VA-ECMO circuit management and complications: flow, sweep, cannulation, anticoagulation and circuit failure

Device-level management of adult venoarterial ECMO, from circuit design, flow, sweep gas, cannulation, and limb perfusion through anticoagulation, hemolysis, kidney support, infection, oxygenator failure, and circuit emergencies.

Venoarterial extracorporeal membrane oxygenation (VA-ECMO) is run through three controls: pump speed, which sets blood flow against the patient's preload and afterload; sweep gas flow, which sets carbon dioxide removal; and the fraction of delivered oxygen (FDO₂), which with blood flow sets oxygen delivery. Adult circuits are built for about 60 mL/kg/min, and a drainage saturation above 66–70% indicates adequate oxygen delivery. Post-oxygenator gases are measured after any setting change, with normocapnia and a post-oxygenator Po₂ near 150 mm Hg as the Extracorporeal Life Support Organization (ELSO) target. Most device harm comes from the cannulas, anticoagulation, hemolysis, infection, and thrombus in the oxygenator, and all have measurable early signals.

VA-ECMO topics covered on the temporary support page

This page covers the circuit and its complications. Six related topics are treated in full on the temporary mechanical circulatory support page.

Circuit design: drainage, centrifugal pump, membrane lung, and heat exchanger

An ECMO circuit has three working components: a pump, a membrane lung (oxygenator), and a heat exchanger, which is usually integrated into the oxygenator. Venous blood is drained through the access cannula, driven through the membrane lung, and returned to an artery. Unlike a cardiopulmonary bypass machine, with its open reservoir, the ECMO circuit is closed and designed for long-term use.

The pump. Adult circuits use magnetically driven centrifugal pumps, in which a spinning rotor creates negative pressure at the central inlet and expels blood laterally. The pump is not occlusive, so its output at a given speed depends on preload (venous drainage) and afterload (return cannula resistance and arterial pressure). Flow is measured by an ultrasonic probe on the tubing, never inferred from pump speed.

The membrane lung. Oxygenators are made of polymethylpentene hollow fibers, with low resistance, easy de-airing, and rare plasma leak; larger adult devices prime with up to about 250 mL. Oxygenators have a rated flow, the volume of blood at 75% saturation that it can raise to 95% per minute.

Tubing and cannulas. Adult circuits use 3/8-inch tubing. Cannula size is stated in French (1 Fr = 0.33 mm, outer diameter). The pressure drop across a cannula is proportional to its length and inversely proportional to the fourth power of its diameter, so doubling the diameter reduces the drop to one-sixteenth. Common practice keeps the drop at or below 100 mm Hg; manufacturers' pressure-flow charts, measured with water, underestimate it with blood.

Safety components. A connector adds stagnant surface and a site for air entry, clot, or disconnection, so access points are kept to a minimum, with at least one before and one after the membrane lung. Access between the patient and the pump inlet is avoided, because that segment runs at negative pressure and entrains air. A bubble detector sits after the oxygenator. The console has a battery that runs the pump for at least 1 hour, a hand crank or backup drive, and an alarm or mechanism against reverse flow in venoarterial mode. A circuit primed under sterile conditions with an electrolyte solution can be held ready for up to 30 days.

MeasurementWhere it is takenWhat it reflects
Inlet (drainage) pressure, P1Between the venous cannula and the pumpSuction needed to drain blood; more negative with hypovolemia, tamponade, malposition, or a drainage kink
Pre-oxygenator pressure, P2Between the pump and the membrane lungPump output pressure; rises with oxygenator, tubing, return cannula, or arterial resistance
Post-oxygenator pressure, P3Between the membrane lung and the return cannulaReturn line and cannula resistance plus arterial pressure
Pressure drop across the oxygenator (P2 − P3)CalculatedInternal resistance of the membrane lung; a rise relative to flow indicates thrombus
Blood flowUltrasonic probeDelivered extracorporeal flow
Pre- and post-oxygenator saturation and gasesCircuit ports or in-line sensorsThe patient's oxygen supply and demand, and membrane lung gas transfer

Circuit pressures and monitored variables, from the 2022 ELSO circuit guideline.

Pump speed, blood flow, and flow targets

ELSO plans circuits for cardiac support to deliver about 3 L/min/m², roughly 60 mL/kg/min in adults. The individual target is physiologic: systemic oxygen delivery at least three times consumption. Because the circuit returns fully saturated blood, the drainage saturation reflects that ratio. A value above 66% indicates delivery more than three times consumption. The 2017 ELSO general guideline gives above 70% as the best single measure of adequate perfusion.

The 2021 ELSO venoarterial guideline sets flow at the lowest level that maintains organ perfusion and avoids anaerobic metabolism, because retrograde flow raises the pressure against which the left ventricle ejects. The balance is reassessed with lactate, drainage saturation, urine output, arterial pulse pressure, and echocardiographic aortic valve opening.

Inlet pressure and suction. When pump demand exceeds the volume available at the cannula, inlet pressure becomes strongly negative, the vessel collapses around the cannula, flow falls and fluctuates, and the drainage line shakes ("chattering"). The 2017 ELSO guideline states that the pump hemolyzes blood when inlet pressure is more negative than −300 mm Hg. Pump speed is lowered until flow is stable while the cause (hypovolemia, bleeding, tamponade, or a malpositioned cannula) is treated.

Outlet pressure and reverse flow. ELSO gives 400 mm Hg as the highest safe circuit pressure. A post-pump pressure above 300 mm Hg at the desired flow points to systemic hypertension, a small or kinked return cannula, kinked tubing, or a resistant oxygenator. A pump that stops, or turns too slowly to overcome arterial pressure, lets blood run backward from the arterial cannula into the venous line, so the return line is clamped whenever the pump stops.

Pattern at constant pump speedLikely siteCommon causesResponse
Flow falls, inlet pressure more negative, line chatteringDrainageHypovolemia, bleeding, tamponade, cannula tip against a wall or in a hepatic veinSpeed reduced to stable flow; volume; imaging of cannula and pericardium
Flow falls, pre- and post-oxygenator pressures high, oxygenator drop unchangedReturn side or patientSystemic hypertension, kinked return tubing or cannula, small return cannulaArterial pressure lowered to the lowest acceptable value; return limb inspected
Flow falls, pressure drop across oxygenator risingMembrane lungThrombus inside the oxygenatorAnticoagulation review, gas transfer tests, planned change
Flow falls with new pump noise or hemolysisPump headThrombus in the pumpPump head or circuit change

Interpretation of flow and pressure changes on a centrifugal circuit, derived from the pressure-flow relationships in the 2022 ELSO circuit guideline and the 2017 ELSO general guideline.

