CardioBoardExam.com

Library › Imaging › Radiation safety in the catheterization laboratory

Radiation safety and x-ray imaging in the catheterization laboratory

X-ray imaging and radiation protection in adult cardiac and peripheral catheterization procedures, from the imaging chain and dose metrics through patient skin injury, operator and staff protection, pregnancy, low-fluoroscopy techniques, and contrast in peripheral work.

Fluoroscopically guided cardiovascular procedures deliver the largest radiation doses in cardiology, to the patient's skin and to the staff at the table. Reference air kerma predicts skin injury, kerma-area product tracks total energy delivered and stochastic risk, and fluoroscopy time measures neither well. Skin reactions begin above a peak skin dose of about 2 Gy. A reference air kerma of 5 Gy, a kerma-area product of 500 Gy·cm², or 60 min of fluoroscopy triggers patient follow-up, and a peak skin dose of 15 Gy to one field is a Joint Commission sentinel event. Low pulse and frame rates, collimation, a high table with the detector close to the patient, shallow projections, and fluoroscopy store reduce dose to patient and operator together.

The dose from a fluoroscopic procedure is set during the procedure by the physician at the table. It depends on how long the beam is on, the rate and type of acquisition, the geometry of the tube, patient, and detector, and the size of the patient. Patient dose and operator dose move together, because nearly all occupational exposure is radiation scattered from the patient.

The general principles of radiation protection (justification, optimization, and dose limitation), radiation biology, dose quantities and units, regulatory dose limits, personnel monitoring, and the inverse-square law are covered in radiation safety in nuclear cardiology. The governing US document is the 2018 ACC/HRS/NASCI/SCAI/SCCT expert consensus document on optimal use of ionizing radiation in cardiovascular imaging (Hirshfeld 2018). It is supplemented by the Society for Cardiovascular Angiography and Interventions (SCAI) radiation safety program statement (Chambers 2011), the SCAI consensus on the pregnant cardiologist (Best 2011), and the SCAI/ASE/HRS/SIR/SVS statement on enhanced radiation protection (Rizik 2026).

Two operators in sterile gowns, caps, and thyroid collars, seen from behind, watching a live fluoroscopic image of catheters in the chest on a large monitor beside hemodynamic tracings, under a ceiling-mounted boom.
Operators watching the live fluoroscopic image during cardiac catheterization. Thyroid collars are visible above the gowns, the image and hemodynamic tracings share one large display, and numeric procedure displays appear at the left edge of the image. Credit: Navy Medicine, Wikimedia Commons; public domain (US government work).

The fluoroscopic imaging chain from x-ray tube to flat-panel detector

A cardiac angiographic system mounts an x-ray tube and an image receptor at opposite ends of a C-arm that rotates around the patient. In standard cardiac work the tube is under the table and the detector above the patient, so the beam enters the patient's back and exits the chest. Biplane systems carry a second, lateral C-arm. The components between the generator and the display determine both image quality and dose.

ComponentFunctionEffect on dose
GeneratorSupplies tube voltage (kV) and tube current (mA) and produces short x-ray pulsesOutput follows the automatic dose-rate control; pulse width governs motion blur
X-ray tubeProduces a polychromatic x-ray spectrum at the focal spotSource of all radiation in the room
Spectral filters (copper, aluminum)Remove low-energy photons that would be absorbed in the skin without reaching the detectorLower skin dose per image; heavier filtration in fluoroscopy, lighter in cine
Collimator and wedge filtersLead shutters restrict the field; semitransparent wedges attenuate bright regions such as lungSmaller field lowers kerma-area product and scatter; wedges even out detector exposure
PatientAttenuates and scatters the beamBody thickness drives tube output; the irradiated volume is the source of scatter to staff
Antiscatter gridLead strips focused on the source absorb obliquely scattered photons before they reach the detectorImproves contrast; raises the output needed for a given detector signal
Flat-panel detectorCesium iodide scintillator over an amorphous silicon photodiode array converts x-rays to a digital imageDetector dose per frame is set by the system; field-of-view selection changes dose rate
Image processing and displayTemporal and spatial noise filtering, edge enhancement, last-image hold, and storageNoise reduction allows lower dose per frame

Components of the fluoroscopic imaging chain and their effect on patient dose. X-ray production, tube voltage and current, and filtration are described in more detail for CT under x-ray production, tube voltage, tube current, and filtration.

Flat-panel detectors replaced image intensifiers in cardiac systems. They have no geometric distortion and no veiling glare, a wide dynamic range, and a rectangular field. Detector technology and image processing have a large effect on dose. One center replaced a system that used conventional image processing with one combining real-time noise reduction and an optimized acquisition chain. The median dose-area product fell by about two-thirds, from 155 to 52 Gy·cm² for diagnostic angiography and from 229 to 86 Gy·cm² for percutaneous coronary intervention (PCI) (Nakamura 2016). A similar system change halved the kerma-area product of transcatheter aortic valve replacement (TAVR), from a median of 153 to 78 Gy·cm² (Maccagni 2018). Equipment age and configuration are therefore dose determinants, and protocol settings are reviewed with the medical physicist when a system is installed.

Automatic dose-rate control, tube voltage, and tube current

Automatic dose-rate control (automatic exposure rate control in the federal standard) holds the signal reaching the detector constant. When the beam must cross more tissue, because the patient is larger, the projection is steeper, or more of the field passes through the spine and diaphragm, the system raises tube current, pulse width, and tube voltage, in an order defined by the manufacturer's dose curve. The skin dose rate at the beam entrance rises steeply as a result, most in steep left anterior oblique (LAO) and angulated projections (Kuon 2004).

Raising tube voltage increases beam penetration and lowers skin dose for a given detector signal but reduces iodine contrast, because the attenuation of iodine falls as photon energy moves away from its absorption edge. Raising tube current keeps contrast and raises dose in proportion. Dose curves balance these two settings, and most systems offer several fluoroscopy dose levels (labeled, for example, low, normal, and high) with different doses per pulse. Selecting the lowest fluoroscopy level that shows the devices in use is a standard first step in dose reduction.

US regulation limits fluoroscopic output. Equipment with automatic exposure rate control cannot exceed an air kerma rate of 88 mGy per minute at the measurement point in normal operation, and an optional high-level control, which must be activated by the operator, cannot exceed 176 mGy per minute (21 CFR 1020.32). Recording of images that are kept after the exposure (cine and digital acquisition) is exempt from these limits, which is one reason acquisition doses are much higher than fluoroscopy doses. The skin dose rate from fluoroscopy is typically 0.02 to 0.05 Gy per minute and can range from 0.01 to more than 0.5 Gy per minute depending on mode and patient size (FDA 1994).

Pulsed fluoroscopy, pulse rate, and low-dose protocols

Contemporary systems produce fluoroscopy as short pulses synchronized to detector readout. Pulse width (a few milliseconds) governs motion blur; pulse rate governs temporal smoothness. If the dose per pulse is constant, halving the pulse rate halves the fluoroscopy dose rate. Some systems raise the dose per pulse at lower rates to preserve image quality, so the realized dose reduction is smaller and depends on configuration.

In a randomized trial of 363 transradial procedures, both groups used a fixed dose of 40 nGy per pulse. Fluoroscopy at 7.5 frames per second, compared with 15, reduced operator dose by 30% and patient dose-area product by 19%, with no difference in fluoroscopy time (Abdelaal 2014). In TAVR on the same system, moving fluoroscopy and cine from 15 to 7.5 frames per second lowered the median kerma-area product from 184 to 107 Gy·cm² and the reference air kerma from 0.97 to 0.64 Gy (Maccagni 2018). In chronic total occlusion (CTO) PCI, an ultra-low-dose protocol of 3.75 frames per second for fluoroscopy and 7.5 frames per second for cine roughly halved kerma-area product in both simple and complex lesions. It did so without loss of procedural success, increase in in-hospital events, or crossover for poor image quality (Bacci 2023).

Low-rate fluoroscopy serves as the default in many laboratories, with higher rates selected for tasks that require them, such as fine device positioning at high heart rates. The determinants of fluoroscopic dose and the techniques that reduce it are set out in the 2018 consensus document (Hirshfeld 2018).

Cine acquisition, fluoroscopy store, and last-image hold

Cine (digital acquisition) uses a much higher dose per frame than fluoroscopy to produce low-noise images for diagnosis and archiving, and its contribution is included in any skin-dose assessment (FDA 1994). In coronary procedures, cine contributes most of the kerma-area product (Zanca 2020). An additional cine run therefore adds more dose than a similar time of fluoroscopy. The most common avoidable acquisition doses come from runs that answer no new question, repeated runs for marginal improvements, and long runs that continue after the contrast washes out.

Fluoroscopy store saves the most recent fluoroscopic sequence to the archive. It documents wire position, balloon inflation, stent expansion, and device deployment at fluoroscopic dose, and it replaces cine whenever diagnostic image quality is not needed. Last-image hold displays the final fluoroscopic frame after the pedal is released, so the operator can study a position without further exposure. Federal rules require last-image hold on current fluoroscopic equipment and require the display to show clearly whether the image is a held image or live fluoroscopy (21 CFR 1020.32). Virtual collimation and table panning on the held image allow the field to be adjusted before the beam is turned on again.

Image quality, quantum noise, and dose

The quality of a fluoroscopic image is limited by quantum noise: the random variation in the number of photons reaching a detector element. Noise falls with the square root of detector dose, so halving noise requires four times the dose. A grainy fluoroscopic image that shows wires, markers, and device edges is adequate for most of a procedure; a diagnostic cine run needs lower noise. Temporal filtering (averaging successive frames) and spatial filtering reduce visible noise but introduce lag and edge blurring, which is why fast-moving catheters can leave a faint trail at low doses.

Dose reduction is limited by the clinical task. The FDA advisory on skin injury states the aim as clinically adequate images with minimum fluoroscopic exposure (FDA 1994). A low pulse rate did not lengthen fluoroscopy time in a randomized trial (Abdelaal 2014), and an ultra-low-dose protocol did not lower procedural success in a cohort study (Bacci 2023). An image too noisy to show a dissection or a device edge leads to longer procedures and repeated runs, which add dose.

