Library › Devices and mechanical support › Impella and percutaneous ventricular assist devices
Impella and other percutaneous ventricular assist devices: models, console, purge, access, protected PCI and removal
Device-level reference for the Impella family, TandemHeart, and other percutaneous ventricular assist devices in adults, from models, labeling, access, positioning, console waveforms, purge, and hemolysis through early-unloading protocols, protected high-risk PCI, weaning, and removal.
Percutaneous ventricular assist devices move blood actively from a cardiac chamber into a great artery. The Impella family pumps from the left ventricle into the ascending aorta (or from the inferior vena cava into the pulmonary artery), and the TandemHeart drains the left atrium into the femoral artery. In the United States the Impella CP is labeled for up to 6 hours during high-risk percutaneous coronary intervention (PCI) and up to 4 days in cardiogenic shock, and the Impella 5.5 and the right-sided Impella RP for up to 14 days. Safe use depends on device-level detail: the labeled contraindications, the inlet depth below the aortic annulus, the console waveforms and alarms, the purge, hemolysis surveillance, access and closure, and a planned exit.
This page covers the devices themselves. Shock staging and phenotypes, device selection and comparison, the DanGer Shock overview, pressure-volume physiology, anticoagulation, bleeding, limb protection, and the general escalation and weaning framework are on the temporary mechanical circulatory support page. The hemodynamic definition and management of shock are on the cardiogenic shock page.
Percutaneous ventricular assist devices: definition and device classes
A percutaneous ventricular assist device is a catheter- or cannula-based pump, inserted without sternotomy, that adds forward flow in series with a failing ventricle. Transvalvular microaxial pumps (the Impella family) carry a miniature impeller inside a cannula that crosses the aortic valve, or on right-sided models the tricuspid and pulmonary valves, and generate continuous flow. Extracorporeal centrifugal pumps draw blood through one percutaneous cannula and return it through another: the TandemHeart drains the left atrium by a transseptal cannula, and the ProtekDuo drains the right atrium and returns blood to the pulmonary artery through a dual-lumen jugular cannula. Neither design oxygenates blood unless an oxygenator is added, and a left-sided pump can deliver no more blood than crosses the lungs.
The Impella platform: microaxial pump, controller, and purge
All Impella pumps share the same architecture: a pigtail that stabilizes the tip and keeps the inlet off the wall, an inlet area, a cannula, an outlet area, and a motor housing with an impeller. On left-sided pumps the inlet sits in the left ventricle and the outlet in the ascending aorta, so the pump empties the ventricle in systole and diastole. The motor housing measures 12 French (Fr) on the Impella 2.5, 14 Fr on the CP, and 21 Fr on the 5.0 and LD.
The Automated Impella Controller drives the motor, delivers the purge, displays waveforms, and generates alarms. Support is set as a performance level (P-level): P-0 to P-8 with an AUTO setting on the 2.5 and the original CP, and P-0 to P-9 on the CP with SmartAssist, 5.0, 5.5, and LD. Flow is not measured; the controller estimates it from motor current and the pressure difference between outlet and inlet, so the displayed flow is accurate only when the pump is correctly positioned and the inlet is free.
The purge is a continuous pressurized dextrose infusion, conventionally heparinized, that flows through the motor housing and exits between motor and rotor, opposing the entry of blood into the bearing gap; loss of the purge allows thrombus into the motor and stops the pump. Position is sensed by a fluid-filled placement lumen on older 2.5 and CP pumps, a differential membrane sensor on the 5.0 and LD, and an optical sensor at the outlet on current CP and 5.5 pumps with SmartAssist, which also displays a derived left ventricular pressure, an estimated end-diastolic pressure, and suction detection.
A setting of P-0 produces retrograde flow from aorta to ventricle through the cannula when the pump lies across the valve, and P-1 can do the same. A left-sided Impella is therefore kept at P-2 or above while it crosses the valve. The controller raises motor speed when minimum flow falls below 0.1 L/min.
Impella models: 2.5, CP, 5.0, LD, 5.5, and ECP
| Model | Peak flow | Access and sheath | Position target (annulus to inlet) | Status in the United States |
|---|---|---|---|---|
| Impella 2.5 | About 2.5 L/min | Percutaneous femoral, 13 Fr | About 3.5 cm | No longer marketed; the pump of PROTECT II |
| Impella CP (with SmartAssist) | About 3.5 to 3.7 L/min | Percutaneous femoral or axillary, 14 Fr | About 3.5 cm (3 cm in a small ventricle) | Approved: high-risk PCI up to 6 hours; shock up to 4 days |
| Impella 5.0 | About 5 L/min | Surgical cutdown to femoral or axillary artery; 21 Fr motor housing | About 3.5 cm | Approved for shock up to 14 days; largely replaced by the 5.5 |
| Impella LD | About 5 L/min | Graft on the ascending aorta at sternotomy | About 3.5 cm | Approved for shock up to 14 days; used in the operating room |
| Impella 5.5 (with SmartAssist) | Mean up to 5.5 L/min; peak about 6.2 L/min | 10-mm graft sewn to the axillary artery (direct aortic access possible) | About 5.0 cm | Approved for shock up to 14 days |
| Impella ECP | CP-range flow | Percutaneous femoral, 9 Fr insertion, expands after insertion | Not established | Investigational |
Left-sided Impella pumps. Flows are manufacturer and review figures and are reached only with adequate preload and right ventricular output. Position targets from the US instructions for use and the 2024 American Society of Echocardiography (ASE) recommendations; labeled durations from the current US indications.
Impella 2.5 and CP. The 2.5 was the pump of PROTECT II and ISAR-SHOCK. The CP replaced it and is the pump of IMPRESS in Severe Shock, DanGer Shock, PROTECT III, PROTECT IV, CHIP-BCIS3, and STEMI-DTU.
Impella 5.0 and LD. The 5.0 placed a 21 Fr pump through a surgical cutdown on the femoral or axillary artery, and the LD placed the same pump through a graft on the ascending aorta when the chest was open. The instructions for use note that the 5.0 tends to be drawn into the ventricle at P-9, so position is confirmed at maximal support. The long, flexible 5.0 was difficult to steer from the axilla, and transcaval delivery was developed largely for it; the 5.5 has replaced both in most programs.
Impella 5.5. The 5.5 is shorter and stiffer than the 5.0, is placed through a 10-mm graft sewn to the axillary artery, and delivers a mean flow of up to 5.5 L/min (peak flow about 6.2 L/min) while the patient sits and walks. The first 200 US patients (2019–2020) were supported for a median of 10 days; survival to explant was 74.0% overall and 51.4% when extracorporeal membrane oxygenation (ECMO) was also required (Ramzy 2021). It is the device for prolonged full left-sided support in heart failure–related shock and as a bridge to recovery, durable left ventricular assist device (LVAD), or transplantation. DanGer Shock enrolled only ST-elevation myocardial infarction (STEMI)–related shock, and evidence in heart failure–related shock consists of registries and series. Bridging and heart allocation are on the temporary mechanical circulatory support and durable left ventricular assist device pages.