Sweep gas, FDO₂, and post-oxygenator blood gases

Sweep gas flow removes carbon dioxide; raising it lowers the Pco₂ of blood leaving the oxygenator and has little effect on oxygenation. FDO₂, set on the gas blender, together with blood flow determines oxygen delivery. ELSO describes starting sweep at a flow equal to blood flow (1:1) and adjusting from post-oxygenator gases.

In femoral VA-ECMO, blood leaving the oxygenator perfuses the lower body undiluted. The 2021 ELSO venoarterial guideline therefore requires post-membrane gases after any change in blender setting or gas flow. It targets normocapnia and slight hyperoxemia (post-oxygenator Po₂ about 150 mm Hg) and states that both hyperoxemia and hypocapnia have detrimental effects on outcome.

Early correction of hypercapnia and hyperoxemia. Patients often arrive acidotic and hypercapnic, and a high initial sweep can lower Pco₂ within minutes, causing cerebral vasoconstriction. The 2024 ELSO adult neurologic guideline cites a registry analysis of 11,972 venovenous patients in whom a fall in Pco₂ of more than 50% around cannulation was associated with acute brain injury. It also cites a single-center venoarterial study linking larger falls to intracranial hemorrhage; an analysis of 3,125 ECPR patients found no independent association. The guideline states that severe hyperoxemia (Pao₂ above 300 mm Hg) in the first 24 hours may be associated with brain injury and that avoiding severe hyperoxemia and hypoxemia by adjusting FDO₂ is reasonable.

Two arterial samples. In femoral VA-ECMO, a right radial sample reflects blood that has crossed the native lungs, while a femoral or post-oxygenator sample reflects circuit blood. The ventilator is titrated to the right radial gas, and sweep and FDO₂ to the post-oxygenator gas. ELSO advises lung-protective ventilation with tidal volumes of 6 to 8 mL/kg of ideal body weight and relatively high positive end-expiratory pressure (PEEP), which also opposes pulmonary edema; end-tidal carbon dioxide tracks residual pulmonary blood flow.

Sweep on a venoarterial circuit. Any configuration that returns blood to the arterial side is a right-to-left shunt and requires a working oxygenator. With the sweep off and flow continuing, the circuit delivers venous blood to the aorta, so the sweep is turned off only during a deliberate test with flow reversed or the circuit clamped.

FindingInterpretationAdjustment
Post-oxygenator Pco₂ lowExcess carbon dioxide removalSweep reduced in small steps
Post-oxygenator Pco₂ high despite high sweepReduced membrane carbon dioxide transferBrief high-flow sweep to clear condensation; oxygenator tested
Post-oxygenator Po₂ well above 150 mm HgHyperoxemia delivered to the lower bodyFDO₂ reduced
Post-oxygenator Po₂ low at FDO₂ 1.0Membrane lung failure or flow above rated flowOxygen transfer calculated; oxygenator changed if the trend confirms failure
Right radial Po₂ low, post-oxygenator Po₂ normalNative lung failure with native ejectionVentilator optimized; circuit reconfigured if persistent
Drainage saturation below 66–70%Delivery insufficient for consumptionFlow and hemoglobin increased; causes of high consumption treated

Gas-exchange adjustments on VA-ECMO, from the 2021 ELSO venoarterial guideline, the 2022 ELSO circuit guideline, and the 2017 ELSO general guideline.

Contraindications to VA-ECMO

ELSO treats most contraindications as relative, weighing the risk of the procedure and the use of a scarce resource against death without support. The central question is the exit: a patient with no prospect of recovery, durable left ventricular assist device (LVAD) placement, or transplantation has no exit. Age alone is not an absolute contraindication.

Contraindication (2021 ELSO venoarterial guideline)Reason it limits VA-ECMO
Cardiac recovery unlikely and no candidacy for transplantation or durable LVADNo exit strategy
Poor life expectancy (end-stage organ disease, malignant tumor, massive pulmonary embolism with cancer, chemotherapy-induced chronic cardiomyopathy)Underlying prognosis outweighs the benefit of support
Severe aortic regurgitationRetrograde aortic flow regurgitates into a ventricle that cannot empty, causing rapid distension and pulmonary edema
Severe aortic and peripheral vascular disease, including the axillary arteriesNo safe return site; dissection, embolism, and limb ischemia
Acute type A or B aortic dissection with extensive branch involvement (before repair)Retrograde flow can pressurize the false lumen and malperfuse branches
Severe neurologic impairment (prolonged anoxic brain injury, extensive trauma and bleeding)Neurologic outcome determines benefit; anticoagulation worsens intracranial bleeding
Severe immunologic disease with marked blood and coagulation disordersBleeding and infection on an anticoagulated circuit
Cirrhosis, Child-Pugh class B or CCoagulopathy, bleeding, and poor survival

Contraindications listed in the 2021 ELSO venoarterial guideline.

The 2026 ELSO ultrasound review recommends echocardiography before any initiation to exclude aortic dissection, significant aortic regurgitation, and severe atheromatous aortic disease, and to identify tamponade, tension pneumothorax, or large effusions that require treatment first.

Choosing the return site: femoral, axillary or subclavian, and central

The plan depends on the cause of cardiac failure and its expected course, the state of the lungs, the reserve of both ventricles, valve disease, arterial size and disease, coagulopathy, and any planned durable device or transplantation. ELSO designates peripheral femoral VA-ECMO as the default strategy for potentially reversible cardiogenic shock of any cause.