Geometry: table height, detector position, magnification, and collimation

Geometry is the second large group of dose determinants after beam-on time. Its effects follow from the inverse-square law and the behavior of automatic dose-rate control.

Table height and source-to-skin distance. The x-ray tube is a point source, so the dose rate at the skin falls with the square of the distance from the focal spot. Raising the table moves the patient's back away from the tube and lowers the entrance skin dose rate. Federal rules require stationary fluoroscopes to keep the source-to-skin distance at 38 cm or more (21 CFR 1020.32). The table is raised as high as the operator's working position allows.

Detector close to the patient. The detector is lowered to the patient's chest. A gap between patient and detector magnifies the image geometrically, and the automatic control raises output to keep the detector signal constant across the larger air gap. Lowering the detector reduces patient dose and scatter and sharpens the image.

Electronic magnification. Selecting a smaller field of view on the detector magnifies the image but increases the dose per unit area at the skin, because the system raises output to maintain detector signal over fewer detector elements. Magnified fields are used for the steps that need them and released afterward.

Collimation. Collimating the beam to the region of interest reduces the irradiated area. Kerma-area product falls in proportion; the air kerma at the reference point is unchanged, because it is a dose at one point. Collimation also lowers scatter and improves image contrast. In a phantom study, reducing the fluoroscopic field from 15 × 15 cm to 10 × 10 cm reduced scatter to personnel by more than 45% (Perisinakis 2016). Tight collimation reduces overlap of successive fields on the same patch of skin.

Patient size. Body size is the largest patient-related determinant of dose. In 6,623 percutaneous coronary procedures, body surface area correlated more closely with dose than body mass index, weight, or lean body mass. An increase of 1 m² in body surface area added about 164 Gy·cm² to total dose-area product, and female sex and greater height were associated with lower dose (Manicardi 2022). Obese patients also generate more scatter: more than 25% more in a phantom study (Perisinakis 2016). In obese patients, table height, detector position, magnification, and collimation have larger effects on dose, and the reference air kerma is followed closely during the procedure.

Beam angulation and projection choice

Steep projections lengthen the path of the beam through the body. The automatic control raises output, so patient entrance dose and scatter both rise. In a phantom mapping of fluoroscopic dose-area product per second and operator dose in 10-degree steps, a right anterior oblique (RAO) 20° projection produced the lowest values in that system. Compared with the posteroanterior view, patient dose rate and operator dose rose 3.7-fold and 10.6-fold toward LAO 100° and 3.7-fold and 2.4-fold toward RAO 100°. Adding 30° of cranial or caudal angulation increased both about 2.5-fold. Replacing the caudal LAO 60°/20° view of the left main coronary artery with a posteroanterior 30° caudal view reduced patient dose 2.6-fold and operator dose 8-fold. Replacing the cranial LAO 60°/20° view of the left anterior descending bifurcation with a posteroanterior 30° cranial view reduced patient dose 2.6-fold and operator dose 5-fold (Kuon 2004).

Steep LAO projections raise operator dose more than steep RAO projections because they place the x-ray tube on the patient's right, the side where the operator stands, and scatter is most intense on the tube side of the patient. The standard projections and the segments they show are described in standard angiographic projections. Projections are chosen from prior imaging so that the fewest views answer the question, and steep LAO cranial views are used only when no shallower view separates the segment. During long procedures the angle is varied so that successive fields do not land on the same patch of skin.

Beam filtration and the antiscatter grid

Spectral filtration. Copper filters, added in thicknesses that vary with mode, harden the beam by removing low-energy photons that would deposit their energy in the skin without contributing to the image. Fluoroscopy uses heavier filtration than cine, because cine requires the iodine contrast that softer beams provide. Filtration is set in the protocol by the manufacturer and physicist; the operator affects it indirectly through the choice of mode.

Wedge filters. Semitransparent wedges positioned at the edge of the field attenuate the beam over lung or other low-density regions. They prevent bright, saturated areas, improve the displayed image, and reduce dose to those regions.

Antiscatter grid. Radiation scattered in the patient reaches the detector at oblique angles and degrades contrast. The grid, a lattice of thin lead strips aligned with the focal spot and placed in front of the detector, absorbs most of these oblique photons and part of the primary beam. Automatic dose-rate control compensates for the absorbed primary photons by raising output, so the grid improves contrast and increases patient dose. In adults, the thick chest generates abundant scatter, and the grid remains in place for cardiac imaging. The FDA advisory lists removable grids, with optimum beam filtration and efficient automatic brightness control, among equipment features that help minimize skin dose (FDA 1994).

Patient dose metrics: fluoroscopy time, air kerma, kerma-area product, and peak skin dose

Four quantities describe patient dose in fluoroscopy. They are distinct from the effective dose, absorbed dose, and equivalent dose quantities described under dose quantities, units, and tissue weighting factors.

MetricUnitDefinitionPredictsLimitations
Fluoroscopy timeminCumulative time that fluoroscopy was onOperator and procedure complexity in broad termsExcludes cine, the largest dose contributor; ignores dose rate, pulse rate, field size, angulation, and patient size
Reference air kerma (cumulative air kerma, Ka,r)Gy or mGyKinetic energy released per unit mass of air at the interventional reference point, 15 cm from the isocenter toward the x-ray source, summed over fluoroscopy and acquisitionSkin dose and tissue reactionsFixed point that moves with the C-arm; ignores table height, beam angle changes that spread dose, backscatter, and table attenuation
Kerma-area product (PKA; dose-area product)Gy·cm² (or µGy·m²)Air kerma integrated over the beam area; the same at any distance from the sourceTotal energy delivered to the patient, stochastic risk, and scatter to staffDoes not localize dose; a large field at low dose and a small field at high dose give the same value
Peak skin doseGyHighest absorbed dose to any area of skinSeverity of skin injuryNot displayed directly on most systems; estimated from dose maps or reconstructed by a physicist from procedure logs

Fluoroscopic patient dose metrics. 1 µGy·m² = 0.01 Gy·cm². The reference point definition is the federal standard for C-arm systems (21 CFR 1020.32).

Fluoroscopy time is the oldest and least informative metric. Federal rules require it to be displayed at the operator's position and require an audible signal after 5 minutes of fluoroscopy (21 CFR 1020.32). It is a measure of beam-on time and excludes acquisition, so two procedures with identical fluoroscopy times can differ several-fold in dose.

Reference air kerma is measured or calculated at a fixed point on the beam axis that approximates the entrance skin surface of an average patient positioned at isocenter. Peak skin dose can be lower or higher than reference air kerma. Peak skin dose is lower when the table is high, when the angle changes spread dose over several skin fields, and when the table and pad attenuate the beam. It is higher when the patient's back lies closer to the tube than the reference point and when backscatter from the patient adds to the incident dose. In procedures with a reference air kerma above 5 Gy, the median ratio of peak skin dose to reference air kerma was 0.91 for cardiac interventions, 0.81 for body interventions, and 0.59 for neurological interventions (Fong 2022). In complex endovascular aortic procedures it averaged 0.78 (Kirkwood 2014). Reference air kerma therefore approximates peak skin dose more closely in cardiac procedures than in other interventions, but the ratio in a given patient is uncertain.

Kerma-area product is measured by a transmission chamber at the collimator or calculated from technique factors. Because it is the product of air kerma and field area, it is independent of distance from the source. It is the input for estimates of effective dose, and shielding trials report operator dose normalized to it (Medranda 2026).

Peak skin dose is the quantity that determines skin injury. It is estimated by skin dose mapping software or reconstructed after the procedure by a physicist from the system's radiation dose structured report, which records the geometry and technique of the exposures.

The US dose-display standard: 21 CFR 1020.32

The FDA performance standard for fluoroscopic equipment, revised to improve the display of dose information to physicians (FDA 2023), sets requirements for manufacturers. Fluoroscopic equipment manufactured on or after June 10, 2006, must display, at the operator's working position, the air kerma rate during fluoroscopy and the cumulative air kerma, in mGy, for all x-ray tubes in use. The air kerma rate is updated at least once per second when more than six images per second are produced, and the cumulative value is updated at least once in 5 seconds or shown within 5 seconds after the exposure ends. For C-arm systems, the reference location is 15 cm from the isocenter toward the x-ray source along the beam axis, or a manufacturer-specified point representing the intersection of the beam with the patient's skin. The display can be reset between procedures, the irradiation time is shown in minutes and tenths of a minute, and last-image hold is required (21 CFR 1020.32).

The standard governs manufacturers. The use of fluoroscopy, operator qualifications, quality assurance programs, and facility requirements are regulated by states and other agencies (FDA 2023). The kerma-area product display that most cardiac systems provide is not required by the federal standard. Problems with fluoroscopic equipment, including overexposure, are reported to the FDA through MedWatch, with the protocol, mode, pulse rate, field of view, dose rate, and displayed reference air kerma and kerma-area product included where available (FDA 2023).

Typical patient doses in cardiovascular procedures

Patient dose varies widely with procedure, equipment, protocol, operator, and patient size. The values below come from published series and illustrate the range; they are not reference levels.