Impella ECP. The ECP is inserted through a 9 Fr sheath and expands to deliver CP-range flow, so the arteriotomy can be closed with a small-bore device. It is investigational in the United States. Its single-arm pivotal study enrolled 256 patients undergoing high-risk PCI and was presented at the 2024 Transcatheter Cardiovascular Therapeutics meeting. The study has not been published in a journal. The sponsor reported a 30-day major adverse cardiac and cerebrovascular event rate of 6.3%, below the prespecified performance goal (Johnson & Johnson MedTech 2024).
US labeling: indications and labeled duration
| Indication | Devices | Labeled population and duration |
|---|---|---|
| High-risk PCI | Impella 2.5, CP, CP with SmartAssist | Up to 6 hours during elective or urgent PCI in hemodynamically stable patients with severe coronary disease, when a heart team including a cardiac surgeon has determined that high-risk PCI is the appropriate option |
| Cardiogenic shock (left ventricle) | Impella 2.5, CP, CP with SmartAssist | Up to 4 days, for ongoing shock within 48 hours of acute myocardial infarction or open-heart surgery, or in cardiomyopathy (including peripartum cardiomyopathy) or myocarditis, from isolated left ventricular failure not responsive to optimal medical management |
| Cardiogenic shock (left ventricle) | Impella 5.0, 5.5 with SmartAssist, LD | Up to 14 days, same clinical indication |
| Acute right heart failure | Impella RP, RP with SmartAssist, RP Flex with SmartAssist | Up to 14 days, body surface area 1.5 m² or more, right heart failure of less than 48 hours after LVAD implantation, myocardial infarction, heart transplantation, or open-heart surgery, without profound shock, end-organ failure, or acute neurologic injury |
US indications for use of Impella devices as published by the manufacturer; pediatric weight limits omitted.
The label differs from the evidence: the shock indication predates any randomized mortality benefit, which DanGer Shock later showed only in STEMI-related shock with left ventricular failure. The 2025 acute coronary syndrome guideline of the American College of Cardiology (ACC), the American Heart Association (AHA), and partner societies gives a microaxial pump in selected patients with STEMI and severe or refractory shock a Class 2a recommendation. The 2021 ACC/AHA/Society for Cardiovascular Angiography and Interventions (SCAI) revascularization guideline gives elective hemodynamic support in selected high-risk PCI a Class 2b (B-R) recommendation. The full table is on the temporary mechanical circulatory support page.
Contraindications in the US labeling
| Labeled contraindication (left-sided Impella) | Applies to | Reason |
|---|---|---|
| Mural thrombus in the left ventricle | All indications | The inlet aspirates thrombus, causing pump stoppage or systemic embolism |
| Mechanical aortic valve or heart constrictive device | All indications | The cannula cannot cross a mechanical valve safely |
| Aortic stenosis or calcification equivalent to an orifice area of 0.6 cm² or less | All indications | Crossing is difficult, and the cannula occupies much of the residual orifice |
| Moderate to severe aortic insufficiency (echocardiographic grade 2+ or more) | All indications | Pumped blood regurgitates into the ventricle, so net forward flow falls and the ventricle is not unloaded |
| Severe arterial disease precluding placement | All indications | Large-bore access and passage to the aortic valve are unsafe |
| Atrial or ventricular septal defect, including post-infarct ventricular septal rupture | All indications | Lowering left-sided pressure can reverse the shunt and cause systemic desaturation |
| Significant right heart failure | Cardiogenic shock | An unsupported right ventricle cannot fill the pump |
| Combined cardiorespiratory failure | Cardiogenic shock | The pump provides no gas exchange |
| Left ventricular rupture | Cardiogenic shock | Catheter manipulation and lower cavity pressure risk extending the rupture |
| Cardiac tamponade | Cardiogenic shock | Preload is limited by the pericardium, and drainage takes precedence |
Contraindications in the current US important risk information for the Impella 2.5, CP, CP with SmartAssist, 5.0, 5.5 with SmartAssist, and LD. The reasons column summarizes the mechanism.
Reviews treat left ventricular thrombus and a mechanical aortic valve as absolute, with aortic dissection added, and list obstructive hypertrophic cardiomyopathy, ascending aortic aneurysm, hematologic disorders predisposing to hemolysis, and significant peripheral artery disease as relative. The instructions for use specify echocardiography before insertion to exclude thrombus, a mechanical aortic valve, and severe aortic insufficiency, and imaging of the vasculature and access site.
The label excludes this degree of stenosis, although CP and 5.5 pumps have been placed across stenotic valves in experienced centers to support valvuloplasty or transcatheter replacement. New or worsening aortic regurgitation is also a complication of support and was part of the DanGer Shock adverse-event composite. Ventricular septal rupture is a labeled contraindication, and the balloon pump is the conventional bridge to repair. In one retrospective series of 18 patients, off-label axillary 5.0 and 5.5 pumps bridged to delayed repair, with survival to discharge of 5 of 5, against 4 of 13 with percutaneous pumps (Murakami 2025); such a series is too small and selected to define practice.
Unloading, coronary perfusion, and the door-to-unload hypothesis
A transvalvular pump lowers end-diastolic pressure and volume, wall stress, and oxygen demand while raising aortic pressure, and the pressure-volume loop becomes triangular as support rises; the loop analysis is on the temporary mechanical circulatory support page. The wider coronary perfusion gradient improves the supply-demand balance in failing hearts. In normal goat hearts, however, full Impella 5.5 support raised coronary perfusion pressure while coronary flow fell by about one-third, because unloading raised coronary vascular resistance (Ito 2026).
The door-to-unload hypothesis came from animal models in which 30 minutes of unloading before reperfusion reduced infarct size. The DTU-STEMI pilot (50 patients with anterior STEMI without shock) found delayed reperfusion after 30 minutes of Impella CP unloading feasible. The pivotal STEMI-DTU trial randomized 527 adults with a first anterior STEMI within 1 to 6 hours of symptom onset, at 55 hospitals, to 30 minutes of unloading before PCI or immediate PCI. Infarct size on cardiac magnetic resonance at 3 to 5 days was 30.8% and 31.9% of left ventricular mass (mean difference −1.1%; 95% confidence interval (CI) −4.2 to 2.0; P = 0.50), total ischemic time was longer with unloading, and major bleeding or vascular complications were more frequent. Primary PCI is not delayed for device placement in STEMI without shock.
Femoral insertion and the large-bore sheath
The common femoral artery is the default route for the CP. Contrast computed tomography (CT) or angiography in elective cases, and ultrasound in all cases, define vessel diameter, calcification, tortuosity, and the level of the bifurcation. The artery is entered under ultrasound and fluoroscopic guidance with a single anterior-wall puncture over the femoral head, below the inferior epigastric artery and above the bifurcation; a high puncture risks retroperitoneal hemorrhage, and a low one pseudoaneurysm and occlusion. For planned percutaneous removal, one or two suture-based closure devices are deployed before upsizing ("preclose"); plug-based large-bore devices are alternatives.
After dilation to the 14 Fr introducer, heparin is given to an activated clotting time of 250 seconds or more, per the instructions for use, before the catheter is introduced. The pump is advanced on a 0.018-inch wire around the arch and across the valve under fluoroscopy, with the inlet about 3.5 cm below the annulus and free of the mitral chordae. After the wire is removed, position is rechecked on the placement signal and by imaging, and excess catheter slack is removed so that the pump does not migrate inward. After insertion and until explant, the US instructions for use (2016 and 2020 editions) specify an activated clotting time of 160 to 180 seconds. Prolonged support is usually monitored with activated partial thromboplastin time (aPTT) or anti-Xa activity, as described on the temporary mechanical circulatory support page.