Return siteAdvantagesDisadvantagesTypical use
Femoral artery, percutaneousNo incision; fast, including during cardiopulmonary resuscitation; less bleeding; closed sternum; easy switch to a ventricular assist device (VAD)Leg ischemia; retrograde flow and higher left ventricular afterload; differential hypoxemia; poor suitability for long support; fewer percutaneous venting optionsDefault for reversible cardiogenic shock and ECPR
Axillary or subclavian artery, usually through a graftNo leg ischemia; less differential hypoxemia; mobilization during long support; transition to isolated left ventricular supportArm hyperperfusion with swelling; upper limb compartment syndrome; site bleeding; higher cerebral embolic risk; slower to establishIliofemoral disease, long expected runs, poor gas exchange, after aortic dissection repair
Ascending aorta with right atrial drainage (central)Antegrade flow; better drainage and higher flows; low differential hypoxemia; more venting optionsOpen sternum; high bleeding and infection risk; resternotomy for removal; more cerebral emboli; higher risk of a closed aortic valveFailure to wean from cardiopulmonary bypass with recovery expected within 7 days; salvage when femoral access is inadequate
Carotid arteryAccess when no other site is feasibleHigher risk of acute brain injuryLast resort in adults

Arterial return sites, adapted from Tables 5 to 7 and the cannulation text of the 2021 ELSO venoarterial guideline.

Central against peripheral. A meta-analysis of 17 retrospective series (1,691 patients, Raffa 2019) found no survival difference between central and peripheral VA-ECMO (risk ratio 1.00) and less bleeding, hemofiltration, and transfusion with peripheral cannulation. ELSO considers peripheral cannulation preferable when feasible, even after cardiotomy.

Axillary return. The axillary artery is cannulated through a tube graft or by a pseudo-percutaneous technique. Flow through a graft is not limited by the native vessel, so arm hyperperfusion is the characteristic complication, and ischemia is not typical of this route.

Reconfiguration. The circuit can be modified as the clinical problem changes. Options include a jugular return cannula (venoarteriovenous, or VAV, configuration) for refractory differential hypoxemia and conversion to venovenous ECMO when the heart recovers and the lungs do not. Advancing the femoral drainage cannula to the superior vena cava drains desaturated upper-body venous blood and sends better-saturated inferior vena cava blood through the right heart and native lungs. The centrifugal system with an oxygenator can also serve as a temporary left, right, or biventricular assist device when recovery is expected within about 8 weeks.

Percutaneous femoral cannulation: imaging, technique, and cannula sizing

ELSO recommends image-guided cannulation, with vascular ultrasound first and fluoroscopy when the procedure is done in the catheterization laboratory. The common femoral artery and vein are entered below the inguinal ligament and above their bifurcations. The examination starts in short axis to identify the common femoral vessels and the superficial and profunda femoral arteries, then rotates to long axis. The femoral head marks a compressible segment outside the pelvis, and the needle enters the anterior wall only.

Sequence. A heparin bolus, typically 50 to 100 units/kg, is given before cannula placement, even in a bleeding, coagulopathic patient. Guidewire positions are confirmed before dilation: the venous wire in the superior vena cava on the bicaval or subcostal view, and the arterial wire in the descending or abdominal aorta. The tract is dilated serially, with a stiff wire for obesity, tortuosity, or peripheral artery disease. The 2026 ELSO ultrasound review notes that obturators are removed before tip position is confirmed and that position is rechecked after the patient is moved.

Tip positions and sides. In peripheral VA-ECMO, the 2026 ELSO ultrasound review places the tip of the femoral drainage cannula at the superior vena cava–right atrial junction or above; a multistage cannula drains both caval territories through its side holes, which also makes it less likely to lose drainage when the tip abuts a wall. The right femoral vein is preferred for its straighter course, and ELSO prefers arterial and venous cannulas in separate limbs. A chest radiograph confirms position.

ElementUsual adult practiceBasis
Arterial return cannula15–17 Fr; 19–21 Fr only when higher flow is needed, as in sepsis2021 ELSO venoarterial guideline
Maximum cannula sizeOuter diameter in Fr ≤ 3 × vessel diameter in mm, with exceptions in small patients and severe vasoconstriction2026 ELSO ultrasound review
Drainage cannulaRight femoral vein; sized to target flow from the pressure-flow chart2021 ELSO venoarterial guideline; 2022 ELSO circuit guideline
Pressure drop across a cannulaUsually ≤100 mm Hg at target flow2022 ELSO circuit guideline
Distal perfusion cannula6–8 Fr short armored sheath into the superficial femoral artery, at cannulation2021 ELSO venoarterial guideline
Heparin at cannulationBolus 50–100 units/kg2017 ELSO general guideline

Femoral cannulation for adult VA-ECMO. Larger arterial cannulas increase vascular complications, including limb ischemia.

Percutaneous against surgical. ELSO states that the Seldinger technique may reduce local infection and possibly bleeding, without proven superiority. In 384 patients on femoro-femoral VA-ECMO (Dimberg 2026), percutaneous cannulation was associated with less site bleeding (29.3% against 40.9%) and site infection (8.3% against 31.0%); limb ischemia was associated with distal perfusion catheterization and arterial cannula size; it was not associated with cannulation technique.

Complications of cannulation. Vessel injury, hemorrhage, malposition, and distal ischemia are common with large cannulas, and ELSO advises cannulation only by experienced operators with continuous imaging. A drainage cannula across the interatrial septum or in a hepatic vein causes poor flow and is excluded by echocardiography. Thromboembolic events are underestimated and occurred in more than 25% of patients in series cited by ELSO, and venous thrombosis is screened by ultrasound after decannulation. ELSO recommends surgical decannulation for large cannulas. In 115 patients (Hwang 2016), percutaneous closure with two suture-based devices placed through the cannula had technical success similar to surgical repair (85.7% against 86.4%) and a shorter procedure.

Distal perfusion cannula and limb protection

A femoral arterial cannula fills most of the common femoral lumen, and the leg below depends on flow around it. ELSO recommends ipsilateral distal perfusion at the time of femoral cannulation, to prevent any delay in limb perfusion.

Technique. The cannula is placed under ultrasound guidance or direct vision into the superficial femoral artery, through a puncture of the common femoral artery below the arterial cannula and above the bifurcation. The tip must lie in the superficial femoral artery: perfusion into the profunda femoris does not protect the lower leg. The 2026 ELSO ultrasound review suggests placing the distal cannula first when possible, before the large cannula reduces flow and visibility in the superficial femoral artery. ELSO recommends a short 6 to 8 Fr armored cannula, since longer and larger sheaths cause trauma and spasm. Its side arm is joined with a male-to-male connector to the side port of the arterial return cannula. Flow is confirmed by Doppler or contrast ultrasound of the popliteal artery and, where measurable, distal flow of at least 100 mL/min.