Procedure and settingKerma-area productOther reported metricsSource
Diagnostic coronary angiography, single centerMedian 63 Gy·cm²Median fluoroscopy time 3 min; dose rose with body mass index classZanca 2020
PCI, single centerMedian 125 Gy·cm²Median fluoroscopy time 14 min; cine contributed most of the kerma-area productZanca 2020
Coronary angiography with or without PCI, 6,623 proceduresMedian 63 Gy·cm²Body surface area the strongest anthropometric predictor; angioplasty and more treated vessels associated with higher doseManicardi 2022
Angiography and PCI in acute coronary syndromes, randomized access65 Gy·cm² radial, 59 Gy·cm² femoralFluoroscopy time 10 min radial, 9 min femoralSciahbasi 2017
Diagnostic angiography, older versus noise-reduction systemMedian 155 vs 52 Gy·cm²PCI 229 vs 86 Gy·cm²Nakamura 2016
CTO PCI, 6 institutionsNot reportedFluoroscopy time 45 ± 25 min; maximum entrance skin dose 3.2 ± 2.1 Gy (range 0.5–10.2 Gy); skin injury in 2 of 72 patientsSuzuki 2008
Transfemoral TAVR, 15 vs 7.5 frames per secondMedian 184 vs 107 Gy·cm²Reference air kerma 0.97 vs 0.64 GyMaccagni 2018
Supraventricular tachycardia ablation, conventional fluoroscopyNot reportedMedian effective dose 8.87 mSv; fluoroscopy time 859 sCasella 2016
Three-dimensional rotational angiography of the left atriumNot reportedEffective dose 5–8 mSv conventional; 2.6 mSv with a reduced-frame protocolDe Buck 2013

Reported patient doses in adult cardiovascular procedures. Average effective doses for interventional procedures usually range from 5 to 70 mSv, against an average annual background of about 3 mSv (Mettler 2008).

Effective dose estimated from kerma-area product is useful for comparing protocols and communicating the scale of exposure. It is a reference-population quantity, and its conversion into an individual cancer risk rests on assumptions that do not apply to a single patient, as described under dose quantities.

Real-time dose monitoring and skin dose mapping

The displayed cumulative air kerma is the operator's real-time measure of skin risk. Laboratories set notification levels at which the technologist announces the running value to the operator, so that the operator can change projection, collimate, lower the pulse rate, or reassess the plan. The FDA advisory lists real-time indication of dose rate and cumulative dose among the features that aid in limiting skin dose (FDA 1994), and the SCAI program statement sets out best practices for the procedure itself (Chambers 2011).

Skin dose mapping software uses the system's record of tube angle, table position, field size, and technique to display the distribution of dose on a model of the patient's body. It shows whether dose is concentrated on one field or spread over several, and it supports the decision to rotate the beam during a long procedure. After a high-dose procedure, a physicist can reconstruct the peak skin dose from the radiation dose structured report and, when needed, confirm it with radiochromic film measurements, which agreed closely with calculated peak skin doses in one validation (Kirkwood 2014).

Dose management software that archives reference air kerma, kerma-area product, and fluoroscopy time for all procedures supports patient follow-up, laboratory audit, and operator feedback. In a center that applied a follow-up protocol based on continuous dose monitoring to 10,176 diagnostic and interventional cardiac procedures, 0.52% reached a substantial radiation dose level, and 8 patients under follow-up had some degree of skin injury (Barbosa 2025).

Radiation skin injury: dose thresholds, time course, and appearance

Radiation skin injury is a tissue reaction: it has a threshold, and its severity rises with dose above the threshold. The general biology of tissue reactions is described under biological effects of ionizing radiation. In cardiac procedures, injury appears at the beam entrance site, usually on the back, the posterior chest wall, or the axilla or upper arm when the arm lies in a lateral beam. The FDA advisory that first drew attention to these injuries listed the typical threshold doses and times of onset after a single exposure.

EffectTypical threshold skin doseTime to onsetFluoroscopy time at 0.02 Gy/minFluoroscopy time at 0.2 Gy/min
Early transient erythema2 GyHours1.7 h0.17 h
Temporary epilation3 GyAbout 3 weeks2.5 h0.25 h
Main erythema6 GyAbout 10 days5.0 h0.50 h
Permanent epilation7 GyAbout 3 weeks5.8 h0.58 h
Dry desquamation10 GyAbout 4 weeks8.3 h0.83 h
Invasive fibrosis10 GyMonths8.3 h0.83 h
Dermal atrophy11 GyMore than 14 weeks9.2 h0.92 h
Telangiectasia12 GyMore than 52 weeks10.0 h1.00 h
Moist desquamation15 GyAbout 4 weeks12.5 h1.25 h
Late erythema15 Gy6–10 weeks12.5 h1.25 h
Dermal necrosis18 GyMore than 10 weeks15.0 h1.50 h
Secondary ulceration20 GyMore than 6 weeks16.7 h1.67 h

Radiation-induced skin injuries after a single exposure, with the fluoroscopy time needed to reach the threshold at a stationary field (FDA 1994). 0.02 Gy/min is a usual fluoroscopic skin dose rate for an average patient; 0.2 Gy/min approaches the high-level limit. Acquisition adds dose at higher rates. Onset for invasive fibrosis is not given in the advisory and is shown as months.

The thresholds are typical values. Appearance of injury depends on dose rate, fractionation, the site irradiated, and the characteristics of the patient (FDA 1994). A consensus review that replaced the original table with dose bands confirmed the general pattern. For a single procedure in most patients, noticeable skin changes appear about 1 month after a peak skin dose of several gray, injury to skin and subcutaneous tissue increases with dose, and specialized wound care may be needed above 10 Gy (Balter 2010). Skin irradiated above 3 to 5 Gy often looks normal but reacts abnormally when irradiated again, so doses from earlier procedures to the same skin field are included in the estimate of the expected reaction (Balter 2010).

Clinical course. Apart from transient erythema, the effects may not appear until weeks after the exposure. By then the physicians who performed the procedure are often no longer in contact with the patient, and milder reactions are missed (FDA 1994). Main erythema and epilation appear in the first weeks, desquamation at about 4 weeks, and dermal atrophy, telangiectasia, and necrosis months to more than a year later. Necrosis and ulceration require specialized wound care (Balter 2010). A rash or ulcer of the back, chest wall, or arm with the outline of a beam field, appearing weeks after a fluoroscopic procedure, raises suspicion of radiation injury.

Frequency. Injury is uncommon at the doses of routine coronary work and is concentrated in long procedures, obese patients, and repeated procedures on the same field. In 72 CTO PCI procedures in 6 institutions, the maximum entrance skin dose was 3.2 ± 2.1 Gy, with wide differences among institutions, and two patients had skin injury (Suzuki 2008). In complex endovascular aortic procedures with a reference air kerma of 5 Gy or more and a mean peak skin dose of 6.6 Gy, scheduled follow-up visits and telephone interviews up to 1 year found no skin injury, a frequency lower than the doses predicted (Kirkwood 2014).

Patient factors that raise the risk of radiation injury

FactorMechanismResponse
Obesity and large body surface areaHigher tube output and entrance dose rate; skin closer to the tube; more scatterDose tracked from the start; tight collimation; projection rotation; high table
Repeated procedures on the same skin fieldCumulative dose; previously irradiated skin reacts abnormallyPrior procedure doses and fields reviewed before the procedure; beam angles chosen to avoid the same field
Prior radiation therapy to the chest or backResidual tissue damage lowers toleranceTreatment fields identified before the procedure
Complex procedures: CTO PCI, complex structural, multistep ablationLong fluoroscopy and many acquisitions on few fieldsStaging, dose checkpoints, low-dose protocols
Arm in the beam during lateral or oblique projectionsArm close to the tube receives a high entrance doseArms positioned out of the field
Smoking, diabetes, cytotoxic drugs, fair skinEvaluated as putative risk factors for skin injuryRecorded and taken into account in follow-up

Patient factors associated with radiation skin injury. The factors in the last row were evaluated as putative risk factors in a series of fenestrated endovascular aortic repairs in which no skin injury was found (Kirkwood 2015); their effect sizes are not established.

Substantial radiation dose levels and patient follow-up

The National Council on Radiation Protection and Measurements (NCRP), in its report on dose management for fluoroscopically guided procedures (Report 168), defines substantial radiation dose levels (SRDLs): dose values at which follow-up for possible tissue reactions is triggered. They are action levels for follow-up. They are neither thresholds for injury nor limits on the procedure. The values used in cardiovascular practice are a reference air kerma of 5 Gy, a kerma-area product of 500 Gy·cm², or a fluoroscopy time of 60 minutes (Barbosa 2025; Kirkwood 2014). When a peak skin dose estimate from mapping software or a physicist is available, it indicates the expected skin effect, and whether the sentinel event level is reached, more directly than reference air kerma does. Institutions can set procedure-specific levels. Because the ratio of peak skin dose to reference air kerma differs among services, a single reference air kerma level can overestimate or underestimate the need for follow-up (Fong 2022). Trigger levels expressed in kerma-area product have been derived for individual procedure types, including a considerably lower value for radiofrequency ablation (Struelens 2014).

PhaseActionBasis
Before the procedurePrior fluoroscopic procedures and their fields, prior radiation therapy, and body size reviewed; for an expected high-dose procedure, skin injury included in the consent discussionFDA 1994; Balter 2010
During the procedureCumulative air kerma announced at set values; projection, collimation, pulse rate, and staging reassessed at the notificationsFDA 1994; Chambers 2011
At the end of the procedureReference air kerma, kerma-area product, fluoroscopy time, and the irradiated skin fields recorded in the procedure noteFDA 1994; FDA 1995
When an SRDL is reachedPatient counseled on possible skin effects and the area at risk; follow-up arranged by visit, telephone, or photographs of the irradiated skinFDA 1994; Barbosa 2025
Follow-upSkin examination or patient contact in the weeks after the procedure, when erythema and epilation appear, and later for delayed effects; schedules in published programs include examinations at 2 and 6 weeks and 3 months, 6 months, and 1 yearKirkwood 2014; Suzuki 2008
Suspected sentinel eventPeak skin dose reconstructed by a medical physicist; doses to the same field from procedures within about 6 months summed; event reviewed under the institution's sentinel event policyWeinberg 2013; Kirkwood 2014
Injury foundReferral to dermatology; specialized wound care for high-dose injury; further irradiation of the same field avoided where possibleBarbosa 2025; Balter 2010

Patient dose management before, during, and after fluoroscopically guided procedures.

The FDA recommends that facilities record in the medical record an unambiguous identification of the areas of skin that may have received an absorbed dose approaching or exceeding the threshold for injury, by diagram or narrative. It also recommends that facilities determine which of their own procedures are likely to approach that threshold, including procedures likely to be repeated (FDA 1995). The record allows the injury to be recognized if symptoms develop and allows later procedures to avoid the same field.