When support continues, the peel-away introducer is removed and the catheter is secured through the repositioning sheath, which permits sterile repositioning. In protected PCI the coronary guide can be passed through the Impella sheath (single access), avoiding a second large puncture. Limb monitoring, distal perfusion catheters, and contralateral bypass are on the temporary mechanical circulatory support page.
Axillary and transcaval access
Axillary access. The axillary artery is the alternative when the iliofemoral route is unsuitable and the preferred route for prolonged support, because it frees the groin. The CP can be placed through an ultrasound-guided percutaneous axillary puncture with preclosure, with a balloon and covered stent available from the femoral side; insertion under regional nerve block has been reported. The 5.0 and 5.5 require surgical cutdown with an end-to-side graft; for the 5.5, a 10-mm graft is tunneled through a counterincision and the pump is advanced over a stiff 0.018-inch wire under fluoroscopy and echocardiography. The right axillary artery is commonly used. The axillary artery is smaller and more fragile than the femoral artery and lies against the brachial plexus, so hematoma with plexus compression, dissection, upper-limb ischemia, and graft infection are the specific complications. Brachial cutdown through a 6-mm graft has served as a single-access route when femoral and axillary arteries were both unsuitable.
Transcaval access. A wire is passed from the inferior vena cava into the adjacent abdominal aorta at a calcium-free window identified on CT, and the arterial device is delivered through a venous sheath; at removal the aortocaval track is closed with a nitinol occluder or, when unsuitable, a covered aortic stent. In 72 patients in shock (2015–2023), transcaval delivery of an Impella 5.0 succeeded in all; no access-site vascular complication occurred, and in-hospital survival was 44.4% (Mohammed 2025). The technique requires CT planning and immediate endovascular bailout.
Positioning by fluoroscopy and echocardiography
Correct position places the inlet mid-ventricle, pointing toward the apex and free of the mitral apparatus and walls, and the outlet in the ascending aorta. The ASE measures depth on parasternal or transesophageal long-axis views from the annulus to the middle of the inlet area: about 3.5 cm for the CP and about 5.0 cm for the 5.5 (ASE 2024), and 3 cm for the CP in a small ventricle per the instructions for use. Position changes with movement, resuscitation, transport, ventricular size, and catheter slack, so it is reassessed daily and after any transfer, resuscitation, alarm change, unexplained flow change, or new hemolysis. The report names the device and records depth, inlet orientation, valve function, right ventricular function, and the pericardium.
| Finding | Echocardiographic appearance | Consequence |
|---|---|---|
| Correct position | Inlet about 3.5 cm (CP) or 5.0 cm (5.5) below the annulus, mid-cavity, toward the apex; mosaic outlet jet in the proximal ascending aorta | Expected flow and unloading |
| Too shallow | Inlet near the valve; aliased flow at the inlet | Recirculation, low effective flow, risk of slipping into the aorta |
| Too deep | Inlet against the wall or among the chordae | Suction, ectopy, hemolysis, new mitral regurgitation; outlet below the valve |
| Malrotated | Correct depth, inlet facing the lateral wall or mitral valve | Suction, hemolysis, incomplete unloading |
| In the aorta | Inlet and outlet both above the valve | No unloading |
| Complication | New effusion, new or worse aortic regurgitation, aortic wall injury | Perforation, valve injury, dissection |
Echocardiographic patterns of microaxial pump position, after the 2024 ASE recommendations and the US instructions for use. The high-velocity jet just above the aortic valve is pump output and can be mistaken for aortic regurgitation.
Repositioning is done at P-2 under echocardiographic guidance, by advancing, withdrawing, or rotating the catheter through the repositioning sheath. Because two-dimensional imaging cannot fully show the inlet's relation to the ventricle, the instructions for use recommend repositioning a pump with suction or hemolysis even when the image looks correct.
Console waveforms: placement signal, motor current, and left ventricular pressure
SmartAssist pumps display three signals. The placement signal (red), the pressure at the outlet sensor, has an aortic contour when the outlet is in the aorta. The motor current (green) is pulsatile when inlet and outlet lie on opposite sides of the valve, because the pressure difference across the cannula changes through the cardiac cycle. The derived left ventricular pressure (white), aortic pressure minus the pressure head estimated from motor speed and current, has a ventricular contour and gives an estimated end-diastolic pressure.
| Console pattern | Interpretation | Response |
|---|---|---|
| Aortic placement signal, ventricular white trace, pulsatile motor current | Correct position | Depth confirmed daily by echocardiography |
| Placement signal becomes ventricular; motor current flattens | Whole pump in the ventricle | P-2, withdrawal under imaging, P-level restored |
| White trace becomes arterial; motor current flattens | Whole pump in the aorta | P-2, recrossing under imaging, over a wire if needed |
| Low-amplitude signals; position unknown | Very poor contraction, ventricular fibrillation, or damped signal | Rhythm and native function assessed; position confirmed by echocardiography |
| Diastolic left ventricular pressure below −40 mm Hg, normal systolic contour | Diastolic suction: low preload or right ventricular failure | Volume status, right atrial pressure, pulmonary artery pulsatility index (PAPi), and right ventricle assessed |
| Continuously negative left ventricular pressure | Inlet against the wall or chordae | P-level reduced; pump repositioned |
| Rising estimated end-diastolic pressure at stable flow | Inadequate unloading: malposition, aortic regurgitation, rising native load, or pump dysfunction | Position, valve, and pump reassessed; escalation considered |
Interpretation of SmartAssist waveforms, after the US instructions for use and the 2026 Structural Heart review. Loss of pulsatility on the patient's arterial line has the same two explanations as flat motor current: displacement or very poor native contraction.
Pumps without SmartAssist show only the placement signal and motor current, and their fluid-filled placement lumen can damp and mimic a position change. Derived pressures support bedside judgment and do not replace a pulmonary artery catheter and echocardiography.
P-levels, flow, and cardiac power on the controller
At a fixed P-level, flow rises when the aortic-ventricular pressure difference falls (a contracting ventricle, lower afterload) and falls when it widens (vasoconstriction, a poorly contracting ventricle), so displayed flow changes without any change in setting. A high P-level in an underfilled ventricle produces suction without added flow. In AUTO the controller lowers speed when it detects suction and displays "Impella Flow Reduced"; if suction persists at the lowest speed it displays "Suction." BOOST maximizes flow for 5 minutes. When a measured cardiac output is entered, SmartAssist displays native output and cardiac power output, and the trend screen plots them over time. Rising native output at constant pump flow is one of the clearest signs of recovery. The National Cardiogenic Shock Initiative targeted cardiac power output above 0.6 W; the calculation is on the cardiogenic shock page.