Monitoring. Near-infrared spectroscopy (NIRS) on both calves compares tissue oxygenation. With 15 and 17 Fr cannulas, ELSO states that NIRS can inform the need for distal perfusion, using a tissue saturation above 50% (preferably 60%) and a difference of less than 20% between legs.

Evidence. ELSO cites a fall in leg ischemia from 25.4% to 9.7% with a distal perfusion cannula. A meta-analysis of 17 studies (2,040 patients, Gouchoe 2024) found less acute limb ischemia with a distal perfusion cannula regardless of size (relative risk 0.49), a larger reduction with prophylactic than reactive placement (relative risk 0.41), and no difference in mortality or bleeding.

Retrograde perfusion and distal cannula failure. Retrograde perfusion through the posterior tibial or dorsalis pedis artery needs no fluoroscopy. A distal cannula can kink, thrombose, or migrate into the profunda, and falling calf NIRS or loss of the popliteal signal leads to imaging. ELSO notes a risk of thrombus in the femoral segment between the two cannulas, one reason some clinicians favor surgical decannulation. Compartment syndrome is treated with fasciotomy, with embolectomy, vessel repair, or revascularization when needed.

Pulmonary artery venting: configurations, effectiveness, and place in practice

A pulmonary artery vent drains the main pulmonary artery into the venous limb of the circuit. By diverting right ventricular output away from the lungs, it reduces left atrial and left ventricular filling. It unloads indirectly: it lowers preload and does not move blood out of the left ventricle.

Configurations. In central ECMO after cardiac surgery, a cannula is placed directly in the pulmonary artery. Percutaneously, a cannula is introduced through the right internal jugular vein into the pulmonary artery. A dual-lumen right atrial–pulmonary artery cannula designed for right ventricular support has been run in drainage mode for this purpose (Kumar 2022). When a cannula is added to the circuit, ELSO describes flow-controlling clamps and additional flowmeters to apportion flow between limbs.

Effectiveness. In the ELSO table of unloading options, pulmonary artery venting receives three of five marks, equal to the lower range of atrial septostomy and below a transaortic microaxial pump or direct apical venting. In a porcine shock model on VA-ECMO (Meani 2021), a microaxial pump reduced the pressure-volume area by 34.7% and a pulmonary artery cannula by 9.7%, with similar end-organ perfusion. Clinical data are limited to small series; at one center, 5 of 78 vented VA-ECMO patients received a pulmonary artery vent (Robinson 2026).

Place in practice. The pulmonary artery vent is an option when a transaortic pump is not feasible (left ventricular thrombus, a mechanical aortic valve, a small ventricle, or no second arterial access), when the right internal jugular route is available, or with an open chest. Like other drainage-based vents, it reduces the native-lung blood available for the left ventricle to eject, an effect that ELSO lists among the configuration responses to differential oxygenation.

Prophylactic or rescue venting: the timing decision

The choice is between venting all patients at or soon after cannulation and venting only when distension appears. ELSO sets a low threshold for venting, guided by clinical and imaging signs, and states that impaired unloading compromises recovery and is addressed immediately.

Evidence. Two observational meta-analyses favor venting. Among 17 studies (3,997 patients, Russo 2019), mortality was 54% with unloading and 65% without (risk ratio 0.79). Among 62 studies (7,995 patients, Al-Fares 2019), venting was associated with improved weaning and lower 30-day mortality, but not lower in-hospital or 6-month mortality, and with more hemolysis (risk ratio 2.18). The short-term association was present for venting within 12 hours and absent for later venting. The only randomized test of timing, EARLY-UNLOAD, found no 30-day mortality difference between early transseptal venting and a conventional approach with rescue venting. Half the conventional group crossed over (unloading section). The data support venting when distension is present or imminent and do not support routine prophylactic venting.

SituationApproach supported by current data
Aortic valve opening, pulse pressure above 10 mm Hg, no pulmonary edemaLowest adequate flow, echocardiographic surveillance, pulmonary artery catheter for diastolic and wedge pressures; no vent
Closed aortic valve and absent pulsatility from the start, with severe left ventricular dysfunctionEarly mechanical venting considered; observational data favor venting within 12 hours
Pulse pressure below 5–10 mm Hg, closed valve with a dilating ventricle, smoke in the ventricle, rising wedge pressure or pulmonary edemaFlow reduction, vasodilation, moderate inotrope, higher PEEP, and diuresis, then mechanical venting without delay
Distension with failing native lungsDecompression that limits ejection (atrial septostomy, left atrial or pulmonary artery drainage) or VAV reconfiguration; a microaxial pump can worsen differential hypoxemia
Microaxial pump not feasibleAtrial septostomy, transseptal left atrial cannula, pulmonary artery vent, or surgical vent
Long support or bridge to durable LVAD expectedAxillary return, or conversion to a temporary surgical VAD with an oxygenator

Decision-level approach to venting, from the 2021 ELSO venoarterial guideline, the Russo 2019 and Al-Fares 2019 meta-analyses, and EARLY-UNLOAD. The methods are compared on the temporary support page.

Anticoagulation on VA-ECMO: agents, monitoring, and titration

The 2021 ELSO anticoagulation guideline recommends routine anticoagulation on VA-ECMO because of the risk of systemic embolism; retrospective reports of VA-ECMO without anticoagulation await randomized confirmation. The cross-device overview is in the anticoagulation section of the temporary support page.

Unfractionated heparin acts through antithrombin, binds plasma proteins and cells unpredictably, and has a half-life of 60 to 90 minutes in adults. A bolus of 50 to 100 units/kg is given at cannulation and an infusion titrated thereafter. Protamine reversal is almost never indicated during ECMO because the circuit can clot. Heparin-induced thrombocytopenia occurs in 0.2% to 5% of adults; its management is on the parenteral anticoagulants and heparin-induced thrombocytopenia page.

Direct thrombin inhibitors. Bivalirudin binds circulating and clot-bound thrombin, does not need antithrombin, has a half-life of about 25 minutes, is cleared mainly by proteolysis with about 20% renal excretion, and is removed by renal replacement therapy and plasmapheresis. Its proteolysis in stagnant blood is the concern on VA-ECMO: thrombus can form in a nonejecting left ventricle, and sampling lines and distal perfusion cannulas clot more often. Argatroban has a half-life of 39 to 51 minutes and hepatic clearance. Neither has an antidote. In a meta-analysis of six adult studies (Kido 2024), bivalirudin was associated with less thromboembolism (odds ratio 0.61) and circuit thrombosis (odds ratio 0.51) than heparin, without differences in mortality or major bleeding.

