse events, did not differ (35.1% against 40.1%; P = 0.227 by intention to treat). At 90 days the rates were 40.6% and 49.3% (P = 0.066) by intention to treat, and the difference reached significance only in the per-protocol population. The trial was stopped early for futility. The 2021 ACC/AHA/SCAI revascularization guideline states that elective insertion of an appropriate hemodynamic support device as an adjunct to PCI may be reasonable in selected high-risk patients to prevent hemodynamic compromise (Class 2b, B-R). PROTECT IV (NCT04763200) randomized patients with complex coronary disease and reduced ejection fraction to Impella CP support or PCI with or without a balloon pump, with a 3-year composite primary end point (ClinicalTrials.gov NCT04763200). Large-bore access complications are the main hazard of protected PCI in elective cases, and the device is removed at the end of the procedure when hemodynamics allow.
Other procedural uses include ablation of hemodynamically unstable ventricular tachycardia, high-risk transcatheter aortic valve replacement or balloon aortic valvuloplasty in patients with severe left ventricular dysfunction, and high-risk cardiac surgery, in which a microaxial pump may be placed before or after the operation to prevent or treat postcardiotomy shock. Randomized data for these uses are lacking.
Postcardiotomy shock
Postcardiotomy cardiogenic shock is failure to wean from cardiopulmonary bypass, or shock developing after cardiac surgery, despite inotropes and correction of surgical problems. Temporary support is provided by VA-ECMO (continued from bypass through central cannulas or converted to peripheral cannulation), by surgically placed centrifugal VADs, by the Impella 5.5, or by the IABP. The 2020 EACTS/ELSO/STS/AATS expert consensus sets out indications, timing, configuration, management, weaning, and the decision to stop. It emphasizes early institution before profound multiorgan failure, attention to residual surgical lesions by transesophageal echocardiography and angiography, left ventricular venting, and early recognition of futility.
Outcomes remain poor. In the PELS-1 multicenter cohort of 2,058 adults on postcardiotomy ECMO, in-hospital mortality was 60.5%, and age and preoperative cardiac arrest were independently associated with death. Among hospital survivors, survival was 89.5% at 1 year and 65.9% at 10 years, so about two thirds of hospital survivors were alive at 10 years.
Disease-specific uses of temporary support
Fulminant myocarditis. Fulminant myocarditis presents with shock, often with ventricular arrhythmia or heart block, in a previously healthy adult, and myocardial recovery is common when the patient survives the acute phase. The 2024 ACC expert consensus decision pathway on myocarditis recommends early transfer to a center capable of temporary MCS and endomyocardial biopsy, vasoactive support as needed, and prompt progression to MCS when shock is refractory. The device follows the phenotype: VA-ECMO for biventricular failure, arrhythmic storm, or arrest; a microaxial pump, often the Impella 5.5, for left ventricle–dominant failure; and surgical biventricular support when prolonged support is anticipated. Biopsy identifies giant cell, eosinophilic, and other immune-mediated forms that require immunosuppression. Support is planned for days to weeks, and transplantation is considered when recovery does not occur.
Mechanical complications of myocardial infarction. Papillary muscle rupture, ventricular septal rupture, and free wall rupture require urgent surgical or transcatheter repair. Temporary support stabilizes the patient as a bridge to repair. The balloon pump reduces afterload and the regurgitant or shunt fraction; a microaxial pump unloads the left ventricle but can increase right-to-left flow across a septal rupture when it lowers left ventricular pressure below right ventricular pressure; and VA-ECMO supports both ventricles but can increase left ventricular afterload and regurgitation. These lesions are described on the STEMI page.
Takotsubo syndrome. Shock in takotsubo syndrome results either from severe pump failure or from dynamic left ventricular outflow obstruction with systolic anterior motion of the mitral valve. Obstruction is worsened by inotropes and by afterload reduction, so the balloon pump and a left-sided microaxial pump are avoided in that phenotype, and VA-ECMO is used when vasoconstrictor support, beta blockade, and volume fail. Pump failure without obstruction can be supported by a microaxial pump or VA-ECMO, and recovery usually occurs within days to weeks.
Pulmonary embolism. VA-ECMO is reasonable in refractory cardiogenic shock from pulmonary embolism (Class 2a in the 2026 multisociety AHA/ACC pulmonary embolism guideline), with continued parenteral anticoagulation in the absence of bleeding; the added benefit of thrombolysis, catheter therapy, or surgical embolectomy once ECMO is running is uncertain. Details are on the pulmonary embolism page.