Position and suction alarms
| Alarm | Common causes | Management |
|---|---|---|
| Impella Flow Reduced (advisory) | Suction detected in AUTO; speed lowered by the controller | Evaluated as suction; flow is restored once suction clears |
| Suction | Hypovolemia or bleeding; right ventricular failure or high pulmonary vascular resistance; inlet against the wall, malrotation, or deep position; tamponade; P-level above what the ventricle supplies | Position checked by imaging and the inlet moved slightly; volume for low filling pressures; inotropes, pulmonary vasodilators, or right-sided support for right ventricular failure; effusion drained; P-level reduced, then returned to the pre-alarm setting |
| Impella Position in Ventricle | Inward migration, often after sitting up or loss of slack | P-2, withdrawal under imaging |
| Impella Position in Aorta | Outward displacement from traction or movement | P-2, repositioning or recrossing under imaging |
| Impella Position Unknown | Low pulsatility from poor function, arrest, or arrhythmia; damped sensor | Rhythm and function assessed; echocardiography |
| Sudden low flow or suction at start-up | Thrombus in the inlet or pump | Catheter withdrawn and inspected per the instructions for use; pump replaced if flow does not recover |
Microaxial pump alarms, after the US instructions for use and the 2026 Structural Heart review. Alarm wording varies between software versions.
In diastolic suction, a pulmonary artery catheter separates hypovolemia (low wedge and right atrial pressures) from right ventricular failure (low or normal wedge pressure, high right atrial pressure, PAPi below about 1). In the National Cardiogenic Shock Initiative, diastolic suction alarms in the first 24 hours were more frequent with right ventricular dysfunction and served as an early marker of it.
Purge system management
Purge composition, heparin and bicarbonate solutions, and the heparin dose delivered by the purge are on the temporary mechanical circulatory support page; this section covers controller behavior and troubleshooting.
The controller adjusts purge flow to hold a target purge pressure. In the US instructions for use (2016 edition; the 2020 Impella CP with SmartAssist and Impella 5.5 with SmartAssist editions give the same range), set-up runs at 600 mm Hg and the standard configuration holds a pressure between 300 and 1,100 mm Hg at flows of 2 to 30 mL/h, with alarms below 300 or above 1,100 mm Hg. Purge flow falls as pressure rises, so a falling purge flow shifts more of the heparin requirement to the systemic infusion. Dextrose 5% in water with heparin 25 or 50 units/mL is the standard solution once support continues beyond the procedure; saline and direct thrombin inhibitors are not placed in the purge. Sodium bicarbonate, 25 mEq/L in 5% dextrose (the approved US labeling allows 25 or 50 mEq/L), is the labeled alternative when heparin is contraindicated, including heparin-induced thrombocytopenia, and whenever in the physician's judgment heparin should not be used. Viscosity sets pressure: the SCAI quality tip describes raising dextrose up to 40%, with 20% preferred, for a low-pressure alarm and returning to 5% for a high-pressure alarm.
| Purge finding | Causes | Response |
|---|---|---|
| High purge pressure | Kinked tubing or catheter; deposits narrowing the purge lumen; concentrated dextrose | Kinks corrected; dextrose reduced to 5%; persistent rise with falling purge flow treated as impending pump thrombosis |
| Low purge pressure | Leak or disconnection; damaged purge lumen; dilute solution | Connections and cassette checked; dextrose concentration increased; persistent loss of the barrier requires device exchange |
| Rising pressure, falling purge flow, rising lactate dehydrogenase | Pump thrombosis | Anticoagulation reviewed; pump exchanged if hemolysis or flow loss persists |
| Bleeding on support | Systemic plus purge heparin | Systemic heparin stopped with heparin kept in the purge; then bicarbonate purge; plain dextrose if bicarbonate is unavailable |
Purge troubleshooting, after the US instructions for use (2016 and 2020 editions), the 2022 SCAI quality improvement tip, and the 2021 multidisciplinary anticoagulation consensus.
Any change of purge flow or solution changes the heparin delivered, and the systemic infusion is retitrated after it. The SCAI quality tip cites a systemic anti-Xa target of 0.2 to 0.4 IU/mL or the equivalent aPTT; targets vary across centers.
Hemolysis on the microaxial pump
Red cells are damaged by shear in the cannula, contact with the impeller, and negative pressure at an obstructed inlet. In 40 patients supported for more than 6 hours in shock, cumulative incidence of hemolysis was 62.5% (Badiye 2016); a 2026 review reports hemolysis in at least one-quarter of supported patients. Plasma free hemoglobin is the best indicator in the instructions for use; no threshold is specific to microaxial pumps. The Mechanical Circulatory Support Academic Research Consortium defines hemolysis as plasma free hemoglobin above 20 mg/dL or lactate dehydrogenase above 2.5 times the upper limit of normal after the first 72 hours of support. Lactate dehydrogenase, bilirubin, haptoglobin, potassium, falling hemoglobin, and dark urine support the diagnosis, and daily sampling, more often after suction or position alarms, detects it early.
| Cause | Console and clinical clues | Correction |
|---|---|---|
| Inlet near the wall or chordae; wrong position | Suction, Impella Flow Reduced, or position alarms; unexpected flow | Pump rotated or moved under imaging; P-level reduced until repositioned |
| Higher flow setting than needed | Often no alarm; normal hemodynamics, native recovery | P-level reduced until pressure begins to fall, then raised slowly to the lowest adequate level |
| Inadequate filling | Suction alarms, low central venous and wedge pressures, bleeding | Volume, fluid balance, treatment of bleeding |
| Right ventricular failure | Suction alarms, high right atrial pressure with low wedge pressure, low PAPi | Inotropes, pulmonary vasodilators, right-sided support |
| Pump thrombosis | Rising purge pressure, falling purge flow | Anticoagulation reviewed; device exchanged |
| No correctable cause | Correct position and adequate filling | Device exchange, upsizing to a 5.5 that delivers the same flow at lower speed, or removal |
Causes and management of hemolysis, after Table 7.1 of the US instructions for use and the 2026 Structural Heart review. Patient causes such as sickle cell disease and thrombotic microangiopathy are considered when no device cause is found.
A lower P-level does not always stop hemolysis, because an underfilled ventricle continues to generate suction at low speed. Hemolysis on all devices is covered on the temporary mechanical circulatory support page.
Device-specific complications, infection, and mobilization
Vascular injury, bleeding, limb ischemia, acquired von Willebrand syndrome, and stroke are described on the temporary mechanical circulatory support page. Microaxial pumps add their own complications.
- Aortic valve injury. Crossing a calcified valve can tear a leaflet, and leaflets closing against the cannula are damaged during prolonged support; new regurgitation is seen below the valve.
- Ventricular perforation. Wire or pigtail perforation, especially through necrotic myocardium after repeated resuscitation, causes effusion and dislocation of the pump into the pericardium; a new effusion on support is treated as perforation until proved otherwise.
- Mitral apparatus injury. A deep pump entangles chordae, producing regurgitation, hemolysis, and difficult withdrawal.
- Ventricular arrhythmia. Inlet contact with the septum or papillary muscles triggers ectopy and ventricular tachycardia, and repositioning is part of treatment. The instructions for use specify reducing flow during cardiopulmonary resuscitation, avoiding contact with the catheter during defibrillation, and reassessing placement afterward.
- Pump thrombosis and embolism. Thrombus forms in the motor when the purge fails or anticoagulation is inadequate, and presents as hemolysis, rising purge pressure, falling flow, or embolic stroke.