Assays. The activated clotting time (ACT) is a bedside whole-blood test affected by platelets, fibrinogen, temperature, and hemodilution, and ACT devices are not interchangeable. The activated partial thromboplastin time (aPTT) is the standard test for direct thrombin inhibitors and is shortened by the high factor VIII and fibrinogen of inflammation. Anti-factor Xa (anti-Xa) activity is falsely low with plasma free hemoglobin above 50 mg/dL, triglycerides above 500 mg/dL, or bilirubin above 6 mg/dL, all common on ECMO. ELSO recommends that centers have both a plasma test of anticoagulant effect and a whole-blood test of hemostasis such as viscoelastic testing.

TestSuggested frequency
ACTAt 1- to 2-hour intervals
aPTT, anti-Xa, platelet count2 to 4 times daily
International normalized ratio (INR), fibrinogen, complete blood countOnce or twice daily
AntithrombinDaily or as needed
Plasma free hemoglobinDaily
Thromboelastography or thromboelastometryDaily or as needed for bleeding or thrombosis

Monitoring schedule suggested in the 2021 ELSO anticoagulation guideline. In a stable patient without bleeding or clotting, once-daily blood count and coagulation tests are reasonable.

Anti-Xa (goal 0.3–0.7 IU/mL)aPTT lowaPTT in range (60–90 s)aPTT high
LowHeparin increasedHeparin increasedFactor deficiency considered; plasma given
In rangeHypercoagulable state consideredNo changeaPTT repeated; plasma given if still high
HighUnlikely; tests repeatedTests repeated; hypercoagulable state consideredHeparin decreased

ELSO algorithm for discordant anti-Xa and aPTT results. With anti-Xa titration, the infusion changes by 10% to 20% per value outside the chosen range (0.3–0.5, 0.4–0.6, or 0.5–0.7 IU/mL), and the ACT goal by 10 to 20 seconds.

Antithrombin replacement and blood product goals. A pilot randomized trial in venovenous ECMO found no benefit from antithrombin supplementation, and ELSO does not recommend routine replacement; it is considered when antithrombin is below 50% to 80% and goals cannot be reached at a maximal heparin dose.

ComponentBleeding patientNonbleeding patient
Platelets≥100,000/µL≥50,000–100,000/µL
INR<1.5<3
Fibrinogen>150 mg/dL (also before surgery)>100 mg/dL
Hemoglobin>7–9 g/dL; a restrictive strategy is reasonable in nonbleeding adults

Blood product goals from the 2021 ELSO anticoagulation guideline (hemoglobin and fibrinogen converted from g/L). In adult center surveys the most common thresholds were a platelet count of 50,000/µL and a fibrinogen of 200 mg/dL.

Holding heparin for bleeding. For excessive bleeding, particularly after cardiotomy, ELSO considers holding heparin for 4 to 6 hours reasonable, and up to 12 hours or longer, with frequent assessment of circuit thrombus. Low-flow periods, including weaning trials, require adequate anticoagulation.

Hemolysis on the circuit

A well-functioning circuit causes little hemolysis: plasma free hemoglobin is normally below 10 mg/dL, and the 2017 ELSO guideline calls for investigation above 50 mg/dL. In the ELSO registry (2014–2018), plasma free hemoglobin above 50 mg/dL was reported in 3.4% of adult venoarterial runs, with 32% survival to discharge. In 154 ECMO patients (Omar 2015), a level above 50 mg/dL at 24 hours occurred in 3.9% of survivors and 15.5% of nonsurvivors and independently predicted mortality (odds ratio 3.4).

Free hemoglobin scavenges nitric oxide and injures renal tubules. ELSO recommends daily plasma free hemoglobin; lactate dehydrogenase, bilirubin, and haptoglobin are confounded by liver injury and transfusion. The general approach to device hemolysis is in the hemolysis section of the temporary support page.

SourceCluesCorrection
Excess inlet suctionInlet pressure more negative than −300 mm Hg; chatteringLower pump speed; volume; drainage cannula repositioned or a second added
Pump head thrombusAbrupt, severe hemolysis; new pump noise; falling flow at the same speedPump head or circuit change
Oxygenator thrombusRising pressure drop relative to flow; rising D-dimer; falling platelets and fibrinogenPlanned oxygenator or circuit change
Flow through a narrow orificeHigh post-pump pressure; small or kinked return cannulaKink relieved; larger or additional return cannula; flow reduced
Attached deviceHemolysis after a hemofilter, continuous renal replacement therapy (CRRT), or plasmapheresis connectionConnection and pressures reviewed
Microaxial ventSuction alarms or malpositionRepositioning; preload corrected; speed reduced

Sources of hemolysis on VA-ECMO, from the 2017 ELSO general guideline, the Al-Fares 2019 meta-analysis (venting), and the 2021 ELSO venoarterial guideline, which reports pump head clots in fewer than 4% of runs, causing massive hemolysis.

The circuit is inspected with a flashlight for dark, nonmoving areas and white platelet-fibrin deposits at connectors. In venovenous patients who underwent circuit exchange (Hoffman 2023), plasma free hemoglobin rose 42-fold before the exchange, and the abnormalities reversed afterward, with lower membrane resistance.

Acute kidney injury and renal replacement therapy on the circuit

Acute kidney injury affects 50% to 70% of patients on ECMO, more often on venoarterial support, and more than half of those affected receive renal replacement therapy, according to the 2026 joint consensus of the Acute Disease Quality Initiative (ADQI) and ELSO. In the ELSO registry, renal replacement was used in 29.6% of adult venoarterial runs, with 31% survival; a meta-analysis of 1,866 adults (Cheng 2014) found acute kidney injury in 55.6% and renal replacement in 46.0%. A positive fluid balance is associated with worse outcomes, and ELSO advises a neutral to negative balance.

Fluid accumulation is the most common indication for renal replacement during ECMO, and continuous techniques are preferred. Modality, blood flow, and dose follow local protocol, because no comparative studies exist in patients on ECMO. Intermittent hemodialysis is generally avoided but may be considered during de-escalation.