Skin-dose follow-up calculator

Skin-dose follow-up

Substantial radiation dose levels and expected skin effects
Cumulative reference air kerma—
Peak skin dose—
Substantial radiation dose level—
—

Substantial radiation dose levels: reference air kerma 5 Gy, kerma-area product 500 Gy·cm², or fluoroscopy time 60 minutes in the current procedure (Barbosa 2025; Kirkwood 2014). When no peak skin dose is entered, it is estimated as 0.91 × cumulative reference air kerma, the median ratio in cardiac interventions (Fong 2022); individual ratios vary widely with table height and projection spread. Expected effects use the threshold doses of the FDA advisory (FDA 1994) and the 10-Gy wound-care level of the consensus review (Balter 2010). Sentinel event definition: peak skin dose of 15 Gy or more to a single field (Weinberg 2013). The 7.5-Gy screening value comes from one institution (Weinberg 2013).

The default values describe a complex PCI in which reference air kerma reached 6.2 Gy while kerma-area product and fluoroscopy time stayed below their levels. Reference air kerma alone reaches the SRDL, the estimated peak skin dose is 5.6 Gy, and temporary epilation is possible. If the same patient had a procedure 3 months earlier with 4.0 Gy on the same field, the cumulative reference air kerma is 10.2 Gy, the estimated peak skin dose 9.3 Gy, and a physicist estimate is obtained to exclude a sentinel event. A diagnostic angiogram with 0.8 Gy, 45 Gy·cm², and 4 minutes reaches no level, with an estimated peak skin dose of 0.7 Gy.

The Joint Commission sentinel event and dose review

The Joint Commission defines prolonged fluoroscopy resulting in a cumulative skin dose of 15 Gy or more to a single field as a reviewable sentinel event (Weinberg 2013). The dose refers to peak skin dose, summed over procedures delivered to the same field; published programs sum doses from procedures within about 6 months, consistent with Joint Commission recommendations (Kirkwood 2014). The event is then reviewed under the institution's sentinel event policy.

Peak skin dose is rarely displayed, so institutions screen with displayed metrics. One institution reviewed 183 cases meeting screening criteria of fluoroscopy time above 150 minutes or reference air kerma above 6 Gy. Of these, 54% were cardiology procedures, and reference air kerma alone identified 96.7% of the cases. When the ratio of peak skin dose to reference air kerma was applied, a reference air kerma of 7.5 Gy identified all procedures that would exceed 15 Gy in peak skin dose at that institution (Weinberg 2013). Screening values depend on equipment and practice and are set locally with the medical physicist.

Stochastic risk, cumulative exposure, and cancer risk in patients

Cancer risk from ionizing radiation is estimated with the linear no-threshold model described under stochastic effects. For fluoroscopic procedures, the kerma-area product is the input for effective dose, and the projected risk depends on the patient's age and sex and on the organs in the field, including breast tissue in women undergoing coronary procedures.

A population-based cohort study followed 82,861 adults after acute myocardial infarction without a history of cancer. Within the first year, 77% underwent at least one cardiac imaging or therapeutic procedure involving ionizing radiation, and cumulative exposure averaged 5.3 mSv per patient-year, 84% of it in the first year. Cancer risk rose with dose: per 10 mSv, age- and sex-adjusted cancer risk increased by about 3% over a mean follow-up of 5 years (hazard ratio 1.003 per mSv) (Eisenberg 2011). The association is observational and cannot exclude confounding by the illness that led to imaging. Most patients undergoing catheterization are older, and their projected risk is small beside the risk of the cardiovascular disease being treated. Risk weighs more in younger patients and in those who will undergo many procedures over a lifetime, such as patients with adult congenital heart disease, recurrent arrhythmias, or repeated revascularization.

Cumulative exposure is driven by serial testing and repeated procedures. Laboratory records that track patient dose across procedures allow the cumulative exposure to inform the choice of test and the technique of the next procedure.

Consent for a procedure that is expected to approach an SRDL, such as CTO PCI, complex structural intervention, or a repeat procedure on a previously irradiated field, includes the possibility of skin injury, its delayed appearance, and the need to report skin changes. The SCAI radiation safety program statement includes this preprocedural assessment among its best practices (Chambers 2011). For routine diagnostic angiography and uncomplicated PCI, the radiation risk is small, and the consent discussion describes it in proportion to its size. The general approach to communicating radiation risk is described under communicating radiation risk to patients.

The procedure report records the reference air kerma, kerma-area product, fluoroscopy time, and contrast volume. When an SRDL is reached, the report also records the irradiated fields, the patient notification, and the follow-up plan (FDA 1995). Consistent recording makes the dose available to the next operator and to the laboratory's quality program.

The pregnant patient: fetal dose and procedure planning

Coronary and structural procedures in pregnancy irradiate the chest, and the fetus receives dose mainly from radiation scattered inside the mother's body. Indications, fetal dose thresholds, and abdominal shielding are covered in coronary angiography, fetal radiation, and coronary CT in pregnancy. PCI is the preferred revascularization strategy for appropriate indications in pregnancy, with procedural modifications that minimize radiation exposure (Bolaji 2025). Pregnancy loss, congenital malformations, and growth retardation have threshold fetal doses above 100–200 mGy, and the risk is considered negligible at 50 mGy (Gomes 2015). Abdominal shielding can compromise automatic exposure control: in a phantom study of CT pulmonary angiography in pregnancy, shielding raised the patient effective dose by 47% (Begano 2020). Pregnancy screening and fetal dose thresholds for other radiation exposures are described under pregnancy, breastfeeding, and patients of reproductive age.

Technique determines fetal dose. Radial access, low pulse and frame rates, tight collimation to the heart, avoidance of steep angulation, fluoroscopy store in place of cine, and the most experienced available operator reduce both maternal and fetal exposure. Echocardiographic guidance reduces fluoroscopy further in structural procedures: in three patients who underwent patent foramen ovale closure during the second trimester under intracardiac echocardiographic guidance, estimated uterine doses were below 0.005 mGy (Schrale 2007). Catheter ablation in pregnancy uses electroanatomic mapping with minimal or no fluoroscopy, as described under supraventricular tachycardia in pregnancy. Contrast in pregnancy is covered under pregnancy and breastfeeding.

Scatter radiation and the patient as its source

Staff standing near the table receive almost all their dose from radiation scattered out of the patient. Leakage from the tube housing is small, and staff do not stand in the primary beam except when hands enter it. Three properties of scatter determine where staff dose accrues.

Scatter dose at the operator's eye level and waist varies with projection, field size, and patient size. Variations of ±10 cm in source-to-skin distance or ±10° in angulation of a given projection changed measured scatter by less than 10%, whereas field size and patient size had large effects (Perisinakis 2016).

Operator technique: beam-on time, distance, and hands out of the primary beam

Operator dose is set more by technique than by any single device. The highest-yield practices are the same practices that reduce patient dose: the beam is on only when the operator is watching the image, pulse rate and dose level are the lowest adequate, the field is collimated, and cine is replaced by fluoroscopy store when diagnostic quality is not needed.

Hands out of the primary beam. The dose rate inside the primary beam is many times the scatter dose rate beside the table. A hand in the field also adds attenuation, so the automatic control raises tube output and the dose to the patient and to all staff in the room. Hand dose is reduced by keeping fingers outside the collimated field, by using last-image hold to adjust devices, by working through extension tubing and torque devices at a distance from the field, and by avoiding manual injection with the hand near the beam during acquisition. An extremity dosimeter documents hand dose for operators with high workloads; ring dosimetry is described under personnel monitoring.

Position and distance. The operator stands as far from the field as the procedure allows and steps back during cine runs when power injection is used. Staff not needed at the table stand outside the scatter field or behind fixed shields.

Experience and feedback. In the RECAP trial, after correction for procedural complexity, cardiology fellows had a 34% higher operator exposure relative to dose-area product than staff interventional cardiologists (Vlastra 2019). Real-time dosimeters that display operator dose rate during the procedure give immediate feedback on position and shielding.

Radial and femoral access and operator dose

Radial access places the operator closer to the field, especially during right radial procedures with the arm at the patient's side, and early radial procedures were associated with more fluoroscopy. In the RAD-MATRIX substudy of the MATRIX trial in acute coronary syndromes, 18 expert operators wore thorax, wrist, and head dosimeters during 777 procedures randomized to radial or femoral access. The noninferiority end point for radial access was not met: operator thorax dose was 77 µSv with radial and 41 µSv with femoral access, and the difference persisted after normalization for fluoroscopy time or dose-area product. Wrist and head doses did not differ. Right and left radial access did not differ significantly (84 versus 52 µSv). In the full MATRIX population, fluoroscopy time (10 versus 9 minutes) and dose-area product (65 versus 59 Gy·cm²) were slightly higher with radial access (Sciahbasi 2017).

Radial access remains the default for coronary procedures because of its bleeding and vascular benefits, and the radiation difference is addressed with adjunctive shielding over the patient's arm and pelvis, careful arm positioning, and attention to collimation and frame rate. Randomized trials of pelvic and arm drapes in radial procedures are summarized under shields and drapes.

Lead aprons, thyroid collars, and apron integrity

Aprons. Protective aprons are rated by their lead equivalence, the thickness of lead that gives the same attenuation at a stated beam quality. Aprons of 0.25, 0.35, and 0.5 mm lead equivalence are in use (König 2023). Wraparound designs protect the back when the wearer turns away from the table, and two-piece designs separate a vest and a skirt. Aprons protect the trunk and gonads and leave the head, eyes, arms, hands, and lower legs exposed.