- Infection. Insertion-site infection and sepsis are listed adverse events. Risk rises with duration, femoral access, and surgical grafts; femoral sites are inspected daily, axillary grafts are tunneled, and fever on support is evaluated with blood cultures and inspection of the access site and graft.
Mobilization. Femoral pumps confine the patient to bed with the leg straight, because hip flexion displaces the pump. Axillary pumps allow sitting, standing, and walking during weeks of support. In 30 patients on an Impella 5.5, 63.3% of 876 physiotherapy sessions included out-of-bed mobilization; no life-threatening events occurred (Hamazaki 2026). Protocols check position before and after sessions, secure catheter and console, limit abduction of the arm on the graft side, and return the patient to bed for echocardiography after a position alarm.
Early unloading protocols and the National Cardiogenic Shock Initiative
The National Cardiogenic Shock Initiative was a single-arm US study of a standardized protocol for infarct-related shock: rapid recognition, Impella placement ideally before PCI, right heart catheterization, and reassessment within 24 hours with escalation when perfusion or cardiac power remained inadequate. Criteria followed the SHOCK trial, and patients already on a balloon pump were excluded.
In the 2019 report (171 patients, 35 sites), survival to discharge was 72%. Creatinine of 2 mg/dL or more, lactate above 4 mmol/L, cardiac power output below 0.6 W, and age 70 years or more predicted death. The final report (406 patients, 80 sites, 2016–2020) included 17% with witnessed out-of-hospital arrest and 9% under active resuscitation during implantation. By 24 hours lactate fell from 4.8 to 2.7 mmol/L, and cardiac power output rose from 0.67 to 1.0 W; survival was 71% to discharge and 53% at 1 year. Right ventricular dysfunction (right atrial pressure above 12 mm Hg with PAPi below 1) affected 28% and lowered survival to discharge (61.8% against 73.4%).
The initiative showed that early support with invasive hemodynamics and defined targets is feasible across hospitals, but it was not randomized. DanGer Shock then randomized 360 patients with STEMI-related shock and ejection fraction below 45%, excluding comatose arrest survivors. Death at 180 days was 45.8% with an Impella CP and 58.5% with standard care, with more bleeding, limb ischemia, hemolysis, and renal replacement therapy (details on the temporary mechanical circulatory support page). The 2025 ACC concise clinical guidance discourages routine support in all shock and supports a microaxial pump in selected STEMI-related shock with predominantly left ventricular failure.
Protected high-risk PCI: patient selection
Protected PCI is elective or urgent PCI with hemodynamic support in a patient without shock whose anatomy and ventricular function predict collapse during balloon inflation, atherectomy, or a complication. The labeled indication requires a heart team, including a cardiac surgeon, to judge high-risk PCI the appropriate option, which in practice means bypass surgery has been declined or judged prohibitive.
| Domain | Features used to identify high-risk PCI |
|---|---|
| Ventricular function | Ejection fraction of 35% or less; trial thresholds were 30% or less (BCIS-1; PROTECT II three-vessel stratum), 35% or less, or 45% or less with severe mitral regurgitation (CHIP-BCIS3), and 40% or less (PROTECT IV) |
| Myocardium at risk | Unprotected left main disease; last patent vessel or conduit; three-vessel disease; BCIS-1 jeopardy score of 8 or more of 12 |
| Lesion complexity | Heavy calcification needing atherectomy; chronic total occlusion with retrograde techniques; bifurcation left main disease; diffuse disease |
| Hemodynamics | Elevated end-diastolic or wedge pressure; low cardiac index; pulmonary hypertension; recent decompensation |
| Clinical risk | Older age, kidney disease, recent acute coronary syndrome, prior bypass surgery, high Society of Thoracic Surgeons (STS) score, surgical turndown |
| Access | Iliofemoral anatomy adequate for a 14 Fr sheath, or an alternative route |
Features used in trials and consensus documents to define complex high-risk PCI.
Scores predict risk and do not predict benefit. In PROTECT III, 30-day major adverse events after Impella-supported PCI were 5.1%, 10.1%, and 17.6% with STS predicted mortality below 4%, 4% to 8%, and above 8%. The STS score underestimated mortality (observed-to-expected ratio 1.23; Falah 2026). The 2021 revascularization guideline (Class 2b) and the 2015 multisociety consensus statement both predate CHIP-BCIS3.
Randomized and registry evidence for protected PCI
| Study | Design and population | Main result |
|---|---|---|
| BCIS-1 (2010; long-term 2013) | 301 patients, ejection fraction 30% or less and jeopardy score 8 or more; elective balloon pump against no planned support | MACCE at discharge 15.2% vs 16.0% (P = 0.85); major procedural complications 1.3% vs 10.7%; at a median of 51 months, 42 vs 58 deaths (HR 0.66; 95% CI 0.44–0.98) |
| PROTECT II (2012) | 452 patients with unprotected left main or last patent vessel and ejection fraction 35% or less, or three-vessel disease and ejection fraction 30% or less; Impella 2.5 against balloon pump | 30-day major adverse events 35.1% vs 40.1% (P = 0.227); 90 days 40.6% vs 49.3% (P = 0.066); stopped early for futility |
| PROTECT III (2017–2020) | Prospective single-arm study of Impella-supported PCI; 504 patients met PROTECT II criteria | 90-day MACCE 15.1% vs 21.9% in the PROTECT II Impella arm; bleeding requiring transfusion 1.8% vs 9.3% |
| CHIP-BCIS3 (2026) | 300 patients at UK centers, severe left ventricular dysfunction and extensive coronary disease; elective microaxial pump against standard care during complex PCI | Win ratio 0.85 (95% CI 0.63–1.15; P = 0.30); death 47 vs 33 patients (HR 1.54; 95% CI 0.99–2.41) |
| PROTECT IV (NCT04763200) | Planned 1,252 patients, complex disease, ejection fraction 40% or less, excessive surgical risk; Impella CP against PCI with or without balloon pump | 3-year composite primary end point (ClinicalTrials.gov registration) |
Randomized trials and the principal registry of hemodynamic support in high-risk PCI. MACCE, major adverse cardiac and cerebrovascular events; HR, hazard ratio.
BCIS-1 and PROTECT II. Elective balloon support did not reduce major adverse events at discharge, although it reduced procedural complications, mainly hypotension; the long-term mortality difference was hypothesis-generating. In PROTECT II the Impella 2.5 gave better hemodynamic support, but adverse events did not differ by intention to treat, and the 90-day difference was significant only per protocol and in a three-vessel substudy (39.5% against 51.0%; P = 0.039). Secondary analyses of a neutral trial do not establish benefit.
PROTECT III. Compared with the PROTECT II Impella arm, PROTECT III patients had more atherectomy (37.1% against 14.8%) and less procedural hypotension (2.2% against 10.1%); after propensity matching, 90-day MACCE was 10.4% and 16.9% (P = 0.048). A pooled win-ratio analysis favored Impella over the balloon pump (win ratio 1.69; Panoulas 2026). These are nonrandomized, manufacturer-supported comparisons across eras of practice.