ConfigurationAdvantagesDisadvantages
Separate (parallel) access through a dialysis catheterNo interference between circuits; full CRRT monitoring; precise fluid removalNew central access in an anticoagulated patient (bleeding, thrombosis, air embolism, bloodstream infection)
CRRT machine connected to the ECMO circuitNo additional access; full CRRT monitoring; longer filter life in several cohortsCRRT pressure alarms from ECMO pressures; air entrainment, embolization, thrombosis, disconnection bleeding
In-line hemofilterSimple, inexpensive, no extra accessInaccurate fluid removal; no pressure or air monitoring; inefficient clearance

Configurations for CRRT during ECMO, from the 2026 ADQI–ELSO consensus. An existing dialysis catheter favors a parallel circuit.

Integration. For a CRRT machine connected to the circuit, the consensus favors drawing from a positive-pressure point and returning before the membrane lung, so that the hollow fibers catch bubbles, clots, and particles before they reach the patient, a protection it calls particularly relevant in VA-ECMO. Recirculation through the oxygenator is about 2% to 3% at 3 to 5 L/min. ECMO pump flow determines the pressures the CRRT machine sees, which can exceed its alarm limits. In one matched cohort summarized in a 2026 review (Tomovic 2026), filter life was 48 hours with integrated CRRT against 20 hours with a separate catheter. Heparin for the ECMO circuit usually anticoagulates the filter, and regional citrate data in VA-ECMO are lacking.

Interactions with the oxygenator. The 2022 ELSO circuit guideline warns that a clotted CRRT circuit can falsely raise oxygenator pressure readings taken at the connection and can seed thrombus into the oxygenator, so a clotted set is removed promptly and repeated clotting leads to a different access. Drug clearance by the combined circuits is unpredictable, and drug levels guide dosing where available.

Infection on VA-ECMO

In 220 adults supported for more than 48 hours for cardiogenic shock (Schmidt 2012), 64% developed a nosocomial infection: ventilator-associated pneumonia in 55%, bloodstream infection in 18%, cannula infection in 10%, and mediastinitis in 11%. Infection with severe sepsis or septic shock independently predicted death (odds ratio 1.93). The ELSO registry recorded culture-proven infection in 7.6% of adult venoarterial runs, and a meta-analysis of 1,866 adults reported significant infection in 30.4%; rates vary with definitions and surveillance.

ECMO-specific sources. Cannula sites, an open or recently closed sternum, and multiple central lines add to the usual ICU sources. Central cannulation carries more infection than peripheral, and surgical cannulation more site infection than percutaneous. Bacteremia may arise from a circuit component but usually comes from another source, and the cannulas cannot be exchanged like an ordinary central line.

Diagnosis. The heat exchanger masks fever or hypothermia, and pulmonary edema confounds the chest radiograph (Abrams 2020). New instability, rising vasoactive needs, or falling oxygenator performance can be the first sign.

Prophylaxis and dosing. The 2017 ELSO general guideline states that no standard policy exists for prophylactic antibiotics in patients on extracorporeal life support, and the ELSO infectious disease task force recommended against routine prophylaxis, although many centers use it. In the same cohort (Schmidt 2012), antibiotics at cannulation were not significantly associated with fewer infections. Cannula sites are cleaned frequently with antiseptic. Antibiotic pharmacokinetics are altered by the circuit and by critical illness, and ELSO advises dose adaptation and drug-level monitoring.

An ECMO cart at an intensive care bedside. The upper shelf holds the ECMO system console, marked A, with tubing running to the patient's bed. The lower shelf holds a separate thermoregulatory heater-cooler device, marked B, with two displays and blue water hoses.
Bedside ECMO setup with (A) the ECMO system and (B) a water-filled thermoregulatory (heater-cooler) device that warms blood through the oxygenator's heat exchanger. Mycobacterium chimaera, linked to heater-cooler units in cardiac surgery, was found in 9 of 18 water samples from 10 such ECMO devices, and in bronchial specimens of 3 of 76 tested patients without signs of infection. Credit: Trudzinski FC, Schlotthauer U, Kamp A, et al. Euro Surveill 2016;21(46):30398, Figure 1, via Wikimedia Commons (source PMC5144944), CC BY 4.0.

Oxygenator failure and circuit thrombosis

ELSO reports progressive thrombotic deposition, especially in the oxygenator, in more than 10% of runs, which can lead to failure that requires exchange; the ELSO registry recorded circuit component clots in 9.2% of adult venoarterial runs. The registry defines a failing oxygenator as one requiring change because of clot, gas exchange failure, or blood leak.

ELSO permits observation of small clots of 1 to 5 mm at connectors; clots larger than 5 mm or enlarging after the membrane lung are removed by replacing that segment, or the whole circuit if there are many. Thrombus inside the oxygenator consumes platelets and fibrinogen and can produce a consumptive coagulopathy resembling disseminated intravascular coagulation, which resolves only when the device is changed.

IndicatorFinding that suggests failure
Pressure dropRising pressure drop, or drop divided by flow, with falling flow at the same pump speed (a rise without a flow change suggests measurement error)
Oxygen transferPost-oxygenator Po₂ below 300 mm Hg at FDO₂ 1.0; outlet saturation below 95% at flow below rated flow with inlet saturation of 70% or higher; falling daily transfer
Carbon dioxide transferPre-to-post Pco₂ difference below 10 mm Hg at maximal sweep, after a brief high-flow sweep to clear condensation
Coagulation and hemolysisRapidly rising D-dimer, falling platelets and fibrinogen, rising plasma free hemoglobin, after other causes are excluded
InspectionVisible clot growth with reduced flow, coagulopathy, or hemolysis; blood or plasma leak

Indicators of oxygenator failure from the 2022 ELSO circuit guideline and the 2017 ELSO general guideline.

Wet membrane and gas embolism. Water condensing in the gas path reduces carbon dioxide clearance; a few seconds of high sweep flow clears it. Sweep is kept below the manufacturer's do-not-exceed rate to protect the membrane. If blood pressure in the membrane lung falls below gas pressure, as when the pump stops with the oxygenator above the patient, gas can cross into the blood.

Planned and emergency oxygenator change. ELSO advises a planned change once the oxygenator no longer allows adequate flow or gas transfer or causes consumptive coagulopathy, because an urgent change under poor conditions is more dangerous. Unnecessary changes are also avoided, since any change interrupts support and exposes a new surface. A primed replacement circuit is kept available.