Lightweight and lead-free aprons. Lead-composite aprons combine lead with lighter attenuating materials, and lead-free aprons omit lead. In a comparison of aprons from seven manufacturers at tube voltages from 70 to 130 kV, new-generation and conventional lead aprons shielded similarly below 90 kV. Above 90 kV, conventional lead shielded best, and only new-generation aprons of 0.5 mm lead equivalence matched conventional 0.25- and 0.35-mm lead aprons (König 2023). The stated lead equivalence of a lightweight apron therefore depends on the beam quality at which it was measured. The weight saving of lightweight materials is also smaller than it appears once equivalent protection at cardiac beam energies is required.

Thyroid collars. The thyroid lies outside the apron and in the path of scatter directed toward the operator's neck. A thyroid collar of lead-equivalent material is worn by staff at the table.

Integrity testing. Aprons crack and tear with folding, hanging, and wear, and defects are invisible from outside. A 4-year program examined 1,011 items of protective equipment under fluoroscopy yearly. In total, 47.3% had tears, 6.0% of newly purchased items developed tears in their first year of use, and 48% of repaired items failed again the following year (Kellens 2022). Regular x-ray inspection of aprons and collars, which that program performed yearly against written rejection criteria, detects defects before they expose the wearer (Kellens 2022).

Leaded eyewear, the lens of the eye, and radiation cataract

The lens of the eye is a radiosensitive tissue that lies outside the apron. The left eye of a right-handed operator standing on the patient's right faces the scatter source: in interventional cardiologists, mean monthly lens dose was 0.83 mSv for the left eye and 0.35 mSv for the right (Thrapsanioti 2017). Posterior subcapsular lens changes are characteristic of radiation exposure; a survey with direct eye examinations found them in 50% of interventional cardiologists and 41% of nurses and technicians, against fewer than 10% of controls (Vano 2013).

Evidence of harm. In a meta-analysis of 8 studies with 2,559 participants, posterior lens opacities were more frequent in interventional cardiologists than in controls (relative risk 3.21), whereas cortical and nuclear opacities did not differ (Elmaraezy 2017). In the IC-CATARACT study, cortical and posterior subcapsular lens changes, mostly subclinical, were found in 47% of exposed participants and 17% of unexposed controls; occupational exposure was an independent predictor (odds ratio 6.07) (Karatasakis 2018).

Dose limits. The International Commission on Radiological Protection (ICRP) threshold for radiation cataract is about 0.5 Gy, and the ICRP recommends an occupational lens limit of 20 mSv per year averaged over 5 years, which European regulation adopted, as described under occupational and public dose limits. The US Nuclear Regulatory Commission (NRC) limit for licensed radioactive material remains 150 mSv per year. NCRP Commentary No. 26 recommended reducing the US occupational lens limit to an absorbed dose of 50 mGy per year (Dauer 2017). Reconstructed operator doses reach these lower limits in practice: among 61 interventional cardiologists whose cumulative lens doses were reconstructed from dosimetry records, none exceeded 0.5 Gy, but 44% were likely to have exceeded 20 mSv to the more exposed eye in at least one year (Domienik-Andrzejewska 2020).

Eyewear and shields. Leaded glasses with side shields protect the lens from scatter arriving from the side and below. Their effect depends on design, fit, and head position: in clinical measurements, lead eyewear reduced the dose to the eye nearer the tube by a factor of 1.1 to 3.4 depending on model and operator position, and the effect of the same model varied almost twofold between working periods (Domienik 2016). In annual measurements, doses outside and inside leaded glasses were 8.1 and 3.9 mSv for the left eye (Matsubara 2017). The ceiling-suspended shield reduced eye doses by a factor of 2.3 in clinical practice. Even with the shield, estimated annual lens doses without eyewear remained above or close to a 20-mSv limit, so operators with high workloads may need both (Domienik 2016).

Ceiling-suspended shields, table skirts, and radiation-absorbing drapes

Fixed and mobile shields reduce operator dose more reliably than garments, because they block scatter before it reaches the operator and they do not depend on the operator's posture.

DeviceStudyPopulationEffect on operator dose
Ceiling-suspended lead acrylic shieldDomienik 2016, clinical measurementsCoronary angiography and PCIEye dose reduced by a factor of 2.3
Disposable scatter-absorbing drape (RADPAD)ATTENUATE, randomized (Medranda 2026)1,000 proceduresOperator dose normalized to dose-area product 45.1% lower overall; 21.9% lower in 720 coronary cases; 54.4% lower in 210 structural cases (not significant)
Lead-free pelvic drapeMcCutcheon 2020, randomized632 angiography and PCI proceduresChest dose 50% lower; relative dose 57% lower; largest absolute reduction at the left finger
Reusable pelvic lead shield, radial accessde Ceuninck 2025, randomized346 procedures analyzedRelative operator dose 25% lower above and 40% lower under the apron; patient scatter dose 38% lower
Flexible multiconfiguration shield (0.5 mm lead equivalent)Davidsen 2023, randomized103 proceduresRelative operator dose 84.4% lower; no change in patient dose
Scatter-absorbing drape and suspended protection systemCammann 2023, prospective comparison229 proceduresFirst-operator exposure 66.4% lower with the drape and 93.9% lower with the suspended system; reductions for the assistant (30.6% and 38.0%) were not significant
Pelvic lead blanketGrabowicz 2017, clinical and phantom measurements100 proceduresNo significant change in eye lens dose; finger dose lower for physicians

Measured effects of fixed and disposable shielding on operator dose. Most trials report dose normalized to kerma-area product, and effects are added to those of standard aprons, collars, and table skirts.

Ceiling-suspended shield. A transparent lead acrylic shield on a ceiling arm, placed close to the patient between the field and the operator's head and upper body, protects the eyes and neck. Its effect depends on placement: a shield left far from the patient, or positioned above the operator's head, protects little. It is repositioned when the projection changes.

Table-mounted skirt. Lead curtains hanging from the table rail block scatter directed downward from the patient's back toward the operator's legs and lower trunk, the most intense scatter when the tube is under the table. Upper extensions over the table edge close the gap between skirt and ceiling shield.

Radiation-absorbing drapes and pads. Sterile disposable drapes and reusable shields placed over the patient's pelvis or arm absorb scatter leaving the patient's surface near the access site. They are placed outside the primary beam; a drape in the field raises tube output and patient dose.

Enhanced radiation protection systems and the multisociety statement

Enhanced radiation protection devices reduce exposure through engineering controls in place of reliance on wearable personal protective equipment. They include suspended full-body protection systems and shielding integrated into the imaging system and procedure room, and reported reductions in operator and staff exposure reach 99% (ACC 2026).

The SCAI/ASE/HRS/SIR/SVS expert consensus statement on enhanced radiation protection, endorsed by the American College of Cardiology (ACC), calls for mandatory implementation of these devices to meet the ALARA (as low as reasonably achievable) standard. It assigns shared responsibility to government, manufacturers, health care institutions, and professional societies. It also calls for updated federal and state regulation, standardized practice across states, real-time dosimetry, and standardized quality metrics, and it asks that laboratories be designed with enhanced protection integrated into the imaging system and room. The statement leaves the choice of system and technology to the department (Rizik 2026). It cites a 2023 SCAI survey in which about 60% of interventional cardiologists reported orthopedic injuries and 17% reported limiting time in the laboratory to reduce radiation exposure (ACC 2026).

Robotic PCI

Robotic PCI systems let the operator advance, retract, and rotate guidewires, balloons, and stents from a shielded cockpit away from the table, while a bedside cassette manipulates the devices. In the multicenter PRECISE study of 164 patients, device technical success was 98.8%, clinical procedural success 97.6% with no device-related complications, and radiation exposure to the primary operator was 95.2% lower than at the table (Weisz 2013). Staff at the bedside remain in the scatter field, and the dose-reduction measures for patient and staff continue to apply.

Brain tumors and other cancers in interventional operators

The operator's head and neck lie above the apron and, during LAO work, close to the tube side of the patient. A case series collected 31 brain and neck tumors in physicians who had worked for prolonged periods in interventional practice: 23 interventional cardiologists, 2 electrophysiologists, and 6 interventional radiologists, with a latency of 12 to 32 years. Glioblastoma accounted for 55%, and among the 26 cases with known laterality, 22 (85%) were left-sided, the side of the head more exposed to scatter (Roguin 2013). A case series cannot establish incidence or causation. The left-sided predominance is consistent with the scatter geometry, and it is a principal reason for head protection: the ceiling shield, eyewear, and the newer suspended and enclosure systems.

Surveys also report cancer among interventional cardiologists. In the 2014 SCAI membership survey of 314 operators, a small but substantial incidence of cancer was reported, and 6.9% had limited their caseload because of radiation exposure (Klein 2015).

Orthopedic injury from protective garments

Protective aprons place a sustained load on the spine, hips, and knees during long procedures. In the 2014 SCAI survey, 49.4% of operators reported at least one orthopedic injury: lumbar spine problems in 34.4%, cervical spine disease in 24.7%, and hip, knee, or ankle problems in 19.6%. Orthopedic illness correlated most strongly with age and was also associated with annual caseload (Klein 2015). Two-piece aprons, lighter materials of verified lead equivalence, adjustable monitor and table height, and above all suspended or enclosure systems that remove the need for heavy garments reduce this load.

Dosimetry for fluoroscopy staff

General requirements for personnel monitoring, the types of dosimeters, and the investigation of high readings are described under personnel monitoring. Fluoroscopy adds one specific practice: staff who wear aprons are monitored with two dosimeters. One badge, worn under the apron, records the dose to the protected trunk, and a second, worn at the collar outside the apron, records the dose to the unprotected head and neck and is used to estimate the lens dose (Matsubara 2017). A single badge worn under the apron underestimates head and lens dose, and a single badge worn outside overestimates the dose to the protected trunk. An extremity dosimeter is added for operators whose hands work close to the field, and a dedicated eye dosimeter near the left eye, outside any eyewear, estimates lens dose more directly.

Dosimeters work only when worn. Laboratories review readings at the dosimeter exchange, investigate unexpected values, and compare operators with similar case mix.