CHIP-BCIS3. The first randomized trial of a microaxial pump against unsupported PCI enrolled 300 patients with an ejection fraction of 35% or less (45% or less with severe mitral regurgitation), a BCIS-1 jeopardy score of 8 or more, and planned complex PCI: left main intervention, calcium modification, or a chronic total occlusion treated by a retrograde approach. The primary outcome was a hierarchical composite of death, disabling stroke, spontaneous myocardial infarction, cardiovascular hospitalization, and periprocedural myocardial injury at a minimum of 12 months. At a median of 22 months the pump did not improve the outcome, deaths were numerically more frequent with the pump, and bleeding and vascular complications did not differ materially. PROTECT IV, with a 3-year composite including death, stroke, myocardial infarction, revascularization, durable LVAD or transplantation, and cardiovascular hospitalization, compares routine support with PCI with or without a balloon pump (ClinicalTrials.gov NCT04763200).
BCIS-1, PROTECT II, and CHIP-BCIS3 were neutral, and CHIP-BCIS3 showed a possible mortality excess, so routine prophylactic support is not supported. It remains a selective heart-team decision for patients in whom transient ischemia is expected to cause collapse that standby support could not rescue, weighed against large-bore access risk.
Procedure in protected PCI. Baseline right heart catheterization, when performed, records filling pressures and output and shows whether the right ventricle can fill a left-sided pump. The pump is placed before the first balloon inflation or atherectomy, with preclosure sutures in place. At the end, a stable patient has the pump removed in the catheterization laboratory with the sutures tied and limb perfusion confirmed; for manual compression the instructions for use specify an activated clotting time below 150 seconds before the introducer is removed. A patient who becomes unstable stays on support through the repositioning sheath, and the indication becomes cardiogenic shock with its own labeled duration.
TandemHeart: circuit, migration, and residual atrial shunt
The TandemHeart is a percutaneous left atrial–to–femoral arterial bypass. A 21 Fr inflow cannula passes from the femoral vein across the interatrial septum by transseptal puncture into the left atrium; an extracorporeal centrifugal pump returns blood through a 15 to 19 Fr femoral (or axillary) arterial cannula at about 3 to 5 L/min, depending on cannula size and speed. It lowers left atrial and left ventricular filling pressures without crossing the aortic valve, so it can be used with aortic stenosis or a mechanical aortic valve; unloading is indirect, through the atrium. An oxygenator can be added, and the same left atrial drainage vents the left ventricle during venoarterial ECMO. In 117 patients in refractory shock (47.9% during active resuscitation), cardiac index rose from 0.52 to 3.0 L/min/m² and 30-day mortality was 40.2% (Kar 2011). The randomized trials against the balloon pump are summarized on the temporary mechanical circulatory support page.
Cannula migration. The inflow cannula is held in the left atrium only by its position across the septum and can be pulled back into the right atrium by hip flexion, turning, or traction. The pump then returns desaturated right atrial blood to the aorta, producing systemic desaturation, and the left atrium is no longer drained, so wedge pressure rises. Sudden desaturation with falling pump flow or rising wedge pressure is treated as migration until imaging shows otherwise. Position is confirmed daily by echocardiography or fluoroscopy, and patients are moved with the leg straight and the cannula secured. Migration into a pulmonary vein or against the atrial wall causes suction and low flow.
Residual atrial septal defect. Removal leaves an iatrogenic defect at the transseptal site, listed among device complications. Most such defects shunt left to right, are hemodynamically trivial, and close spontaneously. A defect matters when high right atrial pressure, right ventricular failure, or left-sided unloading by a durable LVAD reverses the shunt and causes hypoxemia, or when paradoxical embolism occurs. It is assessed by echocardiography with color Doppler and saline contrast, and a defect causing hypoxemia from right-to-left shunting is closed percutaneously with a septal occluder.
Other TandemHeart complications. Bleeding, limb ischemia, and venous thrombosis from the large cannulas, circuit thromboembolism, hemolysis, and perforation or tamponade from the transseptal puncture also occur. Ventricular septal rupture is a contraindication because lowering left-sided pressure increases right-to-left shunting. Weaning reduces pump flow stepwise while wedge pressure and output are followed, and removal includes arterial closure and assessment of the septum.
Right-sided percutaneous support: Impella RP and RP Flex
The Impella RP is placed from the femoral vein across the tricuspid and pulmonary valves, with its inlet in the inferior vena cava near the right atrium and its outlet in the pulmonary artery. It uses a 22 to 23 Fr venous sheath and delivers about 4 L/min at up to 33,000 revolutions per minute. The RP Flex, approved in the United States on March 31, 2023, has a flexible catheter for internal jugular insertion, which frees the groin, and carries SmartAssist sensors. The RP received a humanitarian device exemption in 2015 on the basis of RECOVER RIGHT and later received premarket approval.
The labeled indication and duration are in the labeling table above. Labeled contraindications are disorders of the pulmonary artery wall precluding placement; mechanical valves or severe stenosis or regurgitation of the tricuspid or pulmonary valve; mural thrombus of the right atrium or vena cava; anatomy precluding insertion; and a vena cava filter or interruption device unless there is clear passage for a 22 Fr catheter. Labeled adverse events include tricuspid and pulmonary valve injury, pulmonary insufficiency, arrhythmia, venous thrombosis, and phlegmasia cerulea dolens.
In RECOVER RIGHT, 30 patients with refractory right ventricular failure received the RP; cardiac index rose from 1.8 to 3.3 L/min/m², central venous pressure fell from 19.2 to 12.6 mm Hg, and 30-day survival was 73.3%. The outlet is positioned in the main pulmonary artery, clear of the valve leaflets and short of the bifurcation. Right-sided flow is titrated against the wedge pressure, because pumping more blood across the lungs than the left ventricle can clear causes pulmonary edema. The ProtekDuo, a 29 or 31 Fr dual-lumen jugular cannula with an extracorporeal pump at about 4 to 5 L/min, is used when an oxygenator is needed. Indications, PAPi, and biventricular configurations are on the temporary mechanical circulatory support page.
Other percutaneous devices: iVAC 2L, HeartMate PHP, and Reitan catheter pump
| Device | Design | Status |
|---|---|---|
| iVAC 2L (PulseCath) | Pulsatile extracorporeal membrane pump driven by a standard balloon-pump console, connected to a large-bore transfemoral catheter across the aortic valve; fills from the ventricle and ejects into the aorta in phases timed by the console | CE-marked in Europe since 2014, recertified under the European Medical Device Regulation in 2024; used mainly in high-risk PCI |
| HeartMate PHP | Self-expanding 24 Fr impeller and cannula delivered across the aortic valve in collapsed form | Discontinued; SHIELD II, a randomized comparison with Impella in high-risk PCI, was halted in 2017 after device malfunctions |
| Reitan catheter pump | Foldable propeller in the descending thoracic aorta that lowers afterload above it and raises renal perfusion pressure below it; 10 Fr version | Investigational; studied in low-output heart failure with cardiorenal syndrome |
Percutaneous ventricular assist devices outside the Impella and TandemHeart platforms.
Among 75 roll-in patients of SHIELD II, maximum HeartMate PHP flow exceeded 3.5 L/min in only 26%, and 6.7% had device malfunctions; the expandable concept continues in the Impella ECP. The Reitan pump does not unload the ventricle; it targets decongestion and renal perfusion. None of these devices is part of US practice.