Arrhythmic storm. Refractory ventricular tachycardia or fibrillation with hemodynamic collapse is supported with VA-ECMO, which maintains flow independent of rhythm, as a bridge to ablation, revascularization, or other treatment; management of storm is covered on the electrical storm page.
Poisoning. Refractory shock from cardiotoxic drugs, such as calcium channel blockers, beta blockers, and sodium channel–blocking agents, is a recognized indication for VA-ECMO while the drug is cleared, as described on the cardiovascular toxicology page.
Septic cardiomyopathy and mixed shock. VA-ECMO has been used in selected patients with severe septic cardiomyopathy and low cardiac output; it is not a treatment for vasodilatory septic shock with preserved output, and higher flows and larger arterial cannulas may be required.
Pregnancy and peripartum cardiomyopathy. Microaxial pumps and VA-ECMO are used in peripartum cardiomyopathy and in obstetric cardiovascular collapse, with device, anticoagulation, and timing of delivery planned jointly with obstetric and anesthesia teams.
Futility, goals of care, and palliative care on support
Temporary devices are often placed in minutes, before neurologic status, organ reversibility, or candidacy for advanced therapy is known. The decision to place a device is therefore a time-limited trial. The goals, the destination, and the conditions for withdrawing support are discussed with the patient or surrogate as early as circumstances permit. ELSO recommends early palliative care involvement for all patients on VA-ECMO, to help with expectations, goals of care, and family support.
Support is no longer serving the patient when there is no neurologic recovery after adequate assessment, when multiorgan failure is irreversible, when the patient is not a candidate for durable LVAD or transplantation and native recovery has not occurred, or when complications such as major intracranial hemorrhage have changed the prognosis. Common contraindications to durable LVAD and transplantation listed by ELSO include lack of neurologic recovery, multiorgan failure, advanced age, inadequate social support, and metastatic cancer. Withdrawal of support is planned with the family and the palliative care team, with symptom control and, when appropriate, time for family to be present. Deactivation of a device is withdrawal of a life-sustaining treatment that no longer serves the patient's goals. Outcomes of palliative care involvement in shock are summarized on the cardiogenic shock page.
Sources of error in temporary mechanical support
- A device placed because shock is present, without defining the failing chamber, the flow required, the need for oxygenation, and the exit. The randomized trials of unselected IABP and VA-ECMO use were neutral, and DanGer Shock applies to the population it enrolled.
- Thermodilution cardiac output on VA-ECMO interpreted as total flow. Circuit flow bypasses the pulmonary artery, so thermodilution measures native output only, and total systemic flow is the sum.
- Arterial blood gases drawn from the left radial or a femoral line during femoral VA-ECMO. These sample ECMO-rich blood and miss upper-body hypoxemia; the right radial artery reflects the blood reaching the brain and heart.
- Rising mean arterial pressure on VA-ECMO taken as improvement while pulse pressure narrows, the aortic valve stops opening, and the left ventricle distends.
- Suction alarms on a microaxial pump treated by lowering the P-level alone, when the cause is hypovolemia, malposition, tamponade, or right ventricular failure.
- Purge heparin left out of the anticoagulation plan. The purge delivers a variable, often therapeutic, heparin dose that matters for bleeding, for titration of the systemic infusion, and for heparin-induced thrombocytopenia.
- All platelet falls on a device attributed to HIT, or HIT not considered because thrombocytopenia is expected. Probability scoring and laboratory testing decide; when HIT is confirmed, the purge is changed as well as the systemic infusion.
- A microaxial pump placed in ventricular septal rupture or dynamic outflow obstruction, where unloading the left ventricle can worsen right-to-left shunting or the outflow gradient.
- Femoral arterial cannulation for VA-ECMO without distal perfusion, with limb ischemia recognized late. ELSO recommends distal perfusion at the time of cannulation.
- Weaning VA-ECMO by clamping alone, which does not fully test the right ventricle, or weaning before an advanced heart failure assessment in a patient who will need durable therapy.
- Escalation to a more invasive device after the hemodynamic and metabolic state has become irreversible or before neurologic status is known, without a defined destination.