Nurses, technologists, anesthesia staff, and imaging staff

Staff other than the primary operator receive scatter that depends on their position. Nurses and technologists at the table side share the operator's scatter field and may stand where the ceiling shield and table skirt protect less. The second operator and scrub technologist work beside the first operator; anesthesia staff and echocardiographers at the head of the table are close to the beam during structural procedures, often without a fixed shield between them and the patient's upper body.

Protective devices designed around the primary operator protect others less: in the comparison of a scatter-absorbing drape and a suspended protection system, reductions for the assistant were 30.6% and 38.0% and not statistically significant, against 66.4% and 93.9% for the first operator (Cammann 2023). Echocardiographers performing transesophageal guidance of structural procedures can receive higher doses than the operator; their measured doses, recommended equipment, and ergonomic burden are described under radiation exposure and ergonomics of the interventional imager. All staff in the room wear aprons and collars of verified integrity, are monitored, and use mobile shields when their position allows.

The pregnant operator and staff member

The SCAI consensus on occupational radiation exposure to the pregnant cardiologist and technical personnel concluded that current data do not suggest a significant increase in risk to the fetus of a pregnant woman working in the catheterization laboratory. It also concluded that these data do not justify precluding pregnant physicians from performing procedures, provided that exposure is properly monitored and adequate radiation safety measures are used (Best 2011).

The declaration of pregnancy is voluntary and written. After declaration, US regulations for licensed radioactive material limit the dose to the embryo or fetus to 5 mSv for the pregnancy, as described under the declared pregnant worker; state regulations govern the use of x-ray equipment (FDA 2023). A second dosimeter worn at the waist under the apron monitors fetal dose. The NRC regulations also require efforts to avoid substantial variation above a uniform monthly exposure rate to a declared pregnant worker (10 CFR 20.1208). Practical measures include an apron that wraps fully around the abdomen, consistent use of the ceiling shield and table skirt, a position farther from the field where the work allows, and access to enhanced protection systems.

Career effects remain substantial. In the 2023 SCAI survey cited in the multisociety statement, 28% of women respondents reported being discouraged from working in the catheterization laboratory because of pregnancy or considering pregnancy, and 71% wanted the option to step away from the laboratory during pregnancy (ACC 2026). Policies that allow either continued work with monitoring or temporary reassignment, chosen by the worker, address both concerns.

Electroanatomic mapping and zero-fluoroscopy electrophysiology

Three-dimensional electroanatomic mapping systems localize catheters magnetically or by impedance and build an anatomic shell of the chamber, and intracardiac echocardiography shows the septum, catheters, and pericardium. Together they allow catheter ablation with minimal or no fluoroscopy, which protects the patient and removes the need for staff to wear lead throughout the procedure. Mapping and intracardiac echocardiography in atrial fibrillation ablation are described under the procedure: sedation, vascular access, transseptal puncture, and mapping.

StudyDesign and populationRadiation resultsEfficacy and safety
NO-PARTY (Casella 2016)Randomized, 262 patients, supraventricular tachycardia, 6 centersMedian patient effective dose 0 versus 8.87 mSv; fluoroscopy time 0 versus 859 s; operator dose 1.55 versus 25.33 µSv per procedure; no x-ray used in 72% of minimally fluoroscopic proceduresAcute success and complications similar
Khalid 2026Meta-analysis, 34 studies (5 randomized), 11,635 patients, supraventricular tachycardiaFluoroscopy time 13.75 min shorter; dose-area product lowerAcute success similar; long-term success slightly higher (risk ratio 1.02); complications similar; procedure 4.5 min longer
Debreceni 2023Meta-analysis, 7 studies, 1,593 patients, atrial fibrillationZero fluoroscopy feasible in 95.1%; fluoroscopy time and dose lowerAcute and long-term success similar; complication rate 2.76% overall and similar with both strategies; procedure time shorter

Randomized and pooled evidence for zero- and minimal-fluoroscopy catheter ablation. The NO-PARTY reduction in patient exposure corresponded to a 96% reduction in estimated cancer incidence and mortality risk (Casella 2016).

The randomized and pooled evidence for the zero-fluoroscopy technique comes from supraventricular tachycardia and atrial fibrillation ablation (Casella 2016; Khalid 2026; Debreceni 2023). In a multicenter registry, 18 pregnant women underwent ablation with a standardized zero-fluoroscopy approach; all procedures were completed without fluoroscopy and without serious maternal or fetal complications, and no adverse effects were seen in the children at 5 years (Stec 2026). Fluoroscopy remains available for transseptal puncture when echocardiographic guidance is inadequate, for complications, and for device implantation. Lead placement for pacing and defibrillator implantation still depends on fluoroscopy, and the dose-reduction measures above apply to it. Ablation outcomes are described under catheter ablation outcomes across supraventricular tachycardias.

Rotational angiography and cone-beam CT

Rotational acquisitions rotate the C-arm around the patient during a single injection. Dual-axis rotational coronary angiography records one run for the left and one for the right coronary artery along a predefined path that combines LAO and RAO motion with cranial and caudal motion. Three-dimensional rotational angiography and C-arm cone-beam CT reconstruct a volume from a single rotation of the C-arm, using the flat-panel detector as a cone-beam CT detector, and the volume can be overlaid on live fluoroscopy.

Rotational coronary angiography. Results depend on the comparison. In a meta-analysis of 3 head-to-head studies with 128 patients, dual-axis rotational angiography reduced contrast by about 17 mL and dose-area product by about 12 Gy·cm², with no difference in lesion visualization (Rezende 2026). In a randomized trial of 67 patients in an experienced laboratory using a modern system and low-dose conventional technique, rotational angiography did not reduce fluoroscopy time or dose-area product and used slightly more contrast (46 versus 42 mL) (Hell 2021).

Three-dimensional rotational angiography in electrophysiology. Left atrial three-dimensional rotational angiography had an effective dose of 5 to 8 mSv with conventional acquisition; a protocol with fewer frames and lower dose per frame produced left atrial models of good or optimal quality in 80% of patients at 2.6 mSv and right atrial models of good or optimal quality in all patients at 1.2 mSv (De Buck 2013). Added to cryoballoon ablation, three-dimensional rotational angiography more than doubled radiation dose (998 versus 419 mGy) and contrast (192 versus 83 mL) and lengthened the procedure, without improving 2-year success or complications (Velagic 2021).

Cone-beam CT in structural intervention. Rotational C-arm CT was acquired under rapid ventricular pacing during preprocedural cardiac catheterization in 57 patients with severe aortic stenosis. The resulting three-dimensional model of the aortic root predicted the coplanar deployment angle within the range of interobserver variability and gave annular, sinus, and sinotubular diameters that correlated with expert measurements (Veulemans 2016). Preprocedural gated CT remains the standard for TAVR planning, and the CT-derived projection curve allows deployment angles to be chosen before the procedure without an additional rotational acquisition, as described under access route assessment and fluoroscopic projections. Fusion of CT, echocardiography, and fluoroscopy is described under fusion of echocardiography and fluoroscopy with CT overlay.

Three-dimensional acquisitions deliver a large dose in seconds and often include contrast. They reduce total dose when they replace several conventional runs or guide a procedure that would otherwise need extensive fluoroscopy, and they add dose when they duplicate information available from prior CT. Low-dose protocols, collimation of the rotation to the region of interest, and use only when the volume will guide a decision keep their contribution small.

Radiation and contrast in peripheral endovascular procedures

Peripheral angiography and intervention differ from coronary work in ways that change radiation protection. The fields are long, the table is moved repeatedly along the legs, and digital subtraction angiography with a higher dose per frame is used for diagnosis. The operator often stands at the groin close to the field during long crossing and reentry maneuvers, below-the-knee work, and aortic procedures. A hybrid-room series included 61 complex endovascular procedures, most of them fenestrated aortic repairs, that reached a reference air kerma of 5 Gy. Reference air kerma averaged 8 Gy, and 16 patients had further procedures within 6 months that added to the same skin fields (Kirkwood 2014). The same dose metrics, SRDLs, and follow-up apply. Dose is reduced by fluoroscopy store and roadmapping in place of repeated subtraction runs, collimation to the segment treated, low pulse rates, positioning of the field so the operator's hands stay outside it, and table skirts and mobile shields placed for the leg work.

Contrast in peripheral procedures. Carbon dioxide angiography produces negative contrast on digital subtraction imaging and has no renal toxicity or allergic potential. It is used below the diaphragm and in the legs, and it is avoided in the thoracic aorta, coronary arteries, and cerebral circulation because of the risk of gas embolism, as described under contrast-sparing strategies in the catheterization laboratory. In a meta-analysis of 3 randomized trials and 5 cohort studies of peripheral angiography, contrast-associated acute kidney injury occurred in 8.6% with carbon dioxide and 15.2% with iodinated contrast (risk ratio 0.59, 95% confidence interval 0.33–1.04), a difference that did not reach statistical significance. The residual risk with carbon dioxide means that other risk factors for kidney injury still apply in these patients (Wagner 2022). Imaging choices for peripheral artery disease, including duplex ultrasound and CT and MR angiography before intervention, are described under duplex ultrasound, CT angiography, MR angiography, and catheter angiography.

Contrast-associated kidney injury and contrast reactions in the catheterization laboratory

Contrast safety in the catheterization laboratory is covered in full in iodinated and gadolinium contrast.

Radiation and contrast are linked in practice. Many dose-reduction measures, including fewer and better-planned acquisitions, avoidance of redundant views, and intravascular imaging in place of repeated angiography, also lower contrast volume, and both totals are recorded in the procedure report.

Training, credentialing, and the laboratory radiation safety program

Training and credentialing. The NCRP recommends that facilities ensure that all operators of fluoroscopic systems are trained and understand the operation of the system, including the radiation implications of all modes. It also recommends that physicians performing fluoroscopic procedures be trained and credentialed so that they can assess benefit and risk for individual patients (FDA 2023). The ACCF/AHA/HRS/SCAI clinical competence statement on physician knowledge to optimize patient safety and image quality in fluoroscopically guided invasive cardiovascular procedures defines the knowledge expected of cardiovascular operators (Hirshfeld 2004). The 2018 consensus updates its content (Hirshfeld 2018). Training covers x-ray physics, equipment operation, dose metrics, tissue reactions and stochastic risk, patient dose management, occupational protection, and regulatory requirements.