Membrane lung oxygen transfer and pressure gradient

Oxygen transferred by the membrane lung equals blood flow multiplied by the difference in oxygen content between blood leaving and entering the oxygenator. Adults consume about 3 to 5 mL/kg/min of oxygen. Samples for comparison with the manufacturer's rated transfer are drawn at an FDO₂ of 1.0.

Oxygen transfer—
Oxygen transfer per kg—
Pre-to-post Pco₂ difference—
Pressure drop—
Pressure drop per L/min—
—

Oxygen content (mL/dL) = 1.34 × hemoglobin × saturation + 0.003 × Po₂. Oxygen transfer (mL/min) = flow (L/min) × 10 × (post-oxygenator content − pre-oxygenator content). In the ELSO example, hemoglobin 12 g/dL, saturation rising from 71% to 100%, post-oxygenator Po₂ 365 mm Hg, and flow 5 L/min give 288 mL/min. That example omits dissolved oxygen in pre-oxygenator blood, so the calculator returns 288 mL/min with a pre-oxygenator Po₂ of 0 and 282 mL/min with a pre-oxygenator Po₂ of 40 mm Hg. Thresholds from the 2022 ELSO circuit guideline and the 2017 ELSO general guideline.

Circuit emergencies on VA-ECMO

A patient on full VA-ECMO depends on the circuit for most of the cardiac output, so circuit failure is a circulatory arrest. The emergencies share first steps: the lines are clamped close to the patient when flow must stop, the ventilator is set to support settings, and the cause is found and repaired quickly. Securing the cannulas to the skin in at least two places, frequent checks of position and fixation, and adequate sedation prevent most mechanical emergencies.

EmergencyRecognitionImmediate management
Air in the circuitBubble alarm or visible air; drainage-side air is usually small bubbles caught by the membrane lung; return-side air is the dangerous formPump stopped, lines clamped near the patient, ventilator on support settings; source repaired (cannula insertion site, connector, open stopcock, or infusion line)
Pump or power failureLoss of flow; power or retrograde-flow alarmReturn line clamped; battery (at least 1 hour), backup console, or hand crank; console on an emergency-power outlet
Accidental decannulationBleeding at the site, air in the drainage line, loss of flowBoth lines clamped near the patient; direct pressure; cannula reinserted as soon as possible
Sudden high circuit pressureHigh-pressure alarm with falling flowPump stopped, then flow restored gradually while the occlusion is found
Acute oxygenator failureAbrupt fall in post-oxygenator Po₂ or rise in Pco₂, leak, or pressure riseVentilator and FDO₂ increased; emergency change with a primed circuit
Severe drainage failureChattering, very negative inlet pressure, flow collapseSpeed reduced; volume; tamponade and cannula position imaged

Circuit emergencies, from the 2017 ELSO general guideline and the 2022 ELSO circuit guideline. The post-oxygenator bubble detector is kept active whenever the return side is accessed.

ELSO recommends a dedicated ECMO team with regular retraining and regional hub-and-spoke networks for hospitals that cannot provide ECMO.

EURO SHOCK: VA-ECMO for shock persisting after primary percutaneous coronary intervention

EURO SHOCK was a multicenter European randomized trial in cardiogenic shock complicating acute myocardial infarction that persisted 30 minutes after primary percutaneous coronary intervention (PCI) of the culprit lesion. Patients were assigned to VA-ECMO or continued standard therapy, with 30-day all-cause mortality by intention to treat as the primary end point. The COVID-19 pandemic stopped the trial after 35 patients (17 VA-ECMO, 18 standard therapy).

Thirty-day mortality was 43.8% with VA-ECMO and 61.1% with standard therapy (hazard ratio 0.56; 95% CI 0.21 to 1.45; P = 0.22). Vascular complications occurred in 21.4% and 0% of the VA-ECMO and standard therapy groups, respectively, and bleeding in 35.7% and 5.6%.

EURO SHOCK is too small to change practice. Its distinct question is VA-ECMO for shock that persists after reperfusion, and its excess of bleeding and vascular complications is consistent with the larger trials. The individual patient data meta-analysis of 567 patients found no mortality benefit from early routine VA-ECMO in infarct-related shock; the larger trials and the guideline positions are reviewed on the temporary support and cardiogenic shock pages.