Weaning and removal of microaxial pumps
The general framework and shared criteria are on the temporary mechanical circulatory support page. Weaning begins when the cause has been treated, vasoactive support is low, filling pressures are acceptable, lactate has normalized, and end-organ function is improving. Rising native output on the trend screen, a wider arterial pulse pressure, and more pulsatile placement signal and motor current show a ventricle ejecting more. An advanced heart failure team first decides whether recovery, durable LVAD, or transplantation is the realistic goal.
Stepwise reduction. The instructions for use describe reducing the P-level in two-level steps as cardiac function allows, keeping it at P-2 or above while the pump remains across the valve. After a step, arterial pressure, filling pressures, output and power, lactate, urine output, and echocardiographic ventricular size and outflow velocity-time integral are reassessed. Patients stable at minimal support for the period set by local protocol proceed to removal. Weaning in heart failure–related shock is slower than after infarction, and some patients transition to inotropes.
Removal. The P-level is reduced to P-2 and the pump is pulled back into the aorta; only then is it set to P-0 and withdrawn, through the introducer or with the repositioning sheath. The arteriotomy is closed with preclose sutures, a large-bore closure device, contralateral balloon-assisted hemostasis, or manual compression after the activated clotting time falls below 150 seconds. The instructions for use specify 40 minutes of compression or the local protocol. Axillary 5.5 and 5.0 pumps are removed in the operating room with graft ligation or oversewing. The Impella RP is weaned by reducing flow while right atrial pressure, cardiac index, PAPi, and right ventricular size are followed, often with inhaled pulmonary vasodilators and inotropes.
Failure to wean. Recurrent hypotension, rising filling pressures, falling output, or rising lactate returns the P-level to the previous effective setting. A patient who cannot be weaned within the labeled duration of a CP is assessed for upsizing to a 5.5, durable LVAD, transplantation, or comfort-focused care, as described on the temporary mechanical circulatory support page.
Sources of error in microaxial and percutaneous support
- Reading displayed Impella flow as cardiac output. The console estimates pump flow; with malposition or an obstructed inlet the estimate is wrong, and total output requires native output to be measured.
- Treating suction by raising the P-level. Suction means the inlet is starved or obstructed, and higher speed worsens hemolysis; volume, the right ventricle, position, and tamponade are assessed first.
- Accepting correct depth with a malrotated inlet, which causes suction and incomplete unloading at the right distance from the annulus.
- Omitting the purge from heparin calculations and from bleeding and heparin-induced thrombocytopenia evaluations, or placing saline or a direct thrombin inhibitor in the purge.
- Attributing hemolysis to the device without examining the right ventricle and volume status; suction-related hemolysis persists after a P-level reduction until filling is corrected.
- Treating TandemHeart desaturation as respiratory failure before excluding inflow cannula migration into the right atrium.
- Running right-sided support at a flow the left ventricle cannot clear, which causes pulmonary edema; right- and left-sided flows are titrated together against wedge and right atrial pressures.
References
- Rao SV, O'Donoghue ML, Ruel M, et al. 2025 ACC/AHA/ACEP/NAEMSP/SCAI guideline for the management of patients with acute coronary syndromes: a report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice Guidelines. J Am Coll Cardiol. 2025;85:2135–2237. doi:10.1016/j.jacc.2024.11.009
- Sinha SS, Morrow DA, Kapur NK, Kataria R, Roswell RO. 2025 Concise clinical guidance: an ACC expert consensus statement on the evaluation and management of cardiogenic shock. J Am Coll Cardiol. 2025;85:1618–1641. doi:10.1016/j.jacc.2025.02.018
- Lawton JS, Tamis-Holland JE, Bangalore S, et al. 2021 ACC/AHA/SCAI guideline for coronary artery revascularization: a report of the American College of Cardiology/American Heart Association Joint Committee on Clinical Practice Guidelines. J Am Coll Cardiol. 2022;79:e21–e129. doi:10.1016/j.jacc.2021.09.006
- Rihal CS, Naidu SS, Givertz MM, et al. 2015 SCAI/ACC/HFSA/STS clinical expert consensus statement on the use of percutaneous mechanical circulatory support devices in cardiovascular care. J Am Coll Cardiol. 2015;65:e7–e26. doi:10.1016/j.jacc.2015.03.036
- Estep JD, Nicoara A, Cavalcante J, et al. Recommendations for multimodality imaging of patients with left ventricular assist devices and temporary mechanical support: updated recommendations from the American Society of Echocardiography. J Am Soc Echocardiogr. 2024;37:820–871. doi:10.1016/j.echo.2024.06.005
- Abiomed. Impella 2.5, Impella 5.0, Impella LD, and Impella CP (shock); Impella 2.5 and Impella CP (HRPCI): instructions for use and clinical reference manual (United States). US Food and Drug Administration premarket approval P140003/S018 labeling; 2016. accessdata.fda.gov
- Abiomed (Johnson & Johnson MedTech). Impella left-side and right-side devices: indications for use and important risk information (United States). Accessed November 10, 2026. heartrecovery.com
- US Food and Drug Administration. Premarket approval P140003/S135 (Impella 2.5, CP, CP with SmartAssist, 5.0, 5.5 with SmartAssist, and LD systems): modification of the instructions for use to allow sodium bicarbonate in the purge fluid. Decision date February 6, 2026; earlier approval P140003/S088, April 19, 2022. accessdata.fda.gov
- De Filippo O, Pagliassotto I, Braia V, et al. Axial flow percutaneous ventricular assist devices: physiology, procedural considerations, and clinical applications. A state-of-the-art review. Struct Heart. 2026;10:100773. doi:10.1016/j.shj.2025.100773
- Upadhyay R, Alrayes H, Arno S, Kaushik M, Basir MB. Current landscape of temporary percutaneous mechanical circulatory support technology. US Cardiol. 2021;15:e21. doi:10.15420/usc.2021.15
- Ferrera D, Beavers CJ, Latif F. Impella anticoagulation management in the CCU. Society for Cardiovascular Angiography & Interventions quality improvement tip; April 14, 2022. scai.org
- Beavers CJ, DiDomenico RJ, Dunn SP, et al. Optimizing anticoagulation for patients receiving Impella support. Pharmacotherapy. 2021;41:932–942. doi:10.1002/phar.2629
- Badiye AP, Hernandez GA, Novoa I, Chaparro SV. Incidence of hemolysis in patients with cardiogenic shock treated with Impella percutaneous left ventricular assist device. ASAIO J. 2016;62:11–14. doi:10.1097/MAT.0000000000000290