The radiation safety program. The SCAI statement describes the components of a laboratory program: essential personnel, radiation monitoring, protective shielding, imaging equipment, and training and education, with best practices before, during, and after the procedure (Chambers 2011). The FDA advisory recommends a quality assurance program for the x-ray system supervised by a qualified medical physicist, with calibrated and documented radiation output (FDA 1994). The imaging team, including the physician, technologist, and physicist, is responsible for optimized protocols, regular equipment quality control, and monitoring of patient dose (FDA 2023). The roles of the Radiation Safety Officer and authorized users under NRC licensing are described under authorized users, the Radiation Safety Officer, and training.

Quality program and dose audit. A laboratory quality program tracks reference air kerma, kerma-area product, and fluoroscopy time by procedure type and operator; compares them with local and published reference levels; identifies procedures that reached an SRDL and confirms that follow-up occurred; and reviews occupational dosimetry. Diagnostic reference levels, the 75th percentile of a dose survey used as an investigation level, are defined under diagnostic reference levels and laboratory dose audit. For interventional cardiology, reference levels are more useful when stratified by procedural complexity and body mass index, because kerma-area product and fluoroscopy time rose with body mass index class in diagnostic angiography and fluoroscopy time rose with complexity class in PCI (Zanca 2020). Patient dose differed significantly among cardiologists performing diagnostic angiography in the same laboratory, which supports operator-level feedback and continued training (Zanca 2020).

Patient dose reduction checklist

PhaseMeasureEffect
Before the procedureIndication confirmed; prior CT or angiography reviewed to plan projections and devicesFewer exploratory runs
Prior fluoroscopic doses and fields, prior radiation therapy, body size, and pregnancy status reviewedIdentification of patients at risk of skin injury or fetal exposure
Low-dose protocol selected: 7.5 frames per second or lower for fluoroscopy, low cine rate, lowest adequate dose levelLower dose rate throughout
During the procedureBeam on only while watching the image; last-image hold for planningLess beam-on time
Table high, detector close to the patient, arms out of the fieldLower entrance skin dose and scatter
Tight collimation and wedge filters; magnification only when neededLower kerma-area product and scatter
Shallow projections; steep LAO cranial views limited; beam angle varied to spread skin doseLower skin dose and operator dose
Fluoroscopy store in place of cine when diagnostic quality is not needed; short cine runsLess cine, the largest dose contributor
Cumulative air kerma announced at set values; plan reassessed; staging for complex proceduresSlower dose accumulation toward SRDLs
After the procedureReference air kerma, kerma-area product, fluoroscopy time, and contrast volume recordedData for follow-up and audit
SRDL reached: fields recorded, patient informed, follow-up arrangedEarly recognition of skin injury

Patient dose reduction before, during, and after fluoroscopically guided procedures. The phases follow the SCAI program statement (Chambers 2011), and the 2018 consensus sets out techniques that minimize exposure of both patients and medical personnel (Hirshfeld 2018).