References
- Rao SV, O'Donoghue ML, Ruel M, et al. 2025 ACC/AHA/ACEP/NAEMSP/SCAI guideline for the management of patients with acute coronary syndromes: a report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice Guidelines. J Am Coll Cardiol. 2025;85:2135–2237. doi:10.1016/j.jacc.2024.11.009
- Sinha SS, Morrow DA, Kapur NK, Kataria R, Roswell RO. 2025 Concise clinical guidance: an ACC expert consensus statement on the evaluation and management of cardiogenic shock. J Am Coll Cardiol. 2025;85:1618–1641. doi:10.1016/j.jacc.2025.02.018
- Heidenreich PA, Bozkurt B, Aguilar D, et al. 2022 AHA/ACC/HFSA guideline for the management of heart failure: a report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice Guidelines. Circulation. 2022;145:e895–e1032. doi:10.1161/CIR.0000000000001063
- Lawton JS, Tamis-Holland JE, Bangalore S, et al. 2021 ACC/AHA/SCAI guideline for coronary artery revascularization: a report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice Guidelines. J Am Coll Cardiol. 2022;79:e21–e129. doi:10.1016/j.jacc.2021.09.006
- Geller BJ, Sinha SS, Kapur NK, et al. Escalating and de-escalating temporary mechanical circulatory support in cardiogenic shock: a scientific statement from the American Heart Association. Circulation. 2022;146:e50–e68. doi:10.1161/CIR.0000000000001076
- Rihal CS, Naidu SS, Givertz MM, et al. 2015 SCAI/ACC/HFSA/STS clinical expert consensus statement on the use of percutaneous mechanical circulatory support devices in cardiovascular care. J Am Coll Cardiol. 2015;65:e7–e26. doi:10.1016/j.jacc.2015.03.036
- Naidu SS, Baran DA, Jentzer JC, et al. SCAI SHOCK stage classification expert consensus update: a review and incorporation of validation studies. J Am Coll Cardiol. 2022;79:933–946. doi:10.1016/j.jacc.2022.01.018
- Bernhardt AM, Copeland H, Deswal A, Gluck J, Givertz MM. The International Society for Heart and Lung Transplantation/Heart Failure Society of America guideline on acute mechanical circulatory support. J Heart Lung Transplant. 2023;42:e1–e64 (corrigendum: 2023;42:1770). doi:10.1016/j.healun.2022.10.028
- Lorusso R, Shekar K, MacLaren G, et al. ELSO interim guidelines for venoarterial extracorporeal membrane oxygenation in adult cardiac patients. ASAIO J. 2021;67:827–844 (erratum: 2022;68:e133). doi:10.1097/MAT.0000000000001510
- McMichael ABV, Ryerson LM, Ratano D, Fan E, Faraoni D, Annich GM. 2021 ELSO adult and pediatric anticoagulation guidelines. ASAIO J. 2022;68:303–310. doi:10.1097/MAT.0000000000001652
- Cho SM, Hwang J, Chiarini G, et al. Neurological monitoring and management for adult extracorporeal membrane oxygenation patients: Extracorporeal Life Support Organization consensus guidelines. ASAIO J. 2024;70:e169–e181. doi:10.1097/MAT.0000000000002312
- Lorusso R, Whitman G, Milojevic M, et al. 2020 EACTS/ELSO/STS/AATS expert consensus on post-cardiotomy extracorporeal life support in adult patients. ASAIO J. 2021;67:e1–e43. doi:10.1097/MAT.0000000000001301
- Estep JD, Nicoara A, Cavalcante J, et al. Recommendations for multimodality imaging of patients with left ventricular assist devices and temporary mechanical support: updated recommendations from the American Society of Echocardiography. J Am Soc Echocardiogr. 2024;37:820–871. doi:10.1016/j.echo.2024.06.005
- Byrne RA, Rossello X, Coughlan JJ, et al. 2023 ESC guidelines for the management of acute coronary syndromes. Eur Heart J. 2023;44:3720–3826. doi:10.1093/eurheartj/ehad191
- Drazner MH, Bozkurt B, Cooper LT, et al. 2024 ACC expert consensus decision pathway on strategies and criteria for the diagnosis and management of myocarditis. J Am Coll Cardiol. 2025;85:391–431. doi:10.1016/j.jacc.2024.10.080
- Uriel N, Sayer G, Annamalai S, Kapur NK, Burkhoff D. Mechanical unloading in heart failure. J Am Coll Cardiol. 2018;72:569–580. doi:10.1016/j.jacc.2018.05.038