Sources of error in VA-ECMO circuit management

References

  1. Lorusso R, Shekar K, MacLaren G, et al. ELSO interim guidelines for venoarterial extracorporeal membrane oxygenation in adult cardiac patients. ASAIO J. 2021;67:827–844 (erratum: 2022;68:e133). doi:10.1097/MAT.0000000000001510
  2. McMichael ABV, Ryerson LM, Ratano D, Fan E, Faraoni D, Annich GM. 2021 ELSO adult and pediatric anticoagulation guidelines. ASAIO J. 2022;68:303–310. doi:10.1097/MAT.0000000000001652
  3. Gajkowski EF, Herrera G, Hatton L, Velia Antonini M, Vercaemst L, Cooley E. ELSO guidelines for adult and pediatric extracorporeal membrane oxygenation circuits. ASAIO J. 2022;68:133–152 (erratum: 2022;68:e131). doi:10.1097/MAT.0000000000001630
  4. Extracorporeal Life Support Organization. ELSO general guidelines for all ECLS cases, version 1.4. Ann Arbor, MI: ELSO; August 2017. elso.org
  5. Cho SM, Hwang J, Chiarini G, et al. Neurological monitoring and management for adult extracorporeal membrane oxygenation patients: Extracorporeal Life Support Organization consensus guidelines. ASAIO J. 2024;70:e169–e181. doi:10.1097/MAT.0000000000002312
  6. Douflé G, O'Neil ER, Abreu S, et al. Ultrasound in extracorporeal membrane oxygenation: an ELSO state-of-the-art review. ASAIO J. 2026;72:85–98. doi:10.1097/MAT.0000000000002656
  7. Gist KM, Selewski DT, Kashani KB, et al. Extracorporeal membrane oxygenation, acute kidney injury, fluid balance, and continuous renal replacement therapy: Acute Disease Quality Initiative (ADQI) and Extracorporeal Life Support Organization (ELSO) joint consensus conference. Intensive Care Med. 2026;52:1175–1195. doi:10.1007/s00134-026-08440-3
  8. Tomovic F, Thibault-Baum A, Glogonjac N, et al. Renal replacement therapy in patients under extracorporeal membrane oxygenation: a narrative review. Crit Care. 2026;30:317. doi:10.1186/s13054-026-06033-6
  9. Thiele H, Zeymer U, Akin I, et al. Extracorporeal life support in infarct-related cardiogenic shock (ECLS-SHOCK). N Engl J Med. 2023;389:1286–1297. doi:10.1056/NEJMoa2307227
  10. Ostadal P, Rokyta R, Karasek J, et al. Extracorporeal membrane oxygenation in the therapy of cardiogenic shock: results of the ECMO-CS randomized clinical trial. Circulation. 2023;147:454–464. doi:10.1161/CIRCULATIONAHA.122.062949
  11. Banning AS, Sabaté M, Orban M, et al. Venoarterial extracorporeal membrane oxygenation or standard care in patients with cardiogenic shock complicating acute myocardial infarction: the multicentre, randomised EURO SHOCK trial. EuroIntervention. 2023;19:482–492. doi:10.4244/EIJ-D-23-00204
  12. Zeymer U, Freund A, Hochadel M, et al. Venoarterial extracorporeal membrane oxygenation in patients with infarct-related cardiogenic shock: an individual patient data meta-analysis of randomised trials. Lancet. 2023;402:1338–1346. doi:10.1016/S0140-6736(23)01607-0
  13. Kim MC, Lim Y, Lee SH, et al. Early left ventricular unloading or conventional approach after venoarterial extracorporeal membrane oxygenation: the EARLY-UNLOAD randomized clinical trial. Circulation. 2023;148:1570–1581. doi:10.1161/CIRCULATIONAHA.123.066179
  14. Russo JJ, Aleksova N, Pitcher I, et al. Left ventricular unloading during extracorporeal membrane oxygenation in patients with cardiogenic shock. J Am Coll Cardiol. 2019;73:654–662. doi:10.1016/j.jacc.2018.10.085
  15. Al-Fares AA, Randhawa VK, Englesakis M, et al. Optimal strategy and timing of left ventricular venting during veno-arterial extracorporeal life support for adults in cardiogenic shock: a systematic review and meta-analysis. Circ Heart Fail. 2019;12:e006486. doi:10.1161/CIRCHEARTFAILURE.119.006486
  16. Meani P, Mlcek M, Kowalewski M, et al. Transaortic or pulmonary artery drainage for left ventricular unloading in venoarterial extracorporeal life support: a porcine cardiogenic shock model. Semin Thorac Cardiovasc Surg. 2021;33:724–732. doi:10.1053/j.semtcvs.2020.11.001
  17. Robinson EM, Yoshida S, Bryner B, et al. Comparison of left ventricular unloading strategies during extracorporeal membrane oxygenation. ASAIO J. 2026;72:750–756. doi:10.1097/MAT.0000000000002580
  18. Kumar K, Coonse K, Zakhary B, Cigarroa JE. Novel method for left ventricular unloading utilizing percutaneous pulmonary artery drainage in cardiorespiratory failure due to COVID-19 infection. Catheter Cardiovasc Interv. 2022;100:175–178. doi:10.1002/ccd.30212
  19. Raffa GM, Kowalewski M, Brodie D, et al. Meta-analysis of peripheral or central extracorporeal membrane oxygenation in postcardiotomy and non-postcardiotomy shock. Ann Thorac Surg. 2019;107:311–321. doi:10.1016/j.athoracsur.2018.05.063
  20. Gouchoe DA, Chaurasia S, Henn MC, et al. Does size matter? The effect of size of distal perfusion catheter on acute limb ischemia: a meta-analysis. ASAIO J. 2024;70:853–860. doi:10.1097/MAT.0000000000002178
  21. Dimberg A, Dalén M, Franco-Cereceda A, Fux T. Percutaneous versus surgical cannulation for femoro-femoral venoarterial extracorporeal membrane oxygenation: a retrospective cohort study on cannulation-related complications. Artif Organs. 2026;50:440–448. doi:10.1111/aor.70061
  22. Hwang JW, Yang JH, Sung K, et al. Percutaneous removal using Perclose ProGlide closure devices versus surgical removal for weaning after percutaneous cannulation for venoarterial extracorporeal membrane oxygenation. J Vasc Surg. 2016;63:998–1003.e1. doi:10.1016/j.jvs.2015.10.067
  23. Kido K, Kabulski GM, Szymanski TW, et al. Meta-analysis comparing bivalirudin versus unfractionated heparin in adult patients with extracorporeal membrane oxygenation. J Pharm Pract. 2024;37:429–434. doi:10.1177/08971900221143406
  24. Omar HR, Mirsaeidi M, Socias S, et al. Plasma free hemoglobin is an independent predictor of mortality among patients on extracorporeal membrane oxygenation support. PLoS One. 2015;10:e0124034. doi:10.1371/journal.pone.0124034
  25. Hoffman KR, Diehl A, Paul E, Burrell AJC. The hematological effects of extracorporeal membrane oxygenator exchange. ASAIO J. 2023;69:e308–e314. doi:10.1097/MAT.0000000000001976
  26. Schmidt M, Bréchot N, Hariri S, et al. Nosocomial infections in adult cardiogenic shock patients supported by venoarterial extracorporeal membrane oxygenation. Clin Infect Dis. 2012;55:1633–1641. doi:10.1093/cid/cis783
  27. Abrams D, Grasselli G, Schmidt M, Mueller T, Brodie D. ECLS-associated infections in adults: what we know and what we don't yet know. Intensive Care Med. 2020;46:182–191. doi:10.1007/s00134-019-05847-z
  28. Cheng R, Hachamovitch R, Kittleson M, et al. Complications of extracorporeal membrane oxygenation for treatment of cardiogenic shock and cardiac arrest: a meta-analysis of 1,866 adult patients. Ann Thorac Surg. 2014;97:610–616. doi:10.1016/j.athoracsur.2013.09.008
  29. Trudzinski FC, Schlotthauer U, Kamp A, et al. Clinical implications of Mycobacterium chimaera detection in thermoregulatory devices used for extracorporeal membrane oxygenation (ECMO), Germany, 2015 to 2016. Euro Surveill. 2016;21(46):30398. doi:10.2807/1560-7917.ES.2016.21.46.30398