Sources of error in fluoroscopic radiation safety

References

  1. Hirshfeld JW Jr, Ferrari VA, Bengel FM, et al. 2018 ACC/HRS/NASCI/SCAI/SCCT expert consensus document on optimal use of ionizing radiation in cardiovascular imaging: best practices for safety and effectiveness. Catheter Cardiovasc Interv. 2018;92:E35–E97. doi:10.1002/ccd.27659
  2. Rizik DG, Sutton NR, Lansky AJ, et al. SCAI/ASE/HRS/SIR/SVS expert consensus statement on enhanced radiation protection: time for mandatory and urgent action. JACC Cardiovasc Interv. 2026;19:2295–2310. doi:10.1016/j.jcin.2026.06.011
  3. American College of Cardiology. Multisociety statement calls for enhanced radiation protection in cath labs. ACC Journal Scan, July 15, 2026. acc.org
  4. Chambers CE, Fetterly KA, Holzer R, et al. Radiation safety program for the cardiac catheterization laboratory. Catheter Cardiovasc Interv. 2011;77:546–556. doi:10.1002/ccd.22867
  5. Best PJ, Skelding KA, Mehran R, et al. SCAI consensus document on occupational radiation exposure to the pregnant cardiologist and technical personnel. Catheter Cardiovasc Interv. 2011;77:232–241. doi:10.1002/ccd.22877
  6. Hirshfeld JW Jr, Balter S, Brinker JA, et al. ACCF/AHA/HRS/SCAI clinical competence statement on physician knowledge to optimize patient safety and image quality in fluoroscopically guided invasive cardiovascular procedures. J Am Coll Cardiol. 2004;44:2259–2282. doi:10.1016/j.jacc.2004.10.014
  7. US Food and Drug Administration. 21 CFR 1020.32: Fluoroscopic equipment. Electronic Code of Federal Regulations, current through November 20, 2026. eCFR
  8. US Food and Drug Administration. Fluoroscopy. Content current as of February 21, 2023. fda.gov
  9. US Food and Drug Administration. Important information for physicians and other health care professionals: avoidance of serious x-ray-induced skin injuries to patients during fluoroscopically-guided procedures. September 9, 1994. fda.gov
  10. US Food and Drug Administration. Recording information in the patient's medical record that identifies the potential for serious x-ray-induced skin injuries. September 15, 1995. fda.gov
  11. Balter S, Hopewell JW, Miller DL, Wagner LK, Zelefsky MJ. Fluoroscopically guided interventional procedures: a review of radiation effects on patients' skin and hair. Radiology. 2010;254:326–341. doi:10.1148/radiol.2542082312
  12. Barbosa PL, Barbosa AHP, Florian MC, et al. Patient assistance protocol based on continuous dose monitoring in interventional cardiological procedures. Appl Radiat Isot. 2025;224:111925. doi:10.1016/j.apradiso.2025.111925
  13. Kirkwood ML, Arbique GM, Guild JB, et al. Radiation-induced skin injury after complex endovascular procedures. J Vasc Surg. 2014;60:742–748. doi:10.1016/j.jvs.2014.03.236
  14. Kirkwood ML, Arbique GM, Guild JB, et al. Deterministic effects after fenestrated endovascular aortic aneurysm repair. J Vasc Surg. 2015;61:902–906. doi:10.1016/j.jvs.2014.11.044
  15. US Nuclear Regulatory Commission. 10 CFR Part 20: Standards for protection against radiation. Electronic Code of Federal Regulations, current through November 6, 2026. eCFR
  16. Fong G, Wunderle K. Technical note: relationship between peak skin dose and fluoroscopic Ka,r: clinical variations and application in establishing substantial radiation dose levels. Med Phys. 2022;49:935–942. doi:10.1002/mp.15441
  17. Weinberg BD, Vance A, Arbique GM, et al. Evaluation of fluoroscopic cases qualifying as potential fluoroscopic sentinel events. Acad Radiol. 2013;20:457–462. doi:10.1016/j.acra.2013.01.002
  18. Suzuki S, Furui S, Isshiki T, et al. Patients' skin dose during percutaneous coronary intervention for chronic total occlusion. Catheter Cardiovasc Interv. 2008;71:160–164. doi:10.1002/ccd.21284
  19. Struelens L, Bacher K, Bosmans H, et al. Establishment of trigger levels to steer the follow-up of radiation effects in patients undergoing fluoroscopically-guided interventional procedures in Belgium. Phys Med. 2014;30:934–940. doi:10.1016/j.ejmp.2014.09.006
  20. Kuon E, Dahm JB, Empen K, et al. Identification of less-irradiating tube angulations in invasive cardiology. J Am Coll Cardiol. 2004;44:1420–1428. doi:10.1016/j.jacc.2004.06.057
  21. Abdelaal E, Plourde G, MacHaalany J, et al. Effectiveness of low rate fluoroscopy at reducing operator and patient radiation dose during transradial coronary angiography and interventions. JACC Cardiovasc Interv. 2014;7:567–574. doi:10.1016/j.jcin.2014.02.005
  22. Maccagni D, Candilio L, Latib A, et al. Implementation of a low frame-rate protocol and noise-reduction technology to minimize radiation dose in transcatheter aortic valve replacement. J Invasive Cardiol. 2018;30:169–175. PMID 29715165.
  23. Bacci E, Chiarito M, Sanz-Sanchez J, et al. Safety and efficacy of an ultra low dose fluoroscopic protocol for chronic total occlusion recanalization. Catheter Cardiovasc Interv. 2023;101:911–917. doi:10.1002/ccd.30605
  24. Nakamura S, Kobayashi T, Funatsu A, et al. Patient radiation dose reduction using an X-ray imaging noise reduction technology for cardiac angiography and intervention. Heart Vessels. 2016;31:655–663. doi:10.1007/s00380-015-0667-z
  25. Perisinakis K, Solomou G, Stratakis J, Damilakis J. Data and methods to assess occupational exposure to personnel involved in cardiac catheterization procedures. Phys Med. 2016;32:386–392. doi:10.1016/j.ejmp.2016.02.006
  26. Manicardi M, Nocetti L, Brigidi A, et al. Anthropometric parameters and radiation doses during percutaneous coronary procedures. Phys Med. 2022;100:164–175. doi:10.1016/j.ejmp.2022.06.013
  27. Zanca F, Collard C, Alexandre N, et al. Patient exposure data and operator dose in coronary interventional procedures: impact of body-mass index and procedure complexity. Phys Med. 2020;76:38–43. doi:10.1016/j.ejmp.2020.05.006
  28. Sciahbasi A, Frigoli E, Sarandrea A, et al. Radiation exposure and vascular access in acute coronary syndromes: the RAD-Matrix trial. J Am Coll Cardiol. 2017;69:2530–2537. doi:10.1016/j.jacc.2017.03.018
  29. Mettler FA Jr, Huda W, Yoshizumi TT, Mahesh M. Effective doses in radiology and diagnostic nuclear medicine: a catalog. Radiology. 2008;248:254–263. doi:10.1148/radiol.2481071451
  30. Eisenberg MJ, Afilalo J, Lawler PR, et al. Cancer risk related to low-dose ionizing radiation from cardiac imaging in patients after acute myocardial infarction. CMAJ. 2011;183:430–436. doi:10.1503/cmaj.100463
  31. Roguin A, Goldstein J, Bar O, Goldstein JA. Brain and neck tumors among physicians performing interventional procedures. Am J Cardiol. 2013;111:1368–1372. doi:10.1016/j.amjcard.2012.12.060
  32. Klein LW, Tra Y, Garratt KN, et al. Occupational health hazards of interventional cardiologists in the current decade: results of the 2014 SCAI membership survey. Catheter Cardiovasc Interv. 2015;86:913–924. doi:10.1002/ccd.25927
  33. Elmaraezy A, Ebraheem Morra M, Tarek Mohammed A, et al. Risk of cataract among interventional cardiologists and catheterization lab staff: a systematic review and meta-analysis. Catheter Cardiovasc Interv. 2017;90:1–9. doi:10.1002/ccd.27114
  34. Karatasakis A, Brilakis HS, Danek BA, et al. Radiation-associated lens changes in the cardiac catheterization laboratory: results from the IC-CATARACT (CATaracts Attributed to RAdiation in the CaTh lab) study. Catheter Cardiovasc Interv. 2018;91:647–654. doi:10.1002/ccd.27173
  35. Dauer LT, Ainsbury EA, Dynlacht J, et al. Guidance on radiation dose limits for the lens of the eye: overview of the recommendations in NCRP Commentary No. 26. Int J Radiat Biol. 2017;93:1015–1023. doi:10.1080/09553002.2017.1304669
  36. Domienik J, Bissinger A, Grabowicz W, et al. The impact of various protective tools on the dose reduction in the eye lens in an interventional cardiology: clinical study. J Radiol Prot. 2016;36:309–318. doi:10.1088/0952-4746/36/2/309
  37. Domienik-Andrzejewska J, Brodecki M, Zmyślony M. Correlation of eye lens doses and personal dose equivalent measured on the arm of interventional cardiologists for a retrospective assessment of doses to operators' eye lens. Radiat Prot Dosimetry. 2020;189:271–278. doi:10.1093/rpd/ncaa039
  38. Matsubara K, Lertsuwunseri V, Srimahachota S, et al. Eye lens dosimetry and the study on radiation cataract in interventional cardiologists. Phys Med. 2017;44:232–235. doi:10.1016/j.ejmp.2017.10.007
  39. König AM, Verbe Zoum J, Fiebich M, Abissi PW, Mahnken AH. Comparison of the radiation protection effect of different radiation protection aprons made of different materials. Eur J Radiol. 2023;164:110862. doi:10.1016/j.ejrad.2023.110862
  40. Kellens PJ, De Hauwere A, Gossye T, et al. Integrity of personal radiation protective equipment (PRPE): a 4-year longitudinal follow-up study. Insights Imaging. 2022;13:183. doi:10.1186/s13244-022-01323-3
  41. Medranda GA, Bliagos D, Case BC. Impact of radiation-attenuating drape on exposure in a contemporary cardiac catheterization laboratory: the ATTENUATE trial. J Soc Cardiovasc Angiogr Interv. 2026;5:105511. doi:10.1016/j.jscai.2026.105511
  42. McCutcheon K, Vanhaverbeke M, Pauwels R, et al. Efficacy of MAVIG X-ray protective drapes in reducing operator radiation dose in the cardiac catheterization laboratory: a randomized controlled trial. Circ Cardiovasc Interv. 2020;13:e009627. doi:10.1161/CIRCINTERVENTIONS.120.009627
  43. de Ceuninck M, Dujardin K, Vanhaverbeke M, et al. Impact of x-ray protective drapes on operator and patient radiation exposure during cardiac catheterization. Catheter Cardiovasc Interv. 2025;105:475–482. doi:10.1002/ccd.31329
  44. Davidsen C, Ytre-Hauge K, Samnøy AT, et al. Efficacy and user experience of a novel x-ray shield on operator radiation exposure during cardiac catheterization: a randomized controlled trial. Circ Cardiovasc Interv. 2023;16:e013199. doi:10.1161/CIRCINTERVENTIONS.123.013199
  45. Cammann VL, Schweiger V, Cieslik M, et al. Effectiveness of radiation protection systems in the cardiac catheterization laboratory: a comparative study. Clin Res Cardiol. 2023;112:605–613. doi:10.1007/s00392-022-02142-8
  46. Weisz G, Metzger DC, Caputo RP, et al. Safety and feasibility of robotic percutaneous coronary intervention: PRECISE (Percutaneous Robotically-Enhanced Coronary Intervention) study. J Am Coll Cardiol. 2013;61:1596–1600. doi:10.1016/j.jacc.2012.12.045
  47. Vlastra W, Claessen BE, Beijk MA, et al. Cardiology fellows-in-training are exposed to relatively high levels of radiation in the cath lab compared with staff interventional cardiologists: insights from the RECAP trial. Neth Heart J. 2019;27:330–333. doi:10.1007/s12471-019-1254-1
  48. Casella M, Dello Russo A, Pelargonio G, et al. Near zerO fluoroscopic exPosure during catheter ablAtion of supRavenTricular arrhYthmias: the NO-PARTY multicentre randomized trial. Europace. 2016;18:1565–1572. doi:10.1093/europace/euv344
  49. Khalid A, Samad ZA, Iftikhar A, et al. Zero- or minimal-fluoroscopy vs conventional fluoroscopy in catheter ablation of supraventricular tachycardias: an updated systematic review and meta-analysis. Heart Rhythm O2. 2026;7:664–674. doi:10.1016/j.hroo.2026.01.008
  50. Debreceni D, Janosi K, Bocz B, et al. Zero fluoroscopy catheter ablation for atrial fibrillation: a systematic review and meta-analysis. Front Cardiovasc Med. 2023;10:1178783. doi:10.3389/fcvm.2023.1178783
  51. De Buck S, Alzand BS, Wielandts JY, et al. Cardiac three-dimensional rotational angiography can be performed with low radiation dose while preserving image quality. Europace. 2013;15:1718–1724. doi:10.1093/europace/eut140
  52. Velagic V, Mugnai G, Kardum D, et al. Intra-procedural three-dimensional rotational angiography in cryoballoon ablation for atrial fibrillation. Int J Cardiovasc Imaging. 2021;37:389–397. doi:10.1007/s10554-020-02029-7
  53. Hell MM, Gilg MD, Röther J, et al. Dual-axis rotational coronary angiography versus conventional coronary angiography: a randomized comparison. Clin Res Cardiol. 2021;110:258–269. doi:10.1007/s00392-020-01743-5
  54. Rezende S, Mattos F, Matoza-Báez MK. Safety and efficacy of dual-axis rotational vs standard coronary angiography: meta-analysis. Arch Cardiol Mex. 2026;96:130–136. doi:10.24875/ACM.24000180
  55. Veulemans V, Mollus S, Saalbach A, et al. Optimal C-arm angulation during transcatheter aortic valve replacement: accuracy of a rotational C-arm computed tomography based three dimensional heart model. World J Cardiol. 2016;8:606–614. doi:10.4330/wjc.v8.i10.606
  56. Wagner G, Glechner A, Persad E, et al. Risk of contrast-associated acute kidney injury in patients undergoing peripheral angiography with carbon dioxide compared to iodine-containing contrast agents: a systematic review and meta-analysis. J Clin Med. 2022;11:7203. doi:10.3390/jcm11237203
  57. Schrale RG, Ormerod J, Ormerod OJ. Percutaneous device closure of the patent foramen ovale during pregnancy. Catheter Cardiovasc Interv. 2007;69:579–583. doi:10.1002/ccd.21031
  58. Grabowicz W, Domienik-Andrzejewska J, Masiarek K, et al. Effectiveness of pelvic lead blanket to reduce the doses to eye lens and hands of interventional cardiologists and assistant nurses. J Radiol Prot. 2017;37:715–727. doi:10.1088/1361-6498/aa7a70
  59. Bolaji OA, Bahar Y, Bahar AR, et al. Percutaneous coronary intervention in pregnancy: a contemporary review. Cardiol Rev. Published online 2025. doi:10.1097/CRD.0000000000000945
  60. Gomes M, Matias A, Macedo F. Risks to the fetus from diagnostic imaging during pregnancy: review and proposal of a clinical protocol. Pediatr Radiol. 2015;45:1916–1929. doi:10.1007/s00247-015-3403-z
  61. Begano D, Söderberg M, Bolejko A. To use or not use patient shielding on pregnant women undergoing CT pulmonary angiography: a phantom study. Radiat Prot Dosimetry. 2020;189:458–465. doi:10.1093/rpd/ncaa059
  62. Stec S, Bronisz M, Kornaszewska J, et al. Zero-fluoroscopy catheter ablation for arrhythmias in pregnancy: validation of a standardized approach with 5-year follow-up. Front Cardiovasc Med. 2026;13:1859227. doi:10.3389/fcvm.2026.1859227
  63. Thrapsanioti Z, Askounis P, Datseris I, et al. Eye lens radiation exposure in Greek interventional cardiology article. Radiat Prot Dosimetry. 2017;175:344–356. doi:10.1093/rpd/ncw356
  64. Vano E, Kleiman NJ, Duran A, et al. Radiation-associated lens opacities in catheterization personnel: results of a survey and direct assessments. J Vasc Interv Radiol. 2013;24:197–204. doi:10.1016/j.jvir.2012.10.016
  65. Brar SS, Aharonian V, Mansukhani P, et al. Haemodynamic-guided fluid administration for the prevention of contrast-induced acute kidney injury: the POSEIDON randomised controlled trial. Lancet. 2014;383:1814–1823. doi:10.1016/S0140-6736(14)60689-9