- Thiele H, Zeymer U, Neumann FJ, et al. Intraaortic balloon support for myocardial infarction with cardiogenic shock (IABP-SHOCK II). N Engl J Med. 2012;367:1287–1296. doi:10.1056/NEJMoa1208410
- Morici N, Sacco A, Frea S, et al. Early intra-aortic balloon support for heart failure-related cardiogenic shock: a randomized clinical trial (Altshock-2). J Am Coll Cardiol. 2025;85:1587–1597. doi:10.1016/j.jacc.2025.03.003
- Møller JE, Engstrøm T, Jensen LO, et al. Microaxial flow pump or standard care in infarct-related cardiogenic shock (DanGer Shock). N Engl J Med. 2024;390:1382–1393. doi:10.1056/NEJMoa2312572
- Møller JE, Beske RP, Engstrøm T, et al. Long-term outcomes of the DanGer Shock trial. N Engl J Med. 2025;393:1037–1038. doi:10.1056/NEJMc2508284
- Ouweneel DM, Eriksen E, Sjauw KD, et al. Percutaneous mechanical circulatory support versus intra-aortic balloon pump in cardiogenic shock after acute myocardial infarction (IMPRESS in Severe Shock). J Am Coll Cardiol. 2017;69:278–287. doi:10.1016/j.jacc.2016.10.022
- Thiele H, Møller JE, Henriques JPS, et al. Temporary mechanical circulatory support in infarct-related cardiogenic shock: an individual patient data meta-analysis of randomised trials with 6-month follow-up. Lancet. 2024;404:1019–1028. doi:10.1016/S0140-6736(24)01448-X
- Dhruva SS, Ross JS, Mortazavi BJ, et al. Association of use of an intravascular microaxial left ventricular assist device vs intra-aortic balloon pump with in-hospital mortality and major bleeding among patients with acute myocardial infarction complicated by cardiogenic shock. JAMA. 2020;323:734–745. doi:10.1001/jama.2020.0254
- O'Neill WW, Kleiman NS, Moses J, et al. A prospective, randomized clinical trial of hemodynamic support with Impella 2.5 versus intra-aortic balloon pump in patients undergoing high-risk percutaneous coronary intervention: the PROTECT II study. Circulation. 2012;126:1717–1727. doi:10.1161/CIRCULATIONAHA.112.098194
- Thiele H, Sick P, Boudriot E, et al. Randomized comparison of intra-aortic balloon support with a percutaneous left ventricular assist device in patients with revascularized acute myocardial infarction complicated by cardiogenic shock. Eur Heart J. 2005;26:1276–1283. doi:10.1093/eurheartj/ehi161
- Burkhoff D, Cohen H, Brunckhorst C, O'Neill WW; TandemHeart Investigators Group. A randomized multicenter clinical study to evaluate the safety and efficacy of the TandemHeart percutaneous ventricular assist device versus conventional therapy with intraaortic balloon pumping for treatment of cardiogenic shock. Am Heart J. 2006;152:469.e1–469.e8. doi:10.1016/j.ahj.2006.05.031
- Thiele H, Zeymer U, Akin I, et al. Extracorporeal life support in infarct-related cardiogenic shock (ECLS-SHOCK). N Engl J Med. 2023;389:1286–1297. doi:10.1056/NEJMoa2307227
- Ostadal P, Rokyta R, Karasek J, et al. Extracorporeal membrane oxygenation in the therapy of cardiogenic shock: results of the ECMO-CS randomized clinical trial. Circulation. 2023;147:454–464. doi:10.1161/CIRCULATIONAHA.122.062949
- Zeymer U, Freund A, Hochadel M, et al. Venoarterial extracorporeal membrane oxygenation in patients with infarct-related cardiogenic shock: an individual patient data meta-analysis of randomised trials. Lancet. 2023;402:1338–1346. doi:10.1016/S0140-6736(23)01607-0
- Schrage B, Becher PM, Bernhardt A, et al. Left ventricular unloading is associated with lower mortality in patients with cardiogenic shock treated with venoarterial extracorporeal membrane oxygenation: results from an international, multicenter cohort study. Circulation. 2020;142:2095–2106. doi:10.1161/CIRCULATIONAHA.120.048792
- Kim MC, Lim Y, Lee SH, et al. Early left ventricular unloading or conventional approach after venoarterial extracorporeal membrane oxygenation: the EARLY-UNLOAD randomized clinical trial. Circulation. 2023;148:1570–1581. doi:10.1161/CIRCULATIONAHA.123.066179
- Torre DE, Pirri C. Harlequin syndrome in venoarterial ECMO and ECPELLA: when ECMO and native or Impella circulations collide. A comprehensive review. Rev Cardiovasc Med. 2025;26:39992. doi:10.31083/RCM39992