- Møller JE, Engstrøm T, Jensen LO, et al. Microaxial flow pump or standard care in infarct-related cardiogenic shock (DanGer Shock). N Engl J Med. 2024;390:1382–1393. doi:10.1056/NEJMoa2312572
- Kapur NK, Alkhouli MA, DeMartini TJ, et al. Unloading the left ventricle before reperfusion in patients with anterior ST-segment-elevation myocardial infarction. Circulation. 2019;139:337–346. doi:10.1161/CIRCULATIONAHA.118.038269
- Kapur NK, Mangner N, Aghili N, et al. Left ventricular unloading in anterior ST-segment elevation myocardial infarction without shock: the ST-Segment Elevation Myocardial Infarction Door to Unload randomized controlled trial. J Am Coll Cardiol. 2026;88:76–90. doi:10.1016/j.jacc.2026.03.071
- Ito T, Saku K, Ohba K, et al. Impact of microaxial flow pump on coronary flow dynamics in normal hearts: an experimental evaluation. J Cardiovasc Transl Res. 2026;19:93. doi:10.1007/s12265-026-10818-z
- Basir MB, Kapur NK, Patel K, et al. Improved outcomes associated with the use of shock protocols: updates from the National Cardiogenic Shock Initiative. Catheter Cardiovasc Interv. 2019;93:1173–1183. doi:10.1002/ccd.28307
- Basir MB, Lemor A, Gorgis S, et al. Early utilization of mechanical circulatory support in acute myocardial infarction complicated by cardiogenic shock: the National Cardiogenic Shock Initiative. J Am Heart Assoc. 2023;12:e031401. doi:10.1161/JAHA.123.031401
- Gorgis S, Gupta K, Lemor A, et al. Impact of right ventricular dysfunction on outcomes in acute myocardial infarction and cardiogenic shock: insights from the National Cardiogenic Shock Initiative. J Card Fail. 2024;30:1275–1284. doi:10.1016/j.cardfail.2024.07.015
- Perera D, Stables R, Thomas M, et al. Elective intra-aortic balloon counterpulsation during high-risk percutaneous coronary intervention: a randomized controlled trial (BCIS-1). JAMA. 2010;304:867–874. doi:10.1001/jama.2010.1190
- Perera D, Stables R, Clayton T, et al. Long-term mortality data from the balloon pump-assisted coronary intervention study (BCIS-1): a randomized, controlled trial of elective balloon counterpulsation during high-risk percutaneous coronary intervention. Circulation. 2013;127:207–212. doi:10.1161/CIRCULATIONAHA.112.132209
- O'Neill WW, Kleiman NS, Moses J, et al. A prospective, randomized clinical trial of hemodynamic support with Impella 2.5 versus intra-aortic balloon pump in patients undergoing high-risk percutaneous coronary intervention: the PROTECT II study. Circulation. 2012;126:1717–1727. doi:10.1161/CIRCULATIONAHA.112.098194
- Kovacic JC, Kini A, Banerjee S, et al. Patients with 3-vessel coronary artery disease and impaired ventricular function undergoing PCI with Impella 2.5 hemodynamic support have improved 90-day outcomes compared to intra-aortic balloon pump: a sub-study of the PROTECT II trial. J Interv Cardiol. 2015;28:32–40. doi:10.1111/joic.12166
- O'Neill WW, Anderson M, Burkhoff D, et al. Improved outcomes in patients with severely depressed LVEF undergoing percutaneous coronary intervention with contemporary practices. Am Heart J. 2022;248:139–149. doi:10.1016/j.ahj.2022.02.006
- Falah B, Thompson J, Shah Y, et al. Performance of the STS risk score in predicting outcomes in Impella-supported high-risk percutaneous coronary intervention: insights from the PROTECT III study. J Soc Cardiovasc Angiogr Interv. 2026;5:105448. doi:10.1016/j.jscai.2026.105448
- Panoulas V, Schreiber T, Tsintzos SI, et al. Impella protected percutaneous coronary intervention outcomes compared with intra-aortic balloon pump: a contemporary view. Am J Cardiol. 2026;269:140–146. doi:10.1016/j.amjcard.2026.03.060
- Perera D, Ryan M, Ezad SM, et al. Left ventricular unloading in high-risk percutaneous coronary intervention (CHIP-BCIS3). N Engl J Med. 2026;394:1779–1789. doi:10.1056/NEJMoa2515704
- Mangner N, Sharma SK, O'Connor C, et al. Mechanical circulatory support in high-risk elective PCI: rationale and design of the PROTECT IV trial. EuroIntervention. 2025;21:e1244–e1258. doi:10.4244/EIJ-D-25-00486
- Johnson & Johnson MedTech. Impella ECP pivotal study demonstrates safety and efficacy for use in high-risk PCI (press release, Transcatheter Cardiovascular Therapeutics 2024). October 2024. jnj.com
- Ramzy D, Anderson M, Batsides G, et al. Early outcomes of the first 200 US patients treated with Impella 5.5: a novel temporary left ventricular assist device. Innovations (Phila). 2021;16:365–372. doi:10.1177/15569845211013329
- Hamazaki N, Ikeda Y, Ichikawa T, et al. Early mobilization trajectories during Impella 5.5 support and associated outcomes in patients with cardiogenic shock: a physiotherapy session-level analysis. Heart Lung. 2026;80:102928. doi:10.1016/j.hrtlng.2026.102928
- Mohammed M, Al-Darzi W, Jabri A, et al. Safety and feasibility of transcaval access for the delivery of Impella microaxial flow pump. J Soc Cardiovasc Angiogr Interv. 2025;4:103789. doi:10.1016/j.jscai.2025.103789
- Murakami K, Hasegawa S, Miki A, et al. Impact of initial axillary microaxial flow pump on post-myocardial infarction ventricular septal defect management. Interdiscip Cardiovasc Thorac Surg. 2025;40:ivaf253. doi:10.1093/icvts/ivaf253
- Kar B, Gregoric ID, Basra SS, Idelchik GM, Loyalka P. The percutaneous ventricular assist device in severe refractory cardiogenic shock. J Am Coll Cardiol. 2011;57:688–696. doi:10.1016/j.jacc.2010.08.613
- Anderson MB, Goldstein J, Milano C, et al. Benefits of a novel percutaneous ventricular assist device for right heart failure: the prospective RECOVER RIGHT study of the Impella RP device. J Heart Lung Transplant. 2015;34:1549–1560. doi:10.1016/j.healun.2015.08.018
- Kapur NK, Jorde UP, Sharma S, et al. Early experience with the HeartMate percutaneous heart pump from the SHIELD II trial. ASAIO J. 2022;68:492–497. doi:10.1097/MAT.0000000000001517
- Van Mieghem NM, Daemen J, Lenzen MJ, et al. The PulseCath iVAC 2L left ventricular assist device: conversion to a percutaneous transfemoral approach. EuroIntervention. 2015;11:835–839. doi:10.4244/EIJV11I7A168
- Napp LC, Mariani S, Ruhparwar A, et al. First-in-man use of the percutaneous 10F Reitan catheter pump for cardiorenal syndrome. ASAIO J. 2022;68:e99–e101. doi:10.1097/MAT.0000000000001498
- Kormos RL, Antonides CFJ, Goldstein DJ, et al. Updated definitions of adverse events for trials and registries of mechanical circulatory support: a consensus statement of the Mechanical Circulatory Support Academic Research Consortium. J Heart Lung Transplant. 2020;39:735–750. doi:10.1016/j.healun.2020.03.010
- Abiomed. Impella CP with SmartAssist: instructions for use and clinical reference manual (United States). 2020. fda.gov
- Abiomed. Impella 5.5 with SmartAssist: instructions for use and clinical reference manual (United States). 2020. fda.gov
- ClinicalTrials.gov. Impella-supported PCI in high-risk patients with complex coronary artery disease and reduced left ventricular function (PROTECT IV). NCT04763200. clinicaltrials.gov/study/NCT04763200