- Anderson MB, Goldstein J, Milano C, et al. Benefits of a novel percutaneous ventricular assist device for right heart failure: the prospective RECOVER RIGHT study of the Impella RP device. J Heart Lung Transplant. 2015;34:1549–1560. doi:10.1016/j.healun.2015.08.018
- Kang G, Ha R, Banerjee D. Pulmonary artery pulsatility index predicts right ventricular failure after left ventricular assist device implantation. J Heart Lung Transplant. 2016;35:67–73. doi:10.1016/j.healun.2015.06.009
- Korabathina R, Heffernan KS, Paruchuri V, et al. The pulmonary artery pulsatility index identifies severe right ventricular dysfunction in acute inferior myocardial infarction. Catheter Cardiovasc Interv. 2012;80:593–600. doi:10.1002/ccd.23309
- Aissaoui N, Luyt CE, Leprince P, et al. Predictors of successful extracorporeal membrane oxygenation (ECMO) weaning after assistance for refractory cardiogenic shock. Intensive Care Med. 2011;37:1738–1745. doi:10.1007/s00134-011-2358-2
- Beavers CJ, DiDomenico RJ, Dunn SP, et al. Optimizing anticoagulation for patients receiving Impella support. Pharmacotherapy. 2021;41:932–942. doi:10.1002/phar.2629
- Reed BN, DiDomenico RJ, Allender JE, et al. Survey of anticoagulation practices with the Impella percutaneous ventricular assist device at high-volume centers. J Interv Cardiol. 2019;2019:3791307. doi:10.1155/2019/3791307
- Patel PJ, Harris JE, Donahue KR, Kim JH, Sigala MI. Evaluation of bleeding rates with sodium bicarbonate versus heparin-based purge solutions for Impella CP and 5.5 ventricular assist devices. Artif Organs. 2026;50:1077–1086. doi:10.1111/aor.70145
- Badiye AP, Hernandez GA, Novoa I, Chaparro SV. Incidence of hemolysis in patients with cardiogenic shock treated with Impella percutaneous left ventricular assist device. ASAIO J. 2016;62:11–14. doi:10.1097/MAT.0000000000000290
- De Filippo O, Pagliassotto I, Braia V, et al. Axial flow percutaneous ventricular assist devices: physiology, procedural considerations, and clinical applications. A state-of-the-art review. Struct Heart. 2026;10:100773. doi:10.1016/j.shj.2025.100773
- Mariani S, Heuts S, van Bussel BCT, et al. Patient and management variables associated with survival after postcardiotomy extracorporeal membrane oxygenation in adults: the PELS-1 multicenter cohort study. J Am Heart Assoc. 2023;12:e029609. doi:10.1161/JAHA.123.029609
- Kremer J, Holler AJ, Schmack B, Ruhparwar A, Karck M, Meyer AL. Percutaneous right ventricular support via ProtekDuo pulmonary artery cannulation. ASAIO J. 2026;72:819–825. doi:10.1097/MAT.0000000000002611
- Creager MA, Barnes GD, Giri J, et al. 2026 AHA/ACC/ACCP/ACEP/CHEST/SCAI/SHM/SIR/SVM/SVN guideline for the evaluation and management of acute pulmonary embolism in adults: a report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice Guidelines. Circulation. 2026;153:e977–e1051. doi:10.1161/CIR.0000000000001415
- Organ Procurement and Transplantation Network. Adult heart allocation: frequently asked questions on the revised policy implemented October 18, 2018. Health Resources and Services Administration; 2018. optn.transplant.hrsa.gov
- US Food and Drug Administration. Premarket approval P140003/S135 (Impella 2.5, CP, CP with SmartAssist, 5.0, 5.5 with SmartAssist, and LD systems): modification of the instructions for use to allow sodium bicarbonate in the purge fluid. Decision date February 6, 2026; earlier approval P140003/S088, April 19, 2022. accessdata.fda.gov
- Organ Procurement and Transplantation Network. OPTN policies. Policy 6: allocation of hearts and heart-lungs. Health Resources and Services Administration. optn.transplant.hrsa.gov
- ClinicalTrials.gov. Impella-supported PCI in high-risk patients with complex coronary artery disease and reduced left ventricular function (PROTECT IV). NCT04763200. clinicaltrials.gov/study/NCT